{| {{table}}
The measured standard is 0.1 mm. The measured result have to be within +- 0.005 mm. The QC is preformed ones a year.
| align="center" |
{| border="1" cellspacing="1" cellpadding="2"  align="center" style="width:200px"
 
! QC Recipe:
! Wet1050
! Dry1050
|-  
| H<sub>2</sub> flow
|3 sccm
|0 sccm
|-
|O<sub>2</sub> flow
|2 sccm
|5 sccm
|-
|Temperature
|1050 C
|1050 C
|-
|Oxidation time
|30 min
|100 min
|-
|}
 
| align="center" valign="top"|
{| border="2" cellspacing="1" cellpadding="2" align="center" style="width:500px"
!QC limits
|Thickness
|Non-uniformity (both over a single wafer
and over the boat)
|-
!Dry1050
|110-116 nm
|3 %
|-
!Wet1050
|305-321 nm
|5 %
|-
|}
|-
|}
|}
 