Specific Process Knowledge/Lithography/Strip: Difference between revisions

From LabAdviser
Tigre (talk | contribs)
Tigre (talk | contribs)
Line 25: Line 25:
|style="background:LightGrey; color:black"|Minimum feature size
|style="background:LightGrey; color:black"|Minimum feature size
|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
*1.25µm down to 1.0µm
*
|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
*1.25µm
*  
|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
|
|
Line 35: Line 35:
|style="background:LightGrey; color:black"|Exposure light/filters/spectrum
|style="background:LightGrey; color:black"|Exposure light/filters/spectrum
|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
* 350W Hg-lamp, 365nm filter, 303nm filter optional
*
* intensity in Constant Intensity(CI) mode 7mJ/cm2
|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
* 350W Hg-lamp, SU8 filter, 365nm filter optional
*
* intensity in Constant Intensity(CI) mode 7mJ/cm2
|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
*  
*  
Line 49: Line 47:
|style="background:LightGrey; color:black"|Exposure mode
|style="background:LightGrey; color:black"|Exposure mode
|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
*proximity, soft, hard, vacuum contact
*
|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
*proximity, soft, hard, vacuum contact
*
*proximity, soft, hard, vacuum contact
|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
*  
*  

Revision as of 12:19, 9 August 2013

Strip Comparison Table

Equipment Plasma Asher 1 Plasma Asher 1 III-V Plasma Asher Rough Acetone Strip Fine Acetone Strip
Purpose
  • All purposes
  • clean wafers only, no metal
  • III-V materials only
Performance Minimum feature size
Exposure light/filters/spectrum
Exposure mode
Process parameter range Positive Process
Negative Process
Substrates Batch size
  • 1 small samples
  • 1 50 mm wafers
  • 1 100 mm wafers
  • 1 150 mm wafers
  • 1 50 mm wafers
  • 1 100 mm wafers
  • 25 150 mm wafers with automatic handling
  • 1 small samples
  • 1 50 mm wafers
  • 1 100 mm wafers
  • 1 150 mm wafers
Allowed materials
  • Si and silicon oxide, silicon nitride
  • Quartz, pyrex
  • Si and silicon oxide, silicon nitride
  • Quartz, pyrex
  • III-V compounds


Plasma Asher 1

The PlasmaAsher1 is placed in Cleanroom 3.

The Plasma Asher1( 300 auto load model) can be used for the following process:

  • Photoresist stripping
  • Surface cleaning after storage
  • Surface cleaning after processes using oil pump or diffusion pump vacuum
  • Surface cleaning as part of photolithography after wet developing of lacquer structures prior to wet or plasma etching
  • Stripping of photoresist layers after etching, including after being exposed to high temperatures as after implantation, ion etching, sputter etching, RIE
  • Removal of organic passivating layers and masks
  • Etching of glass and ceramic
  • Etching of SiO, SiN, Si
  • Removal of polyimide layers

The machine can be used for almost every materials, but if you have any doubt about your materials are compatible with the plasma process it is better to ask photolithography group at Danchip.


Overview of typical processes

Photoresist stripping Surface treatment of plastic, ceramic and metal Ashing of organic material
Process pressure 0.8- 1.0mbar 0.5- 1.0mbar 0.8-1.5mbar
Process gases O2, N2 O, CF, N or their mixtures O2
Process power 600-1000W 150-300W 1000W or less for heat- sensitive materials
Process time 5-60min, depending of photoresist thickness a few seconds to a few minutes Between 0.5 and 20 hours, depending on the material
Batch size 1-10 wafers at a time 1 wafer at a time 1 wafer at a time, use a container: Petri dish, evaporating dish weighing dish, beaker, etc.
Size of substrate 2"-6" 2"-6" 2"-6"
Allowed materials All All All




A typical process time for stripping of 1.5 µm AZ5214e resist is 15-25 min and for stripping of 9.5 µm AZ4562 resist is 20-35 min with the process parameters: 210ml O/min or mixture of 210ml O/min and 70ml N/min, power 1000W.

Plasma Asher 2

Plasma asher for removing AZ resist on 6" wafers: positioned in cleanroom ? (class 10 yellow room)

The Plasma Asher 2 is the same as Plasma Asher 1 but has another loading system which is more convenient for batch loading of 6inch substrates.

In this machine, only O2 and N2 gases are used for processes (in PlasmaAsher1, CF4 is used as well).

The typical process parameters when operating the equipment:

  • Photeresist stripping

Pressure: 0.8 - 1.0 mbar

Gas: O2

Power: 600 - 1000 watts

Time: 5 -30 min., depending on photoresist type and thickness

A typical process time for stripping of 1.5 um AZ5214e resist is 25 min for 6 wafers load in a boat, recipe 1.

A Descum process in manuel mode:O2:70, N2:70, power:150W, time:10min Be sure to wait for cooling if the mashine has been used at 1000W right before. At a load at 2 Fused silicawafers resist removed 0.01-01,5um

The other materials have not been tested yet.

Rough Acetone Strip

Acetone bath "rough" for removing most of the resist
Acetone bath "fine" for removing the rest of the resist incl. ultrasound

This acetone strip is only for wafers without metal and SU-8.

There are two acetone bath: one rough for stripping the most of the resist from the surface and one fine with a ultrasound for cleaning the resists remains.


Here are the main rules for acetone strip use:

  • Place the wafers in a wafer holder and put them in the first bath for 2-5 min, this time is depending how much resist you have on the surface.
  • After the rough strip place your wafers directly in the final bath, switch on for the ultra sound and strip them for 2-3 min.
  • Rinse your wafers for 4-5 min. in running water after stripping .


Lift-off wet bench

Acetone lift-off: positioned in cleanroom 3

This bench is only for wafers with metal!

Here are the main rules for lift-off bench use:

  • Place the wafers in a dedicated wafer holder.
  • Put the holder in the acetone and start the ultrasound. The strip off time is depending of resist thickness.
  • Rinse your wafers for 4-5 min. in running water after stripping.

Find more info about the lift-off process here: Specific Process Knowledge/Photolithography/AZ5214E standard resist - reverse process

Overview of acetone benches

Acetone strip Lift-off
General description

wet stripping of resist

lift-off process

Chemical solution CH3COCH3 CH3COCH3
Process temperature 20 oC 20 oC
Batch size

1-25 wafers at a time

1-25 wafer at a time

Size of substrate

4" wafers

4" wafers

Allowed materials
  • Silicon
  • Silicon Oxide
  • Silicon Nitride
  • Silicon Oxynitride
  • All metals