Specific Process Knowledge/Thin film deposition/Deposition of Nickel: Difference between revisions
Appearance
| (5 intermediate revisions by the same user not shown) | |||
| Line 71: | Line 71: | ||
| | | | ||
*Almost any that does not degas at your intended substrate temperature. See also the [ | *Almost any that does not degas at your intended substrate temperature. See also the [https://labmanager.dtu.dk/function.php?module=XcMachineaction&view=edit&MachID=429 cross-contamination sheet] | ||
| | | | ||
* Almost any that do not degas. | * Almost any that do not degas. See also the [https://labmanager.dtu.dk/function.php?module=XcMachineaction&view=edit&MachID=244 cross-contamination sheet] | ||
| | | | ||
*Almost any that does not degas at your intended substrate temperature. See cross contamination sheets for [ | *Almost any that does not degas at your intended substrate temperature. See cross contamination sheets for [https://labmanager.dtu.dk/function.php?module=XcMachineaction&view=edit&MachID=441 PC1] and [https://labmanager.dtu.dk/function.php?module=XcMachineaction&view=edit&MachID=442 PC3] | ||
|- | |- | ||
| Line 100: | Line 100: | ||
|- | |- | ||
| | | | ||
Nickel deposition is tested occasionally, around 1 time per year in each machine. You are welcome to contact staff to request an extra test of the material. | |||
'''LabManager links to procedures and data''' (require login): | |||
*[http://labmanager.dtu.dk/d4Show.php?id=5862&mach=429 QC procedure for the E-beam Evaporator (Temescal)]<br> | *[http://labmanager.dtu.dk/d4Show.php?id=5862&mach=429 QC procedure for the E-beam Evaporator (Temescal)]<br> | ||
| Line 108: | Line 109: | ||
*[http://labmanager.dtu.dk/d4Show.php?id=5862&mach=511 QC procedure for the E-beam Evaporator (10-pockets)]<br> | *[http://labmanager.dtu.dk/d4Show.php?id=5862&mach=511 QC procedure for the E-beam Evaporator (10-pockets)]<br> | ||
*[https://labmanager.dtu.dk/view_binary.php?type=data&mach=511 The newest QC data for the E-beam Evaporator (10-pockets)]<br> | *[https://labmanager.dtu.dk/view_binary.php?type=data&mach=511 The newest QC data for the E-beam Evaporator (10-pockets)]<br> | ||
{| {{table}} | {| {{table}} | ||
| align="left" | | | align="left" valign="top"| | ||
{| border=" | {| border="2" cellspacing="1" cellpadding="2" align="left" style="width:400px" | ||
! QC Recipe: | ! QC Recipe: | ||
! Standard recipes/Ni | ! Standard recipes/Ni | ||
|- | |- | ||
|Deposition rate | |Deposition rate | ||
| | |2 Å/s | ||
|- | |- | ||
|Thickness | |Thickness | ||
| Line 124: | Line 125: | ||
|Pressure | |Pressure | ||
|Below 1*10<sup>-6</sup> mbar | |Below 1*10<sup>-6</sup> mbar | ||
|- | |||
|} | |||
|- | |- | ||
|} | |} | ||
{| {{table}} | {| {{table}} | ||
| align="left" valign="top"| | | align="left" valign="top"| | ||
{| border="2" cellspacing="1" cellpadding="2" align=" | {| border="2" cellspacing="1" cellpadding="2" align="left" style="width:400px" | ||
!QC limits | !QC limits | ||
!Both E-beam Evaporators | !Both E-beam Evaporators | ||
| Line 145: | Line 149: | ||
|- | |- | ||
|} | |} | ||
Thickness is measured in 5 points with a stylus profiler. <br> | |||
Thickness is measured in 5 points with a stylus profiler (usually DektakXT). <br> | |||
|} | |} | ||