''This section is written by DTU Nanolab internal if nothing else is stated.''
''This section is written by DTU Nanolab internal if nothing else is stated.''
[[index.php?title=Category:314]]
[[Category:314]]
[[index.php?title=Category:314-Preparation]]
[[Category:314-Preparation]]


= Plunge Freezer =
= Plunge Freezer =
Line 13: Line 13:
== Leica EM GP2 ==
== Leica EM GP2 ==
Plunge freezing is a cryo-fixation method used to preserve samples in their native state prior to cryo-electron microscopy. The EM GP2 plunge freezes fluid or extremely thin samples spread on an electron microscopy grid into liquid ethane and afterwards excess fluid is removed by automatic blotting. Prior to freezing, the sample is maintained in a temperature and humidity controlled environmental chamber which is adjustable between +4°C and +60°C and room humidity to 99 %.  
Plunge freezing is a cryo-fixation method used to preserve samples in their native state prior to cryo-electron microscopy. The EM GP2 plunge freezes fluid or extremely thin samples spread on an electron microscopy grid into liquid ethane and afterwards excess fluid is removed by automatic blotting. Prior to freezing, the sample is maintained in a temperature and humidity controlled environmental chamber which is adjustable between +4°C and +60°C and room humidity to 99 %.  
The main steps of this technique is as follows:<br />
1. The sample is spread onto a glow discharged EM grid <br />
2. The liquid droplet is then blotted with filter paper until only a very thin film of fluid remains <br />
3. The grid is then rapidly plunged into a cryogen (usually liquid ethane) <br />
4. The grid is afterwards stored in a grid box submerged in liquid nitrogen until finally loaded into the cryo-electron microscope for imaging <br />




Line 72: Line 65:
[[File:IMG_9461_UC7.jpg|400px|left|thumb|Leica EM UC7 Location at DTU Nanolab Building 307 Room 906]]<br clear="all" />
[[File:IMG_9461_UC7.jpg|400px|left|thumb|Leica EM UC7 Location at DTU Nanolab Building 307 Room 906]]<br clear="all" />


'''Requirement for training:''' Completion of the DTU Epoxy course and purchased of own diamond knife. For further information about the equipment usage or training contact mktracy@dtu.dk [https://www.dtu.dk/english/service/phonebook/person?id=141983&cpid=260636&tab=0].
'''Requirement for training:'''  
* Completion of the DTU Epoxy course
* Purchased of own diamond knife.  
 
 
For further information about the equipment usage or training contact mktracy@dtu.dk [https://www.dtu.dk/english/service/phonebook/person?id=141983&cpid=260636&tab=0].


<br clear="all" />
<br clear="all" />
Line 84: Line 82:
[[File:Leica EM FC7 Cryo-Ultramicrotome.jpg|400px|left|thumb|Leica EM FC7 Location at DTU Nanolab Building 307 Room 906]]<br clear="all" />
[[File:Leica EM FC7 Cryo-Ultramicrotome.jpg|400px|left|thumb|Leica EM FC7 Location at DTU Nanolab Building 307 Room 906]]<br clear="all" />


'''Requirement for training:''' Purchased of own diamond knife. For further information about the equipment usage or training contact mktracy@dtu.dk [https://www.dtu.dk/english/service/phonebook/person?id=141983&cpid=260636&tab=0].
'''Requirement for training:'''  
* Purchased of own diamond knife.  
 
 
For further information about the equipment usage or training contact mktracy@dtu.dk [https://www.dtu.dk/english/service/phonebook/person?id=141983&cpid=260636&tab=0].


<br clear="all" />
<br clear="all" />
Line 153: Line 155:
Allows max. capacity 30 micro test tubes for 1.5/
Allows max. capacity 30 micro test tubes for 1.5/
2.0 mL and operates at 25000 ×g /15350 rpm
2.0 mL and operates at 25000 ×g /15350 rpm
* Rotor F-35-6-30
* Rotor F-35-6-30
Allows max. capacity 6 conical tubes for 50 mL or 6 conical tubes for 15 mL and operates at 7745 ×g /7830 rpm
Allows max. capacity 6 conical tubes for 50 mL or 6 conical tubes for 15 mL and operates at 7745 ×g /7830 rpm




'''Specifications:'''
'''Specifications:'''
'''Specifications:'''
* Max. rotor capacity depending on type of rotor: 30 × 1.5/2.0 mL, 6 × 50 mL, 6 × 15 mL
* Max. rotor capacity depending on type of rotor: 30 × 1.5/2.0 mL, 6 × 50 mL, 6 × 15 mL
Line 167: Line 168:




[https://www.eppendorf.com/product-media/doc/en/335227/Centrifugation_Operating-manual_Centrifuge-5430-R.pdf Instruction manual] ''- requires login'' <br />
[https://www.eppendorf.com/product-media/doc/en/335227/Centrifugation_Operating-manual_Centrifuge-5430-R.pdf Instruction manual] ''- requires login'' [[File:IMG_9457_Centrifuge 5430R.jpg|400px|thumb|left|Eppendorf centrifuge 5430R Location at DTU Nanolab Building 307 Room 903]]<br clear="all" />
 
 
[[File:IMG_9457_Centrifuge 5430R.jpg|400px|thumb|left|Eppendorf centrifuge 5430R Location at DTU Nanolab Building 307 Room 903]]<br clear="all" />


For further information on the equipment usage contact mktracy@dtu.dk [https://www.dtu.dk/english/service/phonebook/person?id=141983&cpid=260636&tab=0].
For further information on the equipment usage contact mktracy@dtu.dk [https://www.dtu.dk/english/service/phonebook/person?id=141983&cpid=260636&tab=0].
Line 188: Line 186:




[https://labmanager.dtu.dk/view_binary.php?class=MiscDocument&id=15&name=Eppendorf_Mini_Spin_plus_centrifuge_Original_Instructions.pdf Instruction manual] ''- requires login'' <br />
[https://labmanager.dtu.dk/view_binary.php?class=MiscDocument&id=15&name=Eppendorf_Mini_Spin_plus_centrifuge_Original_Instructions.pdf Instruction manual] ''- requires login'' <br />[[File:Mini Spin centrifuge.jpg|400px|thumb|left|Eppendorf Mini Spin centrifuge Location at DTU Nanolab Building 307 Room 903]]<br clear="all" />
 
[[File:Mini Spin centrifuge.jpg|400px|thumb|left|Eppendorf Mini Spin centrifuge Location at DTU Nanolab Building 307 Room 903]]<br clear="all" />
For further information on the equipment usage contact mktracy@dtu.dk [https://www.dtu.dk/english/service/phonebook/person?id=141983&cpid=260636&tab=0].
For further information on the equipment usage contact mktracy@dtu.dk [https://www.dtu.dk/english/service/phonebook/person?id=141983&cpid=260636&tab=0].


Line 205: Line 200:
*Can be used at 5 to 40 °C (80% relative humidity), in an incubator or cold room
*Can be used at 5 to 40 °C (80% relative humidity), in an incubator or cold room
*Speed range 100 to 2500 (min-1)
*Speed range 100 to 2500 (min-1)


[[File:myPlate magnetic stirrer.jpg|400px|thumb|left|myPlate magnetic stirrer Location at DTU Nanolab Building 307 Room 903]]<br clear="all" />
[[File:myPlate magnetic stirrer.jpg|400px|thumb|left|myPlate magnetic stirrer Location at DTU Nanolab Building 307 Room 903]]<br clear="all" />
Line 243: Line 239:




[[File:Magnetic hotplate stirrer.jpeg|400px|thumb|left|VWR Magnetic hotplate stirrer Location at DTU Nanolab Building 314 Room 040]]<br clear="all" />
[[File:Magnetic hotplate stirrer.jpeg|400px|thumb|left|VWR Magnetic hotplate stirrer Location at DTU Nanolab Building 307 Room 903]]<br clear="all" />
For further information on the equipment usage contact mktracy@dtu.dk [https://www.dtu.dk/english/service/phonebook/person?id=141983&cpid=260636&tab=0].
For further information on the equipment usage contact mktracy@dtu.dk [https://www.dtu.dk/english/service/phonebook/person?id=141983&cpid=260636&tab=0].


Line 299: Line 295:


[[File:Midi 40 CO2 Incubator.jpg|400px|thumb|left|Midi 40 CO2 Incubator Location at DTU Nanolab Building 307 Room 903]]<br clear="all" />
[[File:Midi 40 CO2 Incubator.jpg|400px|thumb|left|Midi 40 CO2 Incubator Location at DTU Nanolab Building 307 Room 903]]<br clear="all" />
For further information on the equipment usage contact mktracy@dtu.dk [https://www.dtu.dk/english/service/phonebook/person?id=141983&cpid=260636&tab=0].
<br clear="all" />
== Ultrasonic cleaner ==
The GT SONIC ultrasonic cleaner can be used to remove dirt, grease, corrosion or cleaning electronic and small mechanical parts. Millions of tiny bubbles are created when ultrasonic sound is passed through liquid. These tiny bubbles combined with a huge pressure create incredible cleaning results.
Specifications:
* Temperature range is from 30°C to 80°C
* Timer can be set from 1 to 99 minutes
* Capacity is 2L
Four modes are available on the intrument and are as follows:
* '''Degas mode''' - This mode is to be used after a fluid change to rapidly remove air from the fluid so it can provide maximum cleaning effectiveness from the start of the cleaning cycle.
* '''Delicate mode''' - The cleaner runs at half power which provides a less aggressive cleaning action for fragile/delicate items.
* '''Full-power mode''' - A powerful deep clean that will clean heavily contaminated items.
* '''Temperature and Timing mode''' - Allows the user to set both the fluid temperature and duration of the cleaning cycle. These parameters will change depending on the cleaning agent and item being cleaned.
[[File:Ultrasonic cleaner.jpg|400px|thumb|left|GT SONIC Ultrasonic cleaner Location at DTU Nanolab Building 307 Room 903]]<br clear="all" />
For further information on the equipment usage contact mktracy@dtu.dk [https://www.dtu.dk/english/service/phonebook/person?id=141983&cpid=260636&tab=0].
For further information on the equipment usage contact mktracy@dtu.dk [https://www.dtu.dk/english/service/phonebook/person?id=141983&cpid=260636&tab=0].


Line 308: Line 330:


Specifications:
Specifications:
* Temperature range is from 2°C to 250°C
* Temperature range is from ambient to 250°C
* Fan speed is 0 to 10
* Fan speed is 0 to 10
* Capacity is 26L
* Capacity is 26L