Specific Process Knowledge/Thermal Process/E1 Furnace Oxidation (8"): Difference between revisions

From LabAdviser
Paphol (talk | contribs)
Paphol (talk | contribs)
No edit summary
 
(10 intermediate revisions by 2 users not shown)
Line 1: Line 1:
=<span style="background:#FF2800">THIS PAGE IS UNDER CONSTRUCTION</span>[[image:Under_construction.png|200px]]=
'''Feedback to this page''': '''[mailto:thinfilm@nanolab.dtu.dk?Subject=Feed%20back%20from%20page%20http://labadviser.nanolab.dtu.dk/index.php?title=Specific_Process_Knowledge/Thermal_Process/E1_Furnace_Oxidation_(8%22) click here]'''


'''Feedback to this page''': '''[mailto:thinfilm@nanolab.dtu.dk?Subject=Feed%20back%20from%20page%20http://labadviser.nanolab.dtu.dk/index.php?title=Specific_Process_Knowledge/Thermal_Process/E1_Furnace_Oxidation_(8%22) click here]'''
''This page is written by DTU Nanolab  internal''


[[Category: Equipment |Thermal C1]]
[[Category: Equipment |Thermal C1]]
Line 9: Line 9:


==Oxidation (8") furnace (E1)==
==Oxidation (8") furnace (E1)==
[[Image:C1.JPG|thumb|300x300px|Anneal-oxide furnace (C1). Positioned in cleanroom E-6]]
[[Image:E1_furnace.JPG|thumb|300x300px|Oxidation (8") furnace (E1). Positioned in cleanroom E-6/ Photo: DTU Nanolab internal]]


The Oxidation (8") furnace (E1) is a Tempress horizontal furnace for oxidation and annealing of silicon wafers. Both 150 mm and 200 mm wafers can be processed in the furnace.
The Oxidation (8") furnace (E1) is a Tempress horizontal furnace for oxidation and annealing of silicon wafers. Both 150 mm and 200 mm wafers can be processed in the furnace.


The Oxidation (8") furnace is the top furnace tube in the E-stack furnaces, which positioned in cleanroom E-6. Most of wafers have to be RCA cleaned, before they enter the furnace. The only exceptions are brand new wafers, wafers from the A-stack furnaces, wafers from C1 furnace, wafers from the LPCVD furnaces (B- and E-stack furnaces) and wafers from PECVD4. Please check the cross contamination information in LabManager, before you use the furnace.  
The Oxidation (8") furnace is the top furnace tube in the furnace E-stack, which positioned in cleanroom E-6. Most of wafers have to be RCA cleaned, before they enter the furnace. The only exceptions are brand new wafers, wafers from the A-stack furnaces, wafers from C1 furnace, wafers from the LPCVD furnaces (B- and other E-stack furnaces) and wafers from PECVD4 (not III-V materials). Please check the cross contamination information in LabManager, before you use the furnace.  


Oxygen is using as oxidant for dry oxidation, and for wet oxidation wafer vapour generated by a steamer is used as oxidant. The oxidation recipes on the furnace are named e.g. "WET1000" and "DRY1000", where "WET" or "DRY" indicates whether it is a wet or dry oxidation process, and the number indicates the oxidation temperature.  
Oxygen is using as oxidant for dry oxidation, and for wet oxidation wafer vapour generated by a steamer is used as oxidant. The oxidation recipes on the furnace are named e.g. "WET1000" and "DRY1000", where "WET" or "DRY" indicates whether it is a wet or dry oxidation process, and the number indicates the oxidation temperature.  
Line 44: Line 44:
Oxidation:
Oxidation:
*Dry oxidation using O<sub>2</sub>  
*Dry oxidation using O<sub>2</sub>  
*Wet oxidation using H<sub>2</sub>O DI Water steam injection system ( Bronkhorst)  
*Wet oxidation using DI water steam injection system (Bronkhorst)  
|-
|-
!style="background:silver; color:black" align="center"|Performance
!style="background:silver; color:black" align="center"|Performance
Line 55: Line 55:
|style="background:LightGrey; color:black"|Process Temperature
|style="background:LightGrey; color:black"|Process Temperature
|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
*800-1100 <sup>o</sup>C
*800 - 1100 <sup>o</sup>C
|-
|-
|style="background:LightGrey; color:black"|Process pressure
|style="background:LightGrey; color:black"|Process pressure
Line 63: Line 63:
|style="background:LightGrey; color:black"|Gas flows
|style="background:LightGrey; color:black"|Gas flows
|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
*N<sub>2</sub>: 0-10 slm  
*N<sub>2</sub>: 0 - 30 slm  
*O<sub>2</sub>: 0-10 slm
*O<sub>2</sub>: 0 - 30 slm
*Steamer flow : 0-25 liter/minute
*Steamer flow : 0 - 12.5 slm (9.5 slm in standard recipes)
|-
|-
!style="background:silver; color:black" align="center" valign="center" rowspan="2"|Substrates
!style="background:silver; color:black" align="center" valign="center" rowspan="2"|Substrates
|style="background:LightGrey; color:black"|Batch size
|style="background:LightGrey; color:black"|Batch size
|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
*1-30 100 mm or 150 mm wafers (or 50 mm wafers)
*1 - 50 150 mm wafers  
*1 - 50 200 mm wafers  
|-
|-
| style="background:LightGrey; color:black"|Substrate materials allowed
| style="background:LightGrey; color:black"|Substrate materials allowed
|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
*New silicon wafers  
*New silicon wafers  
*Silicon wafers with layers of silicon oxide or silicon nitride (RCA cleaned)
*Silicon wafers with layers of silicon oxide or silicon nitride (processed wafers should be RCA cleaned)
*Wafers from the LPCVD furnaces  
*Wafers from A-stack furnaces
*Wafers from PECVD4
*Wafers from C1 furnace
*Wafers from the LPCVD furnaces (B- and other E-stack furnaces)
*Wafers from PECVD4 (not III-V materials)
|-  
|-  
|}
|}

Latest revision as of 10:56, 2 February 2023

Feedback to this page: click here

This page is written by DTU Nanolab internal


Oxidation (8") furnace (E1)

Oxidation (8") furnace (E1). Positioned in cleanroom E-6/ Photo: DTU Nanolab internal

The Oxidation (8") furnace (E1) is a Tempress horizontal furnace for oxidation and annealing of silicon wafers. Both 150 mm and 200 mm wafers can be processed in the furnace.

The Oxidation (8") furnace is the top furnace tube in the furnace E-stack, which positioned in cleanroom E-6. Most of wafers have to be RCA cleaned, before they enter the furnace. The only exceptions are brand new wafers, wafers from the A-stack furnaces, wafers from C1 furnace, wafers from the LPCVD furnaces (B- and other E-stack furnaces) and wafers from PECVD4 (not III-V materials). Please check the cross contamination information in LabManager, before you use the furnace.

Oxygen is using as oxidant for dry oxidation, and for wet oxidation wafer vapour generated by a steamer is used as oxidant. The oxidation recipes on the furnace are named e.g. "WET1000" and "DRY1000", where "WET" or "DRY" indicates whether it is a wet or dry oxidation process, and the number indicates the oxidation temperature.

Annealing can be done for silicon wafers with layers of e.g. silicon oxide, silicon nitride, polysilicon or BPSG glass (deposited in PECVD4). The annealing recipes are named e.g. "ANN1000" (for annealing at 1000 oC).

The oxidation and annealing temperature can be up to 1100 oC.


The user manual, technical information and contact information can be found in LabManager:

Oxidation (8") furnace (E1)

Process knowledge

  • Wet oxidation. More information can be found here
  • Dry oxidation. More information can be found on the oxidation page
  • Annealing. More information can be found on the annealing page

Overview of the performance of Oxidation (8") furnace and some process related parameters

Purpose
  • Oxidation of 150 mm and 200 mm wafers
  • Annealing of 150 mm and 200 mm wafers

Annealing:

  • Using N2

Oxidation:

  • Dry oxidation using O2
  • Wet oxidation using DI water steam injection system (Bronkhorst)
Performance Film thickness
  • Dry SiO2: ~ 0 nm to 300 nm (it takes too long to grow a thicker dry oxide layers)
  • Wet SiO2: ~ 0 nm to 3 µm (23 hours wet oxidation at 1100 oC)
Process parameter range Process Temperature
  • 800 - 1100 oC
Process pressure
  • 1 atm (no vacuum)
Gas flows
  • N2: 0 - 30 slm
  • O2: 0 - 30 slm
  • Steamer flow : 0 - 12.5 slm (9.5 slm in standard recipes)
Substrates Batch size
  • 1 - 50 150 mm wafers
  • 1 - 50 200 mm wafers
Substrate materials allowed
  • New silicon wafers
  • Silicon wafers with layers of silicon oxide or silicon nitride (processed wafers should be RCA cleaned)
  • Wafers from A-stack furnaces
  • Wafers from C1 furnace
  • Wafers from the LPCVD furnaces (B- and other E-stack furnaces)
  • Wafers from PECVD4 (not III-V materials)