'''Set dose calibration for OneView camera in Digital Micrograph'''
'''Set dose calibration for OneView camera in Digital Micrograph'''


Line 6: Line 13:


1. Go to "Customs" and "SetDoseCalibrationTag_OneView_mod-DTU" <br />
1. Go to "Customs" and "SetDoseCalibrationTag_OneView_mod-DTU" <br />
[[File:Oneview_1.png|600px|center|thumb|Run script]]
[[File:Oneview_1.png|600px|left|thumb|Run script]]
<br clear="all" />
<br clear="all" />


2. Type in the high tension the microscope is running in the pop-up window<br />
2. Type in the high tension the microscope is running in the pop-up window<br />
[[File:Oneview_2.png|300px|center|thumb|Select high tension]]
[[File:Oneview_2.png|300px|left|thumb|Select high tension]]
<br clear="all" />
<br clear="all" />


2. Don't change the conversion factor; just click on "OK"<br />
3. Don't change the conversion factor; just click on "OK"<br />
[[File:Oneview_3.png|300px|center|thumb|Don't change conversion factor]]
[[File:Oneview_3.png|300px|left|thumb|Don't change conversion factor]]
<br clear="all" />
<br clear="all" />
'''Feedback to this page''': '''[mailto:labadviser@nanolab.dtu.dk?Subject=Feed%20back%20from%20page%20http://labadviser.nanolab.dtu.dk/index.php/LabAdviser/314/ETEM/dose-calibration click here]'''

Latest revision as of 08:30, 23 August 2023

Feedback to this page: click here

(content by Jens Kling @DTU Nanolab, March 2020)


Set dose calibration for OneView camera in Digital Micrograph


Whenever the high tension of the microscope is changed, the dose calibration in Digital Micrograph for the OneView camera needs to be adjusted. This is done by running an installed script:

1. Go to "Customs" and "SetDoseCalibrationTag_OneView_mod-DTU"

Run script


2. Type in the high tension the microscope is running in the pop-up window

Select high tension


3. Don't change the conversion factor; just click on "OK"

Don't change conversion factor