==LASER Endpoint System==
==LASER Endpoint System==
{{Template:Author-bghe-thissection}}
The LASER Endpoint System LEP400 can be used for finding an end point during plasma etching.
The LASER Endpoint System LEP400 can be used for finding an end point during plasma etching.
It sends in LASER light through a top view port on the etch chamber down onto the sample. The light reflects from the surface and is is detected by the LEP400. The reflected signal has an intensity that depends in the layer thicknesses and refractive indexes of the layers on the sample. Reflections some the different interfaces between the sample layers interferes and therefore the intensity varies as the top layer is getting thinner during the etch (unless that top layer is opaque). This can be used to find the endpoint between two layers and it can also be used to certain thickness down in one of the layers.
It sends in LASER light through a top view port on the etch chamber down onto the sample. The light reflects from the surface and is is detected by the LEP400. The reflected signal has an intensity that depends in the layer thicknesses and refractive indexes of the layers on the sample. Reflections some the different interfaces between the sample layers interferes and therefore the intensity varies as the top layer is getting thinner during the etch (unless that top layer is opaque). This can be used to find the endpoint between two layers and it can also be used to certain thickness down in one of the layers.


A detailed manual describing how to due this is uploaded here:[[Media:EtchDirector Manual V7.0.PDF]]
A detailed manual describing how to due this is uploaded [https://labmanager.dtu.dk/view_binary.php?fileId=704 ''here (requires login)''].

Latest revision as of 12:04, 3 February 2023

Feedback to this page: click here



LASER Endpoint System

The content on this section was created by Berit Herstrøm, DTU Nanolab

The LASER Endpoint System LEP400 can be used for finding an end point during plasma etching. It sends in LASER light through a top view port on the etch chamber down onto the sample. The light reflects from the surface and is is detected by the LEP400. The reflected signal has an intensity that depends in the layer thicknesses and refractive indexes of the layers on the sample. Reflections some the different interfaces between the sample layers interferes and therefore the intensity varies as the top layer is getting thinner during the etch (unless that top layer is opaque). This can be used to find the endpoint between two layers and it can also be used to certain thickness down in one of the layers.

A detailed manual describing how to due this is uploaded here (requires login).