Template:SEM comparison table 314: Difference between revisions
Appearance
No edit summary |
mNo edit summary |
||
| (3 intermediate revisions by 2 users not shown) | |||
| Line 6: | Line 6: | ||
! Equipment | ! Equipment | ||
<!-- ![[Specific Process Knowledge/Characterization/SEM Inspect S|SEM Inspect S]] --> | <!-- ![[Specific Process Knowledge/Characterization/SEM Inspect S|SEM Inspect S]] --> | ||
![[LabAdviser/314/SEM/Nova|Nova]] | ![[LabAdviser/314/Microscopy 314-307/SEM/Nova|Nova]] | ||
![[LabAdviser/314/SEM/QFEG|QFEG]] | ![[LabAdviser/314/Microscopy 314-307/SEM/QFEG|QFEG]] | ||
![[LabAdviser/314/SEM/AFEG|AFEG]] | ![[LabAdviser/314/Microscopy 314-307/SEM/AFEG|AFEG]] | ||
![[LabAdviser/314/FIB/Helios|Helios]] | ![[LabAdviser/314/Microscopy 314-307/FIB/Helios|Helios]] | ||
|- | |- | ||
| Line 24: | Line 24: | ||
*X Ray Analysis with EDS | *X Ray Analysis with EDS | ||
*Crystallographic analysis using EBSD and both On and Off axis TKD | *Crystallographic analysis using EBSD and both On and Off axis TKD | ||
* In-situ experiments with Heating and Gas injection | |||
| | | | ||
*Conductive samples in High Vac | *Conductive samples in High Vac | ||
| Line 33: | Line 34: | ||
*Conductive samples in High Vac | *Conductive samples in High Vac | ||
*Charge reduction in Low Vac | *Charge reduction in Low Vac | ||
*X Ray Analysis with EDS and WDS | |||
| | | | ||
*Conductive samples in High Vac | *Conductive samples in High Vac | ||
| Line 65: | Line 67: | ||
** > 12nm at 3kV (SE) --> | ** > 12nm at 3kV (SE) --> | ||
| | | | ||
*High Vacuum operation: | *High Vacuum operation in Mode II: | ||
**1.0 nm at 15 kV (TLD detector and optimum working distance) | **1.0 nm at 15 kV (TLD detector and optimum working distance) | ||
**1.8 nm at 1 kV (TLD detector and optimum working distance) | **1.8 nm at 1 kV (TLD detector and optimum working distance) | ||
*Low Vacuum operation: | *Low Vacuum operation in Mode II: | ||
**1.5 nm at 10 kV (Helix detector and optimum working distance) | **1.5 nm at 10 kV (Helix detector and optimum working distance) | ||
**1.8 nm at 3 kV (Helix detector and optimum working distance) | **1.8 nm at 3 kV (Helix detector and optimum working distance) | ||
| Line 97: | Line 99: | ||
**3.0 nm at 3 kV (SE) | **3.0 nm at 3 kV (SE) | ||
| | | | ||
* Electron Column | * Electron Column Operation in Mode II | ||
**0.8nm @15kV | **0.8nm @15kV | ||
**0.9nm @1kV | **0.9nm @1kV | ||
| Line 171: | Line 173: | ||
* Z 50mm | * Z 50mm | ||
* R 360⁰ | * R 360⁰ | ||
* T 70⁰ | * T 70⁰ Manual | ||
| | | | ||
* X 150mm Piezo | * X 150mm Piezo | ||
Latest revision as of 10:46, 17 June 2025
| Equipment | Nova | QFEG | AFEG | Helios | |
|---|---|---|---|---|---|
| Purpose |
|
|
|
| |
| Equipment position | Building 314 Room 060 | Building 314 Room 011 | Building 314 Room 034 | Building 314 Room 061 | |
| Resolution | The resolution of a SEM is strongly dependent on sample type and the operator. Resolution quoted is using sputtered gold on carbon | ||||
|
|
|
| ||
| Detectors |
|
|
|
| |
| Stage specifications |
|
|
|
| |
| Options | B | C | D | E | |
| Max sample size | Consult with DTU Nanolab staff as weight, dimensions, pumping capacity and technique all play a roll in the sample size | ||||