Specific Process Knowledge/Characterization/XRD: Difference between revisions
Appearance
Created page with "=XRD SmartLab= '''Feedback to this page''': '''[mailto:labadviser@danchip.dtu.dk?Subject=Feed%20back%20from%20page%20http://labadviser.danchip.dtu.dk/... click here]''' <!-- ..." |
|||
| (92 intermediate revisions by 4 users not shown) | |||
| Line 1: | Line 1: | ||
=XRD | '''Feedback to this page''': '''[mailto:labadviser@nanolab.dtu.dk?Subject=Feed%20back%20from%20page%20http://labadviser.nanolab.dtu.dk/index.php/Specific_Process_Knowledge/Characterization/XRD click here]''' | ||
<i> Unless otherwise stated, this page is written by <b>DTU Nanolab internal</b></i> | |||
< | |||
=XRD at DTU Nanolab= | |||
We have three X-ray diffraction setups in building 346: | |||
*The [[/XRD_SmartLab|XRD SmartLab]] primarily for thin film analysis inside the cleanroom. | |||
*The [[/XRD SmartLab 9kW Rotating Anode|XRD SmartLab 9kW Rotating Anode]] multipurpose system outside the cleanroom. | |||
*The [[/XRD_Powder|XRD Powder]] for phase analysis of powders outside the cleanroom. | |||
[[ | ==Experiments performed with XRD== | ||
*[[/Process Info|List and description of possible XRD measurements with typical setup requirements]] Note mostly relevant for XRD Smartlab and XRD Rot Anode | |||
==Data analysis== | |||
For data analysis, we recommend using Rigaku SmartLab Studio for both thinfilms and basic powder analysis. | |||
If more advanced powder analysis is needed we provide access to a remote desktop with a licence for the excellent Malvern Panalytical software, HighScore. | |||
*[[/software|Installing SmartLab Studio II]] | |||
*[[/dataconversion|Converting data from XRD Powder to SmartLab Studio II]] | |||
*[[/SLSII_analysis|Guide for using SmartLab Studio II for data analysis]] | |||
*[[/HighScore_analysis|Guide for using HighScore Plus for advanced powder data analysis]] | |||
*[[/XRD_Reference_Data|How to download reference spectra for individual crystal structures]] (for DTU users and others with access to the ICSD). | |||
Apart from commercial software a wide range of free software is available online for data analysis. Please let us know if you find something you'd like to recommend! | |||
== | ==Comparison of the XRDs at Nanolab== | ||
{| border="2" cellspacing="0" cellpadding="2" | {| border="2" cellspacing="0" cellpadding="2" | ||
!colspan="2" border="none" style="background:silver; color:black;" align="center"|Equipment | !colspan="2" border="none" style="background:silver; color:black;" align="center"|Equipment | ||
|style="background:WhiteSmoke; color:black"|<b> | |style="background:WhiteSmoke; color:black"|<b>XRD SmartLab</b> | ||
|style="background:WhiteSmoke; color:black"|<b> | |style="background:WhiteSmoke; color:black"|<b>XRD RotAnode</b> | ||
|style="background:WhiteSmoke; color:black"|<b>XRD Powder</b> | |||
|- | |- | ||
!style="background:silver; color:black;" align="center" width="60"|Purpose | !style="background:silver; color:black;" align="center" width="60"|Purpose | ||
|style="background:LightGrey; color:black"| | |style="background:LightGrey; color:black"| Crystal structure analysis | ||
and thin film thickness measurement | |||
|style="background:WhiteSmoke; color:black"| | |||
*Phase ID | |||
*Crystal Size | |||
*Crystallinity | |||
*Quality and degree of orientation | |||
*3D orientation | |||
*Lattice strain | |||
*Composition | |||
*Twist | |||
*3D lattice constant | |||
*Thickness | |||
*Roughness | |||
*Density | |||
|style="background:WhiteSmoke; color:black"| | |||
*Phase ID | |||
*Crystal Size | |||
*Crystallinity | |||
*Quality and degree of orientation | |||
*3D orientation | |||
*Lattice strain | |||
*Composition | |||
*Twist | |||
*3D lattice constant | |||
*Thickness | |||
*Roughness | |||
*Density | |||
|style="background:WhiteSmoke; color:black"| | |||
*Phase ID | |||
*Crystal Size | |||
*Crystallinity | |||
|- | |||
!style="background:silver; color:black" align="center" valign="center" rowspan="6"|X-ray generator | |||
|style="background:LightGrey; color:black"| | |||
Maximum rated output | |||
|style="background:WhiteSmoke; color:black"| | |||
3 kW | |||
|style="background:WhiteSmoke; color:black"| | |||
9 kW | |||
|style="background:WhiteSmoke; color:black"| | |||
600 W | |||
|- | |||
|style="background:LightGrey; color:black"| | |||
Rated tube voltage | |||
|style="background:WhiteSmoke; color:black"| | |||
20 to 45 kV | |||
|style="background:WhiteSmoke; color:black"| | |style="background:WhiteSmoke; color:black"| | ||
20 to 45 kV | |||
|style="background:WhiteSmoke; color:black"| | |style="background:WhiteSmoke; color:black"| | ||
40 kV | |||
|- | |- | ||
|style="background:LightGrey; color:black"| | |||
|style="background: | Rated tube current | ||
|style="background:WhiteSmoke; color:black"| | |||
2 to 60 mA | |||
|style="background:WhiteSmoke; color:black"| | |style="background:WhiteSmoke; color:black"| | ||
2 to 200 mA | |||
|style="background:WhiteSmoke; color:black"| | |style="background:WhiteSmoke; color:black"| | ||
15 mA | |||
|- | |- | ||
|style="background:LightGrey; color:black"| | |style="background:LightGrey; color:black"| | ||
Type | |||
|style="background:WhiteSmoke; color:black"| | |||
Sealed tube | |||
|style="background:WhiteSmoke; color:black"| | |style="background:WhiteSmoke; color:black"| | ||
Rotating Anode | |||
|style="background:WhiteSmoke; color:black"| | |style="background:WhiteSmoke; color:black"| | ||
Sealed tube | |||
|- | |- | ||
|style="background:LightGrey; color:black"| | |||
|style="background: | Target | ||
|style="background:WhiteSmoke; color:black"| | |||
Cu | |||
|style="background:WhiteSmoke; color:black"| | |style="background:WhiteSmoke; color:black"| | ||
Cu | |||
|style="background:WhiteSmoke; color:black"| | |style="background:WhiteSmoke; color:black"| | ||
Cu | |||
|- | |- | ||
|style="background:LightGrey; color:black"| | |style="background:LightGrey; color:black"| | ||
Focus size | |||
|style="background:WhiteSmoke; color:black"| | |style="background:WhiteSmoke; color:black"| | ||
0.4 mm x 8 mm (Line/Point) | |||
|style="background:WhiteSmoke; color:black"| | |style="background:WhiteSmoke; color:black"| | ||
* | 0.1-0.5 mm x 8 mm (Line/Point) | ||
|style="background:WhiteSmoke; color:black"| | |||
0.4 mm x 12 mm (Line) | |||
|- | |||
!style="background:silver; color:black" align="center" valign="center" rowspan="4"|Goniometer | |||
|style="background:LightGrey; color:black"| | |||
Scanning mode | |||
|style="background:WhiteSmoke; color:black"| | |||
incident / receiver coupled or independent | |||
|style="background:WhiteSmoke; color:black"| | |||
incident / receiver coupled or independent | |||
|style="background:WhiteSmoke; color:black"| | |||
incident / receiver coupled | |||
|- | |||
|style="background:LightGrey; color:black"| | |||
Goniomenter radius | |||
|style="background:WhiteSmoke; color:black"| | |||
300 mm | |||
|style="background:WhiteSmoke; color:black"| | |||
300 mm | |||
|style="background:WhiteSmoke; color:black"| | |||
145 mm | |||
|- | |||
|style="background:LightGrey; color:black"| | |||
Minimum step size | |||
|style="background:WhiteSmoke; color:black"| | |||
0.0001° (0.36") | |||
|style="background:WhiteSmoke; color:black"| | |||
0.0001° (0.36") | |||
|style="background:WhiteSmoke; color:black"| | |||
0.001° (3.6") | |||
|- | |||
|style="background:LightGrey; color:black"| | |||
Sample stage motion | |||
|style="background:WhiteSmoke; color:black"| | |||
*χ:-5~+95° | |||
*φ:0~360° | |||
*Z:-4~+1 mm | |||
*X,Y:±50 mm for a 100 mm wafer | |||
*Rx,Ry:-5~+5° | |||
|style="background:WhiteSmoke; color:black"| | |||
*χ:-5~+95° | |||
*φ:0~360° | |||
*Z:-4~+1 mm | |||
*X,Y:±50 mm for a 100 mm wafer | |||
*Rx,Ry:-5~+5° | |||
|style="background:WhiteSmoke; color:black"| | |||
Fixed with rotation | |||
|- | |||
!style="background:silver; color:black" align="center" valign="center" rowspan="2"|Optics | |||
|style="background:LightGrey; color:black"|Incident side | |||
|style="background:WhiteSmoke; color:black"| | |||
*Cross Beam Optics(CBO) | |||
*Ge(220)x2 monochromator | |||
*In-Plane Parallel Slit Collimator (PSC) | |||
*Soller slit | |||
*Automatic variable divergence slit | |||
*Length limiting slits | |||
|style="background:WhiteSmoke; color:black"| | |||
*Cross Beam Optics(CBO) | |||
*Ge(400)x2 monochromator | |||
*In-Plane Parallel Slit Collimator (PSC) | |||
*Soller slit | |||
*Automatic variable divergence slit | |||
*Length limiting slits | |||
|style="background:WhiteSmoke; color:black"| | |||
*0.04° soller slit | |||
*Ni and Cu filter | |||
*Divergence slits | |||
*Beam masks | |||
|- | |||
|style="background:LightGrey; color:black"|Receiver side | |||
|style="background:WhiteSmoke; color:black"| | |||
*Automatic variable scattering slit | |||
*Automatic variable receiver slit | |||
*Parallel slit analysers (PSA) | |||
*Ge(220)x2 analyser | |||
|style="background:WhiteSmoke; color:black"| | |||
*Automatic variable scattering slit | |||
*Automatic variable receiver slit | |||
*Parallel slit analysers (PSA) | |||
*Ge(400)x2 analyser | |||
|style="background:WhiteSmoke; color:black"| | |||
*0.04° soller slit | |||
*Ni filter | |||
|- | |- | ||
!style="background:silver; color:black" align="center" valign="center" rowspan="3"|Substrates | !style="background:silver; color:black" align="center" valign="center" rowspan="3"|Substrates | ||
|style="background:LightGrey; color:black"| | |style="background:LightGrey; color:black"|Measurement temperature | ||
|style="background:WhiteSmoke; color:black"| | |||
Room temperature | |||
|style="background:WhiteSmoke; color:black"| | |||
Room temperature | |||
|style="background:WhiteSmoke; color:black"| | |||
May be heated in N<sub><sub>2</sub></sub> up to 500°C | |||
|- | |||
|style="background:LightGrey; color:black"|Substrate size | |||
|style="background:WhiteSmoke; color:black"| | |||
up to 150 mm wafers | |||
Thickness max 21 mm | |||
|style="background:WhiteSmoke; color:black"| | |style="background:WhiteSmoke; color:black"| | ||
up to 150 mm wafers | |||
Thickness max 21 mm | |||
|style="background:WhiteSmoke; color:black"| | |style="background:WhiteSmoke; color:black"| | ||
Only for powders | |||
|- | |- | ||
| style="background:LightGrey; color:black"|Allowed materials | | style="background:LightGrey; color:black"|Allowed materials | ||
|style="background:WhiteSmoke; color:black"| | |style="background:WhiteSmoke; color:black"| | ||
All materials approved in the cleanroom. | |||
No powders or dusty materials. | |||
|style="background:WhiteSmoke; color:black"| | |||
All materials have to be approved | |||
|style="background:WhiteSmoke; color:black"| | |style="background:WhiteSmoke; color:black"| | ||
All materials have to be approved | |||
|- | |- | ||
|} | |} | ||
<br clear="all" /> | <br clear="all" /> | ||