''This page is written by DTU Nanolab  internal''


[[image:IMAG0188.jpg|450x450px|right|thumb|The SEM Supra 2 located in cleanroom C-1]]
[[image:IMAG0188.jpg|450x450px|right|thumb|The SEM Supra 2 is located in cleanroom C-1. {{photo1}} ]]


=SEM Supra 2=
=SEM Supra 2=


The SEM Supra 2 a scanning electron microscope. It produces enlarged images of a variety of specimens, achieving magnifications of over 500.000x providing ultra high resolution imaging. This important and widely used analytical tool provides exceptional resolution and depth of field and requires minimal specimen preparation.
The SEM Supra 2 is a scanning electron microscope. It produces enlarged images of a variety of specimens, achieving magnifications of over 500.000x providing ultra high resolution imaging. This important and widely used analytical tool provides exceptional resolution and depth of field and requires minimal specimen preparation.


The SEM is a VP (variable pressure) instrument - Indicating that it is capable of operating at variable pressure. By increasing the pressure in the chamber it is possible to image isolating samples. The higher density of gas molecules will eliminate the charges at the cost of slightly reduced resolution. Also, the Se2 and InLens detectors will no longer work.
The SEM is a VP (variable pressure) instrument - Indicating that it is capable of operating at variable pressure. By increasing the pressure in the chamber it is possible to image isolating samples. The higher density of gas molecules will eliminate the charges at the cost of slightly reduced resolution. Also, the Se2 and InLens detectors will no longer work.


This SEM has large chamber volume compared to the other SEMs at Danchip. An 8" wafer can be loaded in the chamber, however the X and Y ranges only provide full view of a 6" wafer.
This SEM has large chamber volume compared to the other SEM's in the cleanroom at DTU Nanolab. An 8" wafer can be loaded in the chamber, however the X and Y ranges only provide full view of a 6" wafer.


All samples must be introduced into the chamber using an airlock. The airlock greatly reduces sample exchange times, and it accepts up to 8" wafer holders.  
All samples must be introduced into the chamber using an airlock. The airlock greatly reduces sample exchange times, and it accepts up to 8" wafer holders.  
Line 37: Line 38:


!colspan="2" border="none" style="background:silver; color:black;" align="center"|Equipment  
!colspan="2" border="none" style="background:silver; color:black;" align="center"|Equipment  
|style="background:WhiteSmoke; color:black"|<b>SEM Supra 2 (Supra 60VP SEM)</b>
|style="background:WhiteSmoke; color:black"|<b>SEM Supra 2 </b>
|-
|-
!style="background:silver; color:black" align="center" valign="center" rowspan="1"|Purpose
!style="background:silver; color:black" align="center" valign="center" rowspan="1"|Purpose
|style="background:LightGrey; color:black"|Imaging and measurement of
|style="background:LightGrey; color:black"|Imaging and measurement of
|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
*Any sample except bulk insulators such as polymers, glass or quartz wafers
* Conducting samples
* Semi-conducting samples
* Thin (~ 5 µm <) layers of non-conducting materials such as polymers
* Thick polymers, glass or quartz samples
|-
|-
!style="background:silver; color:black;" align="center" width="60"|Location  
!style="background:silver; color:black;" align="center" width="60"|Location  
|style="background:LightGrey; color:black"|
|style="background:LightGrey; color:black"|
|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
*Cleanroom of DTU Danchip
*Cleanroom of DTU Nanolab
|-
|-
!style="background:silver; color:black;" align="center" width="60"|Performance
!style="background:silver; color:black;" align="center" width="60"|Performance
Line 94: Line 98:
*Any standard cleanroom material
*Any standard cleanroom material
|-  
|-  
|}
==Quality control of length measurement ==
{| border="1" cellspacing="2" cellpadding="2" colspan="3"
|bgcolor="#98FB98" |'''Quality Control (QC) for SEM Supra 2'''
|-
|
*[http://labmanager.dtu.dk/d4Show.php?id=3531&mach=327 QC procedure for SEM Supra 2]<br>
*[https://labmanager.dtu.dk/view_binary.php?type=data&mach=327 Newest QC data for SEM Supra 2]<br>
'''QC limits:'''
{| {{table}}
| align="center" valign="top"|
{| border="2" cellspacing="1" cellpadding="2" align="center" style="width:500px"
! Detector
! Settings
! Magnification
! Measured dimensions
! Calibration limit
! Action limit
|-
|SE2
| EHT 10 kV, WD 10 mm
| 2.500 k
| 70x70 μm
| ± 2 %
| ± 3 %
|-
|SE2
| EHT 10 kV, WD 10 mm
| 17.000 k
| 10x10 μm
| ± 2 %
| ± 3 %
|-
|-
|SE2
| EHT 10 kV, WD 10 mm
| 75.000 k
| 2x2 μm
| ± 2 %
| ± 3 %
|-
|SE2
| EHT 5 kV, WD 5 mm
| 17.000 k
| 10x10 μm
| ± 2 %
| ± 3 %
|-
| VPSE
| EHT 10 kV, WD 10 mm, 17 Pa
| 17.000 k
| 10x10 μm
| ± 2 %
| ± 3 %
|-
| InLens
| EHT 5 kV, WD 5 mm
| 200.000 k
| 1x1 μm
| ± 2 %
| ± 3 %
|-
| InLens
| EHT 10 kV, WD 3 mm
| 1.000.000 k
| (imaging only)
| NA
| NA
|-
|}
|-
|}
|}
|}

Latest revision as of 10:49, 28 June 2023

Feedback to this page: click here

This page is written by DTU Nanolab internal

The SEM Supra 2 is located in cleanroom C-1. Photo: DTU Nanolab internal

SEM Supra 2

The SEM Supra 2 is a scanning electron microscope. It produces enlarged images of a variety of specimens, achieving magnifications of over 500.000x providing ultra high resolution imaging. This important and widely used analytical tool provides exceptional resolution and depth of field and requires minimal specimen preparation.

The SEM is a VP (variable pressure) instrument - Indicating that it is capable of operating at variable pressure. By increasing the pressure in the chamber it is possible to image isolating samples. The higher density of gas molecules will eliminate the charges at the cost of slightly reduced resolution. Also, the Se2 and InLens detectors will no longer work.

This SEM has large chamber volume compared to the other SEM's in the cleanroom at DTU Nanolab. An 8" wafer can be loaded in the chamber, however the X and Y ranges only provide full view of a 6" wafer.

All samples must be introduced into the chamber using an airlock. The airlock greatly reduces sample exchange times, and it accepts up to 8" wafer holders.

The SEM is equipped with an Oxford Instruments X-MaxN 50 EDX detector and AZtec software. EDX (Energy dispersive X-ray) detection is used for element composition analysis in a small area. A specific training is needed for users that want to use the EDX detector.

The SEM Supra2 in located in the cleanroom. It was installed in December 2013.


The user manual, control instruction, the user APV and contact information can be found in LabManager:

SEM Supra 2 info page in LabManager


Performance information


Equipment performances

Equipment SEM Supra 2
Purpose Imaging and measurement of
  • Conducting samples
  • Semi-conducting samples
  • Thin (~ 5 µm <) layers of non-conducting materials such as polymers
  • Thick polymers, glass or quartz samples
Location
  • Cleanroom of DTU Nanolab
Performance Resolution
  • 1-2 nm (limited by vibrations)

The resolution is strongly dependent on the type of sample and the skills of the operator.

Instrument specifics Detectors
  • Secondary electron (Se2)
  • Inlens secondary electron (Inlens)
  • 4 Quadrant Backscatter electron (QBSD)
  • Variable pressure secondary electron (VPSE)
Stage
  • X, Y: 150 × 150 mm
  • T: -10 to 70o
  • R: 360o
  • Z: 50 mm
Electron source
  • FEG (Field Emission Gun) source
Operating pressures
  • Fixed at High vacuum (2 × 10-4mbar - 10-6mbar)
  • Variable at Low vacuum (0.1 mbar - 2 mbar)
Options
  • Antivibration platform
  • Fjeld M-200 airlock taking up to 8" wafers
  • Oxford Instruments X-MaxN 50 mm2 SDD EDX detector and AZtec software package
Substrates Batch size
  • Up to 8" wafer with 6" view
Allowed materials
  • Any standard cleanroom material

Quality control of length measurement

Quality Control (QC) for SEM Supra 2

QC limits:

Detector Settings Magnification Measured dimensions Calibration limit Action limit
SE2 EHT 10 kV, WD 10 mm 2.500 k 70x70 μm ± 2 % ± 3 %
SE2 EHT 10 kV, WD 10 mm 17.000 k 10x10 μm ± 2 % ± 3 %
SE2 EHT 10 kV, WD 10 mm 75.000 k 2x2 μm ± 2 % ± 3 %
SE2 EHT 5 kV, WD 5 mm 17.000 k 10x10 μm ± 2 % ± 3 %
VPSE EHT 10 kV, WD 10 mm, 17 Pa 17.000 k 10x10 μm ± 2 % ± 3 %
InLens EHT 5 kV, WD 5 mm 200.000 k 1x1 μm ± 2 % ± 3 %
InLens EHT 10 kV, WD 3 mm 1.000.000 k (imaging only) NA NA