Specific Process Knowledge/Wafer and sample drying: Difference between revisions
(54 intermediate revisions by 8 users not shown) | |||
Line 1: | Line 1: | ||
'''Feedback to this page''': '''[mailto:labadviser@danchip.dtu.dk?Subject=Feed%20back%20from%20page%20http://labadviser.danchip.dtu.dk/index.php/Specific_Process_Knowledge/Wafer_and_sample_drying click here]''' | '''Feedback to this page''': '''[mailto:labadviser@danchip.dtu.dk?Subject=Feed%20back%20from%20page%20http://labadviser.danchip.dtu.dk/index.php/Specific_Process_Knowledge/Wafer_and_sample_drying click here]''' | ||
'''Unless anything else is stated, everything on this page, text and pictures are made by DTU Nanolab.''' | |||
=Drying Comparison Table= | =Drying Comparison Table= | ||
Line 7: | Line 8: | ||
|-style="background:lightgray; color:black" | |-style="background:lightgray; color:black" | ||
!Equipment | !Equipment | ||
|[[Specific_Process_Knowledge/Wafer_and_sample_drying # | |||
|[[Specific_Process_Knowledge/Wafer_and_sample_drying #Spin dryer 2|Spin dryer 2]] | |||
|[[Specific_Process_Knowledge/Wafer_and_sample_drying #Spin | |[[Specific_Process_Knowledge/Wafer_and_sample_drying #Spin dryer 3|Spin dryer 3]] | ||
|[[Specific_Process_Knowledge/Wafer_and_sample_drying #Spin | |[[Specific_Process_Knowledge/Wafer_and_sample_drying #Spin dryer 4|Spin dryer 4]] | ||
|[[Specific_Process_Knowledge/Wafer_and_sample_drying # | |[[Specific_Process_Knowledge/Wafer_and_sample_drying #Spin dryer 5|Spin dryer 5]] | ||
| | |[[Specific_Process_Knowledge/Wafer_and_sample_drying/Critical Point Dryer|Critical point dryer]] | ||
|[[Specific_Process_Knowledge/Wafer_and_sample_drying | |||
|[[Specific_Process_Knowledge/Wafer_and_sample_drying #Ethanol fume dryer|Ethanol fume dryer]] | |[[Specific_Process_Knowledge/Wafer_and_sample_drying #Ethanol fume dryer|Ethanol fume dryer]] | ||
|[[Specific_Process_Knowledge/Wafer_and_sample_drying #Nitrogen guns|N<sub>2</sub> guns]] | |[[Specific_Process_Knowledge/Wafer_and_sample_drying #Nitrogen guns|N<sub>2</sub> guns]] | ||
Line 21: | Line 20: | ||
|-style="background:WhiteSmoke; color:black" | |-style="background:WhiteSmoke; color:black" | ||
!Location | !Location | ||
| | | | ||
*B-1 | *B-1 | ||
Line 28: | Line 25: | ||
*D-3 | *D-3 | ||
| | | | ||
* | *D-3 | ||
| | | | ||
* | *F-3 | ||
| | | | ||
*D-3 | *D-3 | ||
Line 45: | Line 38: | ||
|-style="background:lightgray; color:black" | |-style="background:lightgray; color:black" | ||
!Purpose | !Purpose | ||
| | | | ||
*Drying | *Drying | ||
| | | | ||
*Drying | *Drying | ||
Line 60: | Line 46: | ||
| | | | ||
*Drying | *Drying | ||
*Rinsing + drying | |||
| | | | ||
*Drying sensitive samples. E.g. with cantilevers | *Drying sensitive samples. E.g. with cantilevers | ||
Line 71: | Line 58: | ||
!Batch size | !Batch size | ||
| | | | ||
* | *100 mm wafers: 1-25 | ||
*150 mm wafers: 1-25 | |||
* | |||
| | | | ||
* | *100 mm wafers: 1-25 | ||
*150 mm wafers: 1-25 | |||
| | | | ||
* | *100 mm wafers: 1-25 | ||
*150 mm wafers: 1-25 | |||
| | | | ||
* | *100 mm wafers: 1-25 | ||
*150 mm wafers: 1-25 | |||
*200 mm wafers: 1-25 | |||
| | | | ||
*1 to 5 wafers per run. Sizes: 2”, 4" or 6" | *1 to 5 wafers per run. Sizes: 2”, 4" or 6" | ||
* Pieces (up to 10x10mm) | * Pieces (up to 10x10mm) | ||
| | | | ||
* | *50 mm wafers: 1-25 | ||
*100 mm wafers: 1-25 | |||
*Pieces if a suitable carrier is available | *Pieces if a suitable carrier is available | ||
| | | | ||
Line 100: | Line 83: | ||
|-style="background:lightgray; color:black" | |-style="background:lightgray; color:black" | ||
!Allowed materials | !Allowed materials | ||
| | | | ||
*Only for RCA cleaned wafers | *Only for RCA cleaned wafers | ||
| | | | ||
*No restrictions | *No restrictions | ||
Line 135: | Line 112: | ||
'''The user manual, technical information, and contact information can be found in LabManager: | '''The user manual, technical information, and contact information can be found in LabManager: [http://labmanager.danchip.dtu.dk/function.php?module=Machine&view=view&mach=157 Spin dryer 2] [http://labmanager.danchip.dtu.dk/function.php?module=Machine&view=view&mach=158 Spin dryer 3]''' | ||
== Process information == | == Process information == | ||
Standard process is | Standard process is: | ||
#Rinse (with DI water) for 60 seconds at 500 RPM | |||
#Dry with nitrogen for 150 seconds at 1800 RPM | |||
#Dry with nitrogen and anti-static purge for 30 seconds at 1800 RPM | |||
All spin dryers use the following carriers: | |||
*100 mm wafers: A72-40MB | |||
*150 mm wafers: A182-60MB | |||
*200 mm wafers: A192-80M | |||
< | <span style="color:red">It is not allowed to use other carriers in the spin dryers, as this may damage the motor due to incorrect mass distribution. It may also cause fragile wafers to break due to excessive vibrations caused by the mass imbalance.</span> | ||
== Spin dryer 2 == | |||
[[File:spinDryer_2.jpg|640px|thumb|right]] | |||
Spin dryer 2 is for RCA processing only, and is located in B-1 next to the RCA bench. | |||
*Top chamber: 100 mm wafers | |||
*Bottom chamber: 150 mm wafers | |||
<br clear="all" /> | <br clear="all" /> | ||
= | == Spin dryer 3 == | ||
[[File:spinDryer_3.jpg|640px|thumb|right]] | |||
Spin dryer 3 is located in D-3 between the working table and wet bench 03. | |||
*Top chamber: 100 mm wafers | |||
*Bottom chamber: 150 mm wafers | |||
<br clear="all" /> | <br clear="all" /> | ||
= | == Spin dryer 4 == | ||
[[ | [[File:spinDryer_4.jpg|640px|thumb|right]] | ||
Spin dryer 4 is located in D-3 between the HF vapor phase etcher and wet bench 07. | |||
*Top chamber: 100 mm wafers | |||
*Bottom chamber: 150 mm wafers | |||
<br clear="all" /> | |||
== Spin dryer 5 == | |||
[[File:spinDryer_5.jpg|640px|thumb|right]] | |||
Spin dryer 5 is located in F-3 Next to the Developer: TMAH stepper. It has a single process chamber that is capable of processing 100 mm, 150 mm and 200 mm wafers. The wafer size is easily changed, by changing the rotor insert frame. | |||
<br clear="all" /> | <br clear="all" /> | ||
Line 168: | Line 169: | ||
= Ethanol fume dryer = | = Ethanol fume dryer = | ||
If your wafers are fragile or having thin membranes, cantilevers or suspended bridges that risk to stick or adhere to the surface it can be a good idea to dry your wafers in the ethanol fume dryer. Ethanol is heated to 70°C and your wafers are placed in the fumes of Ethanol which will make the water | If your wafers are fragile or having thin membranes, cantilevers or suspended bridges that risk to stick or adhere to the surface it can be a good idea to dry your wafers in the ethanol fume dryer. Ethanol is heated to 70°C and your wafers are placed in the fumes of Ethanol which will make the water evaporate from the surface. | ||
The Ethanol dryer is placed on a shelf or trolly in D-3 and looks like a US bath. Describtion on how to use is written on the bath. | |||
Latest revision as of 12:49, 30 September 2024
Feedback to this page: click here
Unless anything else is stated, everything on this page, text and pictures are made by DTU Nanolab.
Drying Comparison Table
Equipment | Spin dryer 2 | Spin dryer 3 | Spin dryer 4 | Spin dryer 5 | Critical point dryer | Ethanol fume dryer | N2 guns |
---|---|---|---|---|---|---|---|
Location |
|
|
|
|
|
|
|
Purpose |
|
|
|
|
|
|
|
Batch size |
|
|
|
|
|
|
|
Allowed materials |
|
|
|
|
|
|
|
Spin dryers
A tool used for drying wafers. A single cassette of wafers is placed in a stainless steel rotor. The wafers are centered just off the axis of rotation, and rotated at high RPM so that centrifugal force “locks” the wafers in place.
Some of the tools have DI water connected, and can be used for both rinsing and drying of wafers, while other tools do not and can only be used for drying. All wafers have to be washed in a wetbench cleaning station before going into the spin dryer, regardless of using the rinse function or not.
The user manual, technical information, and contact information can be found in LabManager: Spin dryer 2 Spin dryer 3
Process information
Standard process is:
- Rinse (with DI water) for 60 seconds at 500 RPM
- Dry with nitrogen for 150 seconds at 1800 RPM
- Dry with nitrogen and anti-static purge for 30 seconds at 1800 RPM
All spin dryers use the following carriers:
- 100 mm wafers: A72-40MB
- 150 mm wafers: A182-60MB
- 200 mm wafers: A192-80M
It is not allowed to use other carriers in the spin dryers, as this may damage the motor due to incorrect mass distribution. It may also cause fragile wafers to break due to excessive vibrations caused by the mass imbalance.
Spin dryer 2
Spin dryer 2 is for RCA processing only, and is located in B-1 next to the RCA bench.
- Top chamber: 100 mm wafers
- Bottom chamber: 150 mm wafers
Spin dryer 3
Spin dryer 3 is located in D-3 between the working table and wet bench 03.
- Top chamber: 100 mm wafers
- Bottom chamber: 150 mm wafers
Spin dryer 4
Spin dryer 4 is located in D-3 between the HF vapor phase etcher and wet bench 07.
- Top chamber: 100 mm wafers
- Bottom chamber: 150 mm wafers
Spin dryer 5
Spin dryer 5 is located in F-3 Next to the Developer: TMAH stepper. It has a single process chamber that is capable of processing 100 mm, 150 mm and 200 mm wafers. The wafer size is easily changed, by changing the rotor insert frame.
Ethanol fume dryer
If your wafers are fragile or having thin membranes, cantilevers or suspended bridges that risk to stick or adhere to the surface it can be a good idea to dry your wafers in the ethanol fume dryer. Ethanol is heated to 70°C and your wafers are placed in the fumes of Ethanol which will make the water evaporate from the surface. The Ethanol dryer is placed on a shelf or trolly in D-3 and looks like a US bath. Describtion on how to use is written on the bath.
Nitrogen guns
Nitrogen guns are valuable for drying small substrate pieces or samples that otherwise cannot go into the outer drying tools.