==PhotoLuminescence Mapper: YWafer Mapper RD8-CP Photoluminescence== | |||
The PL mapper is used for non-destructive, spatially resolved optical characterization of wafers and thin-film samples. It is primarily used for photoluminescence mapping, but can also perform reflectance measurements using the integrated halogen lamp. | |||
The PL mapper is a YWafer Mapper RD8-CP photoluminescence mapping system from YSystems, Japan. It contains an enclosed motorized mapping stage, laser excitation sources at 405, 532 and 980 nm, 310-2160 nm detection range, a halogen lamp for reflectance measurements, an integrated control PC, and safety interlocks. The system is operated using the YWaferController software. For PL measurements, the system can map peak wavelength, peak amplitude, integrated intensity and FWHM. The software can also display PL data on a wavelength or energy scale. For reflectance measurements, the system can map reflectance-related parameters across the selected sample area. | |||
*System: YWafer Mapper RD8-CP Photoluminescence Mapping System. | |||
*Manufacturer: YSystems, Japan. | |||
*Full configuration: YWAFER-RD8-CP-VISNIR-QE-NIR22-405-532-980-FT-R-HAL. | |||
*Software: YWaferController for measurements, YWaferViewer for viewing maps and spectra and extracting data. | |||
The user manual and contact information can be found in [http://www.labmanager.dtu.dk/function.php?module=Machine&view=view&mach=574 '''LabManager''']. | |||
{| border="2" cellspacing="0" cellpadding="10" | |||
|- | |||
!style="background:silver; color:black" align="left" rowspan="4" valign="top" |Performance | |||
|style="background:LightGrey; color:black"|Excitation | |||
|style="background:WhiteSmoke; color:black"| | |||
*100 mW @ 405 nm | |||
*80 mW @ 532 nm | |||
*150 mW @ 980 nm | |||
*white light source for reflectance measurements (halogen lamp) | |||
|- | |||
|style="background:LightGrey; color:black"|Detection | |||
|style="background:WhiteSmoke; color:black"| | |||
*Silicon detector and IR/InGaAs detector. The system can switch between them to allow detection in the entire range. | |||
*Silicon detector range: 310-1080 nm. | |||
*IR/InGaAs detector range: 950-2160 nm. | |||
|- | |||
|style="background:LightGrey; color:black"|Sample/tray sizes | |||
|style="background:WhiteSmoke; color:black"| | |||
*2-inch, 4-inch, 2-inch quarter, 4-inch quarter, | |||
*40x40 mm, 76x26 mm (glass microscopy slide) | |||
*odd-sized samples smaller than 4-inch | |||
|- | |||
|style="background:LightGrey; color:black"|Resolution | |||
|style="background:WhiteSmoke; color:black"| | |||
*Minimum mapping resolution: 0.02 x 0.02 mm. | |||
*Spot size: Approx. 100 µm at ~1000 nm. | |||
|- | |||
!style="background:silver; color:black" align="left" rowspan="2" valign="top" |Materials | |||
|style="background:LightGrey; color:black"|Allowed substrate materials | |||
|style="background:WhiteSmoke; color:black"| | |||
* III-V | |||
* Silicon | |||
|- | |||
|style="background:LightGrey; color:black"|Forbidden materials | |||
|style="background:WhiteSmoke; color:black"| | |||
*Do not map any wafers that might leave residues in the machine | |||
|- | |||
!style="background:silver; color:black" align="left" rowspan="6" valign="top" |Software | |||
|style="background:LightGrey; color:black"|YWaferViewer | |||
|style="background:WhiteSmoke; color:black"| | |||
*.ywf map/spectra files can be viewed locally using the publicly available YWaferViewer software [https://www.ysystems.jp/en/component/edocman/public-downloads Here] | |||
|} | |||
==PhotoLuminescence Mapper RPM2000 (This tool is not being serviced by DTU Nanolab anymore)== | |||
[[Image:PL-mapper.jpg|500px|right|thumb|Positioned in the MOCVD room: F-1, {{photo1}}]] | |||
Photoluminescence mapping is a non-contact, non-destructive technique for mapping out uniformity of alloy composition, material quality and defects in substrates and of III-V epiwafers. The Rapid Photoluminiscence Mapper (RPM) is equipped with 3 lasers for PL measurements and a white-light source to map out thickness and reflectance of eg layers, microcavities and VCSELs. | |||
[[Image:Wafer-bonding PL-mapper (LabAdviser) Jehem.jpg|500px|right|thumb|Map of two bonded silicon wafers. Red areas are voids between the wafers, Made by Jens Hemmingsen @ DTU Nanolab]] | |||
It can also be used to map out voids after silicon wafer-bonding. This is done using the reflectance mapping and is using the fact that silicon is transparent for wavelengths above ~1000nm. A void will therefor change the reflectance in that wavelength range. See datasheet below (Thanks to Jens Hemmingsen for the data). | |||
The user manual and contact information can be found in [http://www.labmanager.dtu.dk/function.php?module=Machine&view=view&mach=152 '''LabManager''']. | |||
{| border="2" cellspacing="0" cellpadding="10" | {| border="2" cellspacing="0" cellpadding="10" | ||
| Line 61: | Line 121: | ||
|style="background:LightGrey; color:black"|RPM viewer | |style="background:LightGrey; color:black"|RPM viewer | ||
|style="background:WhiteSmoke; color:black"| | |style="background:WhiteSmoke; color:black"| | ||
* | *Only on the local PC at the system | ||
|} | |} | ||
Latest revision as of 10:05, 7 September 2026
The content on this page, including all images and pictures, was created by DTU Nanolab staff, unless otherwise stated.
Feedback to this page: click here
PhotoLuminescence Mapper: YWafer Mapper RD8-CP Photoluminescence
The PL mapper is used for non-destructive, spatially resolved optical characterization of wafers and thin-film samples. It is primarily used for photoluminescence mapping, but can also perform reflectance measurements using the integrated halogen lamp.
The PL mapper is a YWafer Mapper RD8-CP photoluminescence mapping system from YSystems, Japan. It contains an enclosed motorized mapping stage, laser excitation sources at 405, 532 and 980 nm, 310-2160 nm detection range, a halogen lamp for reflectance measurements, an integrated control PC, and safety interlocks. The system is operated using the YWaferController software. For PL measurements, the system can map peak wavelength, peak amplitude, integrated intensity and FWHM. The software can also display PL data on a wavelength or energy scale. For reflectance measurements, the system can map reflectance-related parameters across the selected sample area.
- System: YWafer Mapper RD8-CP Photoluminescence Mapping System.
- Manufacturer: YSystems, Japan.
- Full configuration: YWAFER-RD8-CP-VISNIR-QE-NIR22-405-532-980-FT-R-HAL.
- Software: YWaferController for measurements, YWaferViewer for viewing maps and spectra and extracting data.
The user manual and contact information can be found in LabManager.
| Performance | Excitation |
|
|---|---|---|
| Detection |
| |
| Sample/tray sizes |
| |
| Resolution |
| |
| Materials | Allowed substrate materials |
|
| Forbidden materials |
| |
| Software | YWaferViewer |
|
PhotoLuminescence Mapper RPM2000 (This tool is not being serviced by DTU Nanolab anymore)

Photoluminescence mapping is a non-contact, non-destructive technique for mapping out uniformity of alloy composition, material quality and defects in substrates and of III-V epiwafers. The Rapid Photoluminiscence Mapper (RPM) is equipped with 3 lasers for PL measurements and a white-light source to map out thickness and reflectance of eg layers, microcavities and VCSELs.

It can also be used to map out voids after silicon wafer-bonding. This is done using the reflectance mapping and is using the fact that silicon is transparent for wavelengths above ~1000nm. A void will therefor change the reflectance in that wavelength range. See datasheet below (Thanks to Jens Hemmingsen for the data).
The user manual and contact information can be found in LabManager.
| Performance | Excitation |
|
|---|---|---|
| Detection |
| |
| Gratings |
| |
| Chuck sizes |
| |
| Resolution |
| |
| Wavelength accuracy |
| |
| Materials | Allowed substrate materials |
|
| Forbidden materials |
| |
| Software | RPM viewer |
|