Specific Process Knowledge/Lithography/EBeamLithography/JEOLAlignment: Difference between revisions

Thope (talk | contribs)
Thope (talk | contribs)
 
(4 intermediate revisions by the same user not shown)
Line 82: Line 82:
{| style="border: none; border-spacing: 0; margin: 1em auto; text-align: center;"
{| style="border: none; border-spacing: 0; margin: 1em auto; text-align: center;"
|-
|-
| [[image:OpticalPrealigner1.png|400px]] || [[image:OpticalPrealigner2.png|400px]]
| [[image:OpticalPrealigner1.png|300px]] || [[image:OpticalPrealigner2.png|300px]]
|-  
|-  
| colspan="2" style="text-align:center;|
| colspan="2" style="text-align:center;|
Line 192: Line 192:
{| style="border: none; border-spacing: 0; margin: 1em auto; text-align: center;"
{| style="border: none; border-spacing: 0; margin: 1em auto; text-align: center;"
|-
|-
| [[image:P_roughscan.png|400px]] || [[image:Q_finescan.png|400px]]  || [[image:PQ_finaloffset.png|400px]]  
| [[image:P_roughscan.png|300px]] || [[image:Q_finescan.png|300px]]  || [[image:PQ_finaloffset.png|300px]]  
|-  
|-  
| colspan="2" style="text-align:center;|
| colspan="2" style="text-align:center;|
Line 203: Line 203:
{| style="border: none; border-spacing: 0; margin: 1em auto; text-align: center;"
{| style="border: none; border-spacing: 0; margin: 1em auto; text-align: center;"
|-
|-
| [[image:CHIPAL2.png|400px]] || [[image:CHIPAL-scan.png|400px]]
| [[image:CHIPAL2.png|300px]] || [[image:CHIPAL-scan.png|300px]]
|-  
|-  
| colspan="2" style="text-align:center;|
| colspan="2" style="text-align:center;|
Line 289: Line 289:
{| style="border: none; border-spacing: 0; margin: 1em auto; text-align: center;"
{| style="border: none; border-spacing: 0; margin: 1em auto; text-align: center;"
|-
|-
| [[image:scantype.png|400px]] || [[image:2Dpane.png|400px]]
| [[image:scantype.png|300px]] || [[image:2Dpane.png|300px]]
|-  
|-  
| colspan="2" style="text-align:enter;|
| colspan="2" style="text-align:enter;|
Line 314: Line 314:
{| style="border: none; border-spacing: 0; margin: 1em auto; text-align: center;"
{| style="border: none; border-spacing: 0; margin: 1em auto; text-align: center;"
|-
|-
| [[image:2Droughscan.png|400px]] || [[image:2Dfinescan.png|400px]]
| [[image:2Droughscan.png|300px]] || [[image:2Dfinescan.png|300px]]
|-  
|-  
| colspan="2" style="text-align:enter;|
| colspan="2" style="text-align:enter;|
Imagescans obtained during alignment. Rough alignment scan (left) is 1000 µm while fine scan (right) is 10 µm. The green cross indicates the center as determined by the system.
Imagescans obtained during alignment. Rough alignment scan (left) is 1000 µm while fine scan (right) is 10 µm. The green cross indicates the center as determined by the system.
|}
|}