This is a standard positive resist from ZEON. The resist is very expensive; DTU Nanolab has thus stopped buying this resist. | |||
Use CSAR instead. | |||
== Spin Curves == | == Spin Curves == | ||
| Line 115: | Line 10: | ||
The thickness is measured on VASE Ellipsometer using a simple Cauchy model for a transparent polymer on a Si wafer with native oxide. The measurements are performed at one incidence angle (70 degrees) only. 9 points on each 4" wafer has been measured; the standard deviation thus | The thickness is measured on VASE Ellipsometer using a simple Cauchy model for a transparent polymer on a Si wafer with native oxide. The measurements are performed at one incidence angle (70 degrees) only. 9 points on each 4" wafer has been measured; the standard deviation thus represents the homogeinity of the film on the 4" wafers. | ||
{|border="1" cellspacing="0" cellpadding="3" style="text-align:left;" style="width: 40%" | {|border="1" cellspacing="0" cellpadding="3" style="text-align:left;" style="width: 40%" | ||
| Line 122: | Line 17: | ||
|- | |- | ||
|-style="background:Black; color:White" | |-style="background:Black; color:White" | ||
!colspan="7"|ZEON ZEP520A1:1 in anisole spinning on Spin Coater: Manual LabSpin A-5, TIGRE, 23-04-2014. Resist was mixed 13-01-2014, i.e. | !colspan="7"|ZEON ZEP520A1:1 in anisole spinning on Spin Coater: Manual LabSpin A-5, TIGRE, 23-04-2014. Resist was mixed 13-01-2014, i.e. approximately 3 months old. | ||
|- | |- | ||
| Line 167: | Line 62: | ||
|- | |- | ||
|-style="background:red; color:White" | |-style="background:red; color:White" | ||
!colspan="7"|ZEON ZEP520A1:1 in anisole spinning on Spin Coater: Manual LabSpin A-5, TIGRE, 21-05-2014. Resist was mixed 15-05-2014, i.e. | !colspan="7"|ZEON ZEP520A1:1 in anisole spinning on Spin Coater: Manual LabSpin A-5, TIGRE, 21-05-2014. Resist was mixed 15-05-2014, i.e. approximately 6 days old. | ||
|- | |- | ||
| Line 227: | Line 122: | ||
|- | |- | ||
|-style="background:green; color:White" | |-style="background:green; color:White" | ||
!colspan="7"|ZEON ZEP520A1:2 in anisole spinning on Spin Coater: Manual LabSpin A-5, TIGRE, 21-05-2014. Resist was mixed 15-05-2014, i.e. | !colspan="7"|ZEON ZEP520A1:2 in anisole spinning on Spin Coater: Manual LabSpin A-5, TIGRE, 21-05-2014. Resist was mixed 15-05-2014, i.e. approximately 6 days old. | ||
|- | |- | ||