[[Image:eLINE-Plus.png| | =Raith eLINE Plus system= | ||
[[Image:eLINE-Plus.png|300px|right|]] | |||
The system was installed in the cleanroom in May 2022. As we get more familiar with the tool this page will be populated with relevant process information. | The system was installed in the cleanroom in May 2022. As we get more familiar with the tool this page will be populated with relevant process information. | ||
<br> | |||
<br> | |||
<br> | |||
== System overview == | == System overview == | ||
[[File:RaithHolders.jpg|600px|thumb|left|4 inch wafer holder and 100 mm Universal Sample Holder (USH)]] | |||
<br clear=all> | |||
[[File:RaithHolders.jpg| | |||
The system is a dual use SEM and EBL exposure tool. For SEM applications the most notable difference from our other SEM's is the automation functionality that allows users to link a design to a substrate and simply from the design define areas to image and the tool will then image those areas without further user input. Once an imaging routine has been setup several hundred images can be acquired per hour. For a brief introduction to this feature see more [https://youtu.be/YoZF_6FeVb4 in this video.] | The system is a dual use SEM and EBL exposure tool. For SEM applications the most notable difference from our other SEM's is the automation functionality that allows users to link a design to a substrate and simply from the design define areas to image and the tool will then image those areas without further user input. Once an imaging routine has been setup several hundred images can be acquired per hour. For a brief introduction to this feature see more [https://youtu.be/YoZF_6FeVb4 in this video.] | ||
| Line 20: | Line 19: | ||
==Dose information== | ==Dose information== | ||
[[Image:eLINE-Areas.png|800x800px|right|thumb|Base doses for exposure must be defined in the Pattern Parameter Calculation window.]] | [[Image:eLINE-Areas.png|800x800px|right|thumb|Base doses for exposure must be defined in the Pattern Parameter Calculation window.]] | ||
| Line 49: | Line 50: | ||
[[Image:DoseChip6.png|300x300px|right|thumb|Overview image of the dose test design provided on the tool.]] | [[Image:DoseChip6.png|300x300px|right|thumb|Overview image of the dose test design provided on the tool.]] | ||
===Table of dose to clear (area or curved elements) on Si substrate=== | ===Table of dose to clear (area or curved elements) on Si substrate=== | ||
The following table is merely at guideline to as to what sort of dose requirements to expect. Actual dose for your design will depend on substrate type and material, resist thickness, acceleration voltage, pattern density (proximity effects) and several other factors and thus users should always do a dose test with their actual pattern. Be aware that AR-N 8200 is in particular very dependent on PEB processing parameters. Refer to [[Specific_Process_Knowledge/Lithography/ARN8200|AR-N 8200 page]] for more information. | The following table is merely at guideline to as to what sort of dose requirements to expect. Actual dose for your design will depend on substrate type and material, resist thickness, acceleration voltage, pattern density (proximity effects) and several other factors and thus users should always do a dose test with their actual pattern. Be aware that AR-N 8200 is in particular very dependent on PEB processing parameters. Refer to [[Specific_Process_Knowledge/Lithography/ARN8200|AR-N 8200 page]] for more information. | ||
| Line 79: | Line 82: | ||
==Typical beam currents== | ==Typical beam currents== | ||
Beam current is a function of acceleration voltage and aperture and thus beam current is locked by the choice of acceleration voltage and aperture. Typical beam currents are given in the table below. The High Current (HC) mode can be activated in the Column Control panel. | |||
The High Current (HC) mode can be activated in the Column Control panel. | |||
| Line 157: | Line 154: | ||
|} | |} | ||
== | |||
Writefield dimension is a trade off between beam shot precision and field stitching. The maximum writefield size is 1000x1000 µm. The beam controller has a limit of | Writing time can be estimated in the software once a job is defined, it is however straight forward to estimate in advance based on area to pattern, A, dose to clear, d0, and beam current, I as: | ||
T = d0*A/I | |||
Notice that this estimate is only the beam dwell time and does not account for beam settling time or stage movement time. Typically process times are 3-4 times higher than the dwell time estimate but this will vary with pattern and writing conditions. | |||
==Writefields== | |||
[[Image:WF_imagescan.png|500x500px|right|thumb|WF alignment by image scan.]] | |||
[[Image:WF_linescan.png|500x500px|right|thumb|WF alignment by line scan.]] | |||
Writefield (WF) dimension is a trade off between beam shot precision and field stitching. The maximum writefield size is 1000x1000 µm. The beam controller has a limit of 50k addressable positions along each axis and hence for a 1000x1000 µm writefield the minimum beam position grid (pitch) is 20 nm. For a 100x100 µm writefield the minimum beam pitch is 2 nm. Thus the precision is higher for smaller writing fields. Smaller writing fields will however fracture a design into more fields and create more field boundaries with higher potential for stitching errors. | |||
To minimize stitching errors it is important to perform WF alignment before writing. WF alignment should be done on either a particle or an existing feature close to the writing area. WF alignment can be done in three different ways: | |||
:1. Manual alignment by image scan to a particle or feature | |||
:2. Automatic alignment by image scan of a particle or feature | |||
:3. Automatic alignment by line scan of feature (preferably a cross) | |||
The concept is illustrated in the two drawings and relies on the high precision of the laser interferometer driven stage (about 1 nm precision). When using image scan for WF alignment the user must center the chosen feature or particle in the image field. When executing the WF alignment the stage will displace a bit less than half a WF in one direction. The beam will then deflect back by the same amount and make an image scan. If the WF is perfectly aligned the deflected image will be show the feature in the center of the image. In manual alignment mode the user will indicate any offset, in automatic mode the system will do image analysis and determine the offset. For WF alignment 3-8 different stage positions will be scanned in this way and at the end a correction is calculated and applied. | |||
WF alignment can also be done by line scanning, preferably over a cross from a previous lithography step. The procedure is the same as above except that at each position instead of making an image scan the system makes two line scans and detects each leg of the cross to calculate the center of the cross and in turn the offset of the cross. Again this is done for 3-8 different positions. | |||