Specific Process Knowledge/Lithography/UVExposure: Difference between revisions

From LabAdviser
Jehem (talk | contribs)
No edit summary
Jehem (talk | contribs)
 
(54 intermediate revisions by 2 users not shown)
Line 1: Line 1:
{{:Specific Process Knowledge/Lithography/authors_generic}}
{{cc-nanolab}}


'''Feedback to this page''': '''[mailto:labadviser@nanolab.dtu.dk?Subject=Feed%20back%20from%20page%20http://labadviser.nanolab.dtu.dk/index.php?title=Specific_Process_Knowledge/Lithography/UVExposure click here]'''
'''Feedback to this page''': '''[mailto:labadviser@nanolab.dtu.dk?Subject=Feed%20back%20from%20page%20http://labadviser.nanolab.dtu.dk/index.php?title=Specific_Process_Knowledge/Lithography/UVExposure click here]'''
Line 11: Line 11:




{| border="2" cellspacing="0" cellpadding="2"  
{|border="1" cellspacing="1" cellpadding="10" style="text-align:left;"  


!colspan="2" border="none" style="background:silver; color:black;" align="center"|Equipment  
!colspan="2" border="none" style="background:silver; color:black;" align="center"|Equipment  
|style="background:WhiteSmoke; color:black"|<b>[[Specific_Process_Knowledge/Lithography/UVExposure#KS Aligner|KS Aligner]]</b>
|style="background:WhiteSmoke; color:black"|<b>[[Specific_Process_Knowledge/Lithography/UVExposure#Aligner: MA6-1|Aligner: MA6-1]]</b>
|style="background:WhiteSmoke; color:black"|<b>[[Specific_Process_Knowledge/Lithography/UVExposure#Aligner:_MA6_-_2|Aligner: MA6 - 2]]</b>
|style="background:WhiteSmoke; color:black"|<b>[[Specific_Process_Knowledge/Lithography/UVExposure#Aligner:_MA6-2|Aligner: MA6-2]]</b>
|style="background:WhiteSmoke; color:black"|<b>[[Specific_Process_Knowledge/Lithography/UVExposure#Aligner: Maskless 01|Aligner: Maskless 01]]</b>
|style="background:WhiteSmoke; color:black"|<b>[[Specific_Process_Knowledge/Lithography/UVExposure#Aligner: Maskless 01|Aligner: Maskless 01]]</b>
|style="background:WhiteSmoke; color:black"|<b>[[Specific_Process_Knowledge/Lithography/UVExposure#Aligner: Maskless 02|Aligner: Maskless 02]]</b>
|style="background:WhiteSmoke; color:black"|<b>[[Specific_Process_Knowledge/Lithography/UVExposure#Aligner: Maskless 02|Aligner: Maskless 02]]</b>
|style="background:WhiteSmoke; color:black"|<b>[[Specific_Process_Knowledge/Lithography/UVExposure#Aligner: Maskless 03|Aligner: Maskless 03]]</b>
|style="background:WhiteSmoke; color:black"|<b>[[Specific_Process_Knowledge/Lithography/UVExposure#Aligner: Maskless 03|Aligner: Maskless 03]]</b>
|style="background:WhiteSmoke; color:black"|<b>[[Specific_Process_Knowledge/Lithography/UVExposure#Inclined UV Lamp|Inclined UV Lamp]]</b>
<!--|style="background:WhiteSmoke; color:black"|<b>[[Specific_Process_Knowledge/Lithography/UVExposure#Inclined UV Lamp|Inclined UV Lamp]]</b>-->


|-
|-
Line 45: Line 45:
*Back Side Alignment
*Back Side Alignment
*Maskless UV exposure
*Maskless UV exposure
|style="background:WhiteSmoke; color:black"|
<!--|style="background:WhiteSmoke; color:black"|
*UV exposure
*UV exposure
*DUV exposure
*DUV exposure-->


|-
|-
Line 53: Line 53:
|style="background:LightGrey; color:black"|Minimum feature size
|style="background:LightGrey; color:black"|Minimum feature size
|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
*1.5µm down to 1µm
~1 µm
|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
*1.5µm down to 0.5µm
~1 µm
|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
*2µm down to 1µm
~1 µm
|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
*1µm down to 0.6µm
~1 µm
|style="background:WhiteSmoke; color:black"|
*2µm down to 1µm
|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
~1 µm
<!--|style="background:WhiteSmoke; color:black"|-->
|-
|-
|style="background:LightGrey; color:black"|Exposure light/filters/spectrum
|style="background:LightGrey; color:black"|Exposure light/filters/spectrum
|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
*350W Hg lamp
*350W Hg lamp
*i-line filter (365nm notch filter), intensity in Constant Intensity mode: 8mW/cm<sup>2</sup> @ 365nm
*i-line filter (365nm notch filter)<br>Intensity in Constant Intensity mode:<br>8mW/cm<sup>2</sup> @ 365nm
*303nm filter optional
*303nm filter optional
|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
*500W Hg-Xe lamp
*500W Hg-Xe lamp
*i-line filter (365nm notch filter), intensity in Constant Intensity mode: 11mW/cm<sup>2</sup> @ 365nm
*i-line filter (365nm notch filter)<br>Intensity in Constant Intensity mode:<br>11mW/cm<sup>2</sup> @ 365nm
*UV350 optics optional
*UV350 optics optional
*UV250 optics optional
*UV250 optics optional
|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
*365nm LED
365nm LED
|style="background:WhiteSmoke; color:black"|
*375nm laserdiodes
or
*405nm laserdiodes
|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
*405nm laserdiodes
375nm laserdiodes
|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
405nm laserdiodes
<!--|style="background:WhiteSmoke; color:black"|
*1000 W Hg-Xe lamp
*1000 W Hg-Xe lamp
*near UV (350-450nm), mid UV (260-320nm), and deep UV (220-260nm)
*Dichroic mirror:
**Near UV (350-450nm)
**Mid UV (260-320nm)
**Deep UV (220-260nm)-->


|-
|-
|style="background:LightGrey; color:black"|Exposure mode
|style="background:LightGrey; color:black"|Exposure mode
|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
*Flood exposure
*Proximity
*Proximity
*Contact
*Contact:
**Soft contact
**Soft contact
**Hard contact
**Hard contact
**Vacuum contact
**Vacuum contact
|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
*Flood exposure
*Proximity
*Proximity
*Contact
*Contact:
**Soft contact
**Soft contact
**Hard contact
**Hard contact
**Vacuum contact
**Vacuum contact
|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
*Projection  
*Projection:
**Pneumatic auto-focus
**Pneumatic auto-focus
|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
*Projection  
*Projection:
**Optical auto-focus
**Optical auto-focus
**Pneumatic auto-focus
**Pneumatic auto-focus
|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
*Projection  
*Projection:
**Pneumatic auto-focus
**Pneumatic auto-focus
|style="background:WhiteSmoke; color:black"|
<!--|style="background:WhiteSmoke; color:black"|
*Flood exposure
*Flood exposure
*Proximity exposure with home-made chuck and maskholder
*Proximity exposure with home-made chuck and maskholder
*Inclined exposure
*Inclined exposure
*Rotating exposure
*Rotating exposure-->


|-
|-
Line 145: Line 149:
*1 100 mm wafer
*1 100 mm wafer
*1 150 mm wafer
*1 150 mm wafer
|style="background:WhiteSmoke; color:black"|
<!--|style="background:WhiteSmoke; color:black"|
*all sizes up to 8inch
*all sizes up to 8inch-->
 
|-
|-
| style="background:LightGrey; color:black"|Allowed materials
| style="background:LightGrey; color:black"|Allowed materials
Line 161: Line 166:
|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
*All cleanroom materials
*All cleanroom materials
|style="background:WhiteSmoke; color:black"|
<!--|style="background:WhiteSmoke; color:black"|
*All cleanroom materials
*All cleanroom materials-->
|-  
|-  
|}
|}


 
==Aligner: MA6-1==
==KS Aligner==
[[Image:KSAligner in E-4.jpg|300x300px|thumb|The Aligner: MA6-1 is located in PolyFabLab]]
[[Image:KSAligner in E-4.jpg|300x300px|thumb|The KS Aligner is placed in E-4]]
 
'''Feedback to this section''': '''[mailto:labadviser@nanolab.dtu.dk?Subject=Feed%20back%20from%20page%20http://labadviser.nanolab.dtu.dk/index.php/Specific_Process_Knowledge/Lithography/UVExposure#KS_Aligner click here]'''


SUSS Mask Aligner MA6 is designed for high resolution photolithography.  
SUSS Mask Aligner MA6 is designed for high resolution photolithography.  
Line 176: Line 178:
Two alignment options are available: top side alignment (TSA) with a split field or a video microscope and back side alignment (BSA) with BSA microscope. It is also possible to make IR- light alignment.
Two alignment options are available: top side alignment (TSA) with a split field or a video microscope and back side alignment (BSA) with BSA microscope. It is also possible to make IR- light alignment.


'''Training videos: [https://www.youtube.com/watch?v=o8IBtfQHNzU Operation] [https://www.youtube.com/watch?v=rvUuXYgw-xU Alignment]''' (on Aligner: MA6-2, but KS Aligner is the same model)


'''The user manual(s), quality control procedure(s) and results and contact information can be found in [http://labmanager.dtu.dk/function.php?module=Machine&view=view&mach=44 LabManager].'''
'''Training videos:'''
 
[https://www.youtube.com/watch?v=o8IBtfQHNzU Operation]
 
[https://www.youtube.com/watch?v=rvUuXYgw-xU Alignment]
 
 
The user manual(s), quality control procedure(s) and results and contact information can be found in [http://labmanager.dtu.dk/function.php?module=Machine&view=view&mach=44 LabManager] - '''requires login'''


===Exposure dose===
===Exposure dose===
[[Specific Process Knowledge/Lithography/Resist#KS_Aligner|Information on UV exposure dose]]
[[Specific Process Knowledge/Lithography/Resist#Aligner: MA6-1|Information on UV exposure dose]]


===Process information===
===Process information===
The KS Aligner has an i-line notch filter installed. This results in an exposure light peak around 365nm with a FWHM of 7nm.
The Aligner: MA6-1 has an i-line notch filter installed. This results in an exposure light peak around 365nm with a FWHM of 7nm.


===Quality Control (QC)===
===Quality Control (QC)===
Line 191: Line 199:
|-
|-
|
|
*[http://labmanager.dtu.dk/d4Show.php?id=5644&mach=44 The QC procedure for KS Aligner]<br>
*[http://labmanager.dtu.dk/d4Show.php?id=5644&mach=44 The QC procedure for Aligner: MA6-1] - '''requires login'''<br>
*[https://labmanager.dtu.dk/view_binary.php?type=data&mach=44 The newest QC data for KS Aligner]
*[https://labmanager.dtu.dk/view_binary.php?type=data&mach=44 The newest QC data for Aligner: MA6-1] - '''requires login'''
{| {{table}}
{| {{table}}
| align="center" |  
| align="center" |  
Line 210: Line 218:
{| border="2" cellspacing="1" cellpadding="2" align="center" style="width:400px"
{| border="2" cellspacing="1" cellpadding="2" align="center" style="width:400px"
!QC limits
!QC limits
!KS Aligner
!Aligner: MA6-1
|-
|-
|Nominal intensity
|Nominal intensity
Line 229: Line 237:
===Equipment performance and process related parameters===
===Equipment performance and process related parameters===


{| border="2" cellspacing="0" cellpadding="2"  
{|border="1" cellspacing="1" cellpadding="10" style="text-align:left;"  




!style="background:silver; color:black;" align="center" width="60"|Purpose  
!style="background:silver; color:black;" align="center" width="60"|Purpose  
|style="background:LightGrey; color:black"|
|style="background:LightGrey; color:black"|
|style="background:WhiteSmoke; color:black" align="center" colspan="2"|
|style="background:WhiteSmoke; color:black" colspan="2"|
Alignment and UV exposure
Alignment and UV exposure
|-
|-
Line 241: Line 249:


|style="background:LightGrey; color:black"|Exposure mode
|style="background:LightGrey; color:black"|Exposure mode
|style="background:WhiteSmoke; color:black" align="center" colspan="2"|
|style="background:WhiteSmoke; color:black" colspan="2"|
vacuum contact, hard contact, soft contact, proximity, flood exposure
vacuum contact, hard contact, soft contact, proximity, flood exposure
|-
|-
| style="background:LightGrey; color:black"|Exposure light/filters
| style="background:LightGrey; color:black"|Exposure light/filters
|style="background:WhiteSmoke; color:black" align="center" colspan="2"|
|style="background:WhiteSmoke; color:black" colspan="2"|
*broadband (i-, g-, h-line)
*broadband (i-, g-, h-line)
*365 nm (i-line)
*365 nm (i-line)
Line 251: Line 259:
|-
|-
|style="background:LightGrey; color:black"|Minimum structure size
|style="background:LightGrey; color:black"|Minimum structure size
|style="background:WhiteSmoke; color:black" align="center" colspan="2"|
|style="background:WhiteSmoke; color:black" colspan="2"|
down to 1.25µm
down to 1.25µm
|-
|-
|style="background:LightGrey; color:black"|Mask size
|style="background:LightGrey; color:black"|Mask size
|style="background:WhiteSmoke; color:black" align="center" colspan="2"|
|style="background:WhiteSmoke; color:black" colspan="2"|
*5x5 inch
*5x5 inch
*7x7 inch
*7x7 inch
Line 261: Line 269:
|-
|-
|style="background:LightGrey; color:black"|Alignment modes
|style="background:LightGrey; color:black"|Alignment modes
|style="background:WhiteSmoke; color:black" align="center" colspan="2"|
|style="background:WhiteSmoke; color:black" colspan="2"|
*Top side (TSA)
*Top side (TSA)
*Backside (BSA)
*Backside (BSA)
Line 267: Line 275:
!style="background:silver; color:black" align="center" valign="center" rowspan="3"|Substrates
!style="background:silver; color:black" align="center" valign="center" rowspan="3"|Substrates
|style="background:LightGrey; color:black"|Substrate size
|style="background:LightGrey; color:black"|Substrate size
|style="background:WhiteSmoke; color:black" align="center" colspan="2"|
|style="background:WhiteSmoke; color:black" colspan="2"|
* small pieces 1x1cm
* small pieces 1x1cm
* 50 mm wafers
* 50 mm wafers
Line 274: Line 282:
|-
|-
| style="background:LightGrey; color:black"|Allowed materials
| style="background:LightGrey; color:black"|Allowed materials
|style="background:WhiteSmoke; color:black" align="center" colspan="2"|
|style="background:WhiteSmoke; color:black" colspan="2"|
All cleanroom materials
All cleanroom materials


Line 280: Line 288:
|-
|-
|style="background:LightGrey; color:black"|Batch
|style="background:LightGrey; color:black"|Batch
|style="background:WhiteSmoke; color:black" align="center" colspan="2"|
|style="background:WhiteSmoke; color:black" colspan="2"|
1  
1  
|-  
|-  
Line 287: Line 295:
<br clear="all" />
<br clear="all" />


==Aligner: MA6 - 2==
==Aligner: MA6-2==
 
[[Image:AlignerMA6-2 in E-4.jpg|300x300px|thumb|The Aligner: MA6-2 is placed in E-4]]
[[Image:AlignerMA6-2 in E-4.jpg|300x300px|thumb|The Aligner: MA6 - 2 is placed in E-4]]
 
'''Feedback to this section''': '''[mailto:labadviser@nanolab.dtu.dk?Subject=Feed%20back%20from%20page%20http://labadviser.nanolab.dtu.dk/index.php/Specific_Process_Knowledge/Lithography/UVExposure#Aligner:_MA6_-_2 click here]'''


The Süss MicroTek Mask Aligner MA6 is designed for high resolution photolithography.  
The Süss MicroTek Mask Aligner MA6 is designed for high resolution photolithography.  
Line 299: Line 304:
The aligner can also be used for bond alignment (for Wafer Bonder 02). Special training is required.
The aligner can also be used for bond alignment (for Wafer Bonder 02). Special training is required.


'''Training videos: [https://www.youtube.com/watch?v=o8IBtfQHNzU Operation] [https://www.youtube.com/watch?v=rvUuXYgw-xU Alignment]'''


'''The user manual, quality control procedures and results, user APVs, and contact information can be found in [http://labmanager.dtu.dk/function.php?module=Machine&view=view&mach=339 LabManager].'''
'''Training videos:'''
 
[https://www.youtube.com/watch?v=o8IBtfQHNzU Operation]
 
[https://www.youtube.com/watch?v=rvUuXYgw-xU Alignment]
 
 
The user manual, quality control procedures and results, user APVs, and contact information can be found in [http://labmanager.dtu.dk/function.php?module=Machine&view=view&mach=339 LabManager] - '''requires login'''


===Exposure dose===
===Exposure dose===
[[Specific Process Knowledge/Lithography/Resist#Aligner:_MA6_-_2|Information on UV exposure dose]]
[[Specific Process Knowledge/Lithography/Resist#Aligner:_MA6-2|Information on UV exposure dose]]


===Process information===
===Process information===
The Aligner: MA6 - 2 has an i-line notch filter installed. This results in an exposure light peak at 365nm. Compared to exposure on the KS Aligner, the optimal dose should be very similar. The 500W Hg-Xe lamp also enables exposure in the DUV range around 240nm. This functionality is not established yet, partly due to safety concerns.
The Aligner: MA6-2 has an i-line notch filter installed. This results in an exposure light peak at 365nm. Compared to exposure on the KS Aligner, the optimal dose should be very similar. The 500W Hg-Xe lamp also enables exposure in the DUV range around 240nm. This functionality is not established yet, partly due to safety concerns.


===Quality Control (QC)===
===Quality Control (QC)===
Line 314: Line 325:
|-
|-
|
|
*[http://labmanager.dtu.dk/d4Show.php?id=5644&mach=339 The QC procedure for Aligner: MA6-2]<br>
*[http://labmanager.dtu.dk/d4Show.php?id=5644&mach=339 The QC procedure for Aligner: MA6-2] - '''requires login'''<br>
*[https://labmanager.dtu.dk/view_binary.php?type=data&mach=339 The newest QC data for Aligner: MA6-2]
*[https://labmanager.dtu.dk/view_binary.php?type=data&mach=339 The newest QC data for Aligner: MA6-2] - '''requires login'''
{| {{table}}
{| {{table}}
| align="center" |  
| align="center" |  
Line 351: Line 362:


===Alignment===
===Alignment===
'''Top Side Alignment:'''
'''Top Side Alignment:'''
*TSA microscope standard objectives: 5X, and 10X (20X available)
*TSA microscope standard objectives: 5X, and 10X (20X available)
Line 386: Line 396:
=== Equipment performance and process related parameters ===
=== Equipment performance and process related parameters ===


{| border="2" cellspacing="0" cellpadding="2"  
{|border="1" cellspacing="1" cellpadding="10" style="text-align:left;"  


!style="background:silver; color:black;" align="center" width="60"|Purpose  
!style="background:silver; color:black;" align="center" width="60"|Purpose  
|style="background:LightGrey; color:black"|
|style="background:LightGrey; color:black"|
|style="background:WhiteSmoke; color:black" align="center" colspan="2"|
|style="background:WhiteSmoke; color:black" colspan="2"|
Mask alignment and UV exposure, potentially DUV exposure <sup>1)</sup>
Mask alignment and UV exposure, potentially DUV exposure <sup>1)</sup>


Line 399: Line 409:


|style="background:LightGrey; color:black"|Exposure mode
|style="background:LightGrey; color:black"|Exposure mode
|style="background:WhiteSmoke; color:black" align="center" colspan="2"|
|style="background:WhiteSmoke; color:black" colspan="2"|
vacuum contact, hard contact, soft contact, proximity, flood exposure
vacuum contact, hard contact, soft contact, proximity, flood exposure
|-
|-
| style="background:LightGrey; color:black"|Exposure light/filters
| style="background:LightGrey; color:black"|Exposure light/filters
|style="background:WhiteSmoke; color:black" align="center" colspan="2"|
|style="background:WhiteSmoke; color:black" colspan="2"|
*broadband (i-, g-, h-line)
*broadband (i-, g-, h-line)
*365 nm (i-line)
*365 nm (i-line)
Line 410: Line 420:
|-
|-
|style="background:LightGrey; color:black"|Minimum structure size
|style="background:LightGrey; color:black"|Minimum structure size
|style="background:WhiteSmoke; color:black" align="center" colspan="2"|
|style="background:WhiteSmoke; color:black" colspan="2"|
Typically 1.25 µm, possibly down to 0.8 µm <sup>1)</sup>
Typically 1.25 µm, possibly down to 0.8 µm <sup>1)</sup>
|-
|-
|style="background:LightGrey; color:black"|Mask size
|style="background:LightGrey; color:black"|Mask size
|style="background:WhiteSmoke; color:black" align="center" colspan="2"|
|style="background:WhiteSmoke; color:black" colspan="2"|
*5x5 inch
*5x5 inch
*7x7 inch
*7x7 inch
Line 420: Line 430:
|-
|-
|style="background:LightGrey; color:black"|Alignment modes
|style="background:LightGrey; color:black"|Alignment modes
|style="background:WhiteSmoke; color:black" align="center" colspan="2"|
|style="background:WhiteSmoke; color:black" colspan="2"|
*Top side (TSA)
*Top side (TSA)
*Backside (BSA)
*Backside (BSA)
Line 426: Line 436:
!style="background:silver; color:black" align="center" valign="center" rowspan="3"|Substrates
!style="background:silver; color:black" align="center" valign="center" rowspan="3"|Substrates
|style="background:LightGrey; color:black"|Substrate size
|style="background:LightGrey; color:black"|Substrate size
|style="background:WhiteSmoke; color:black" align="center" colspan="2"|
|style="background:WhiteSmoke; color:black" colspan="2"|
* small pieces 1x1cm
* small pieces 1x1cm
* 50 mm wafers
* 50 mm wafers
Line 433: Line 443:
|-
|-
| style="background:LightGrey; color:black"|Allowed materials
| style="background:LightGrey; color:black"|Allowed materials
|style="background:WhiteSmoke; color:black" align="center" colspan="2"|
|style="background:WhiteSmoke; color:black" colspan="2"|
All cleanroom materials except copper and steel
All cleanroom materials except copper and steel


Line 439: Line 449:
|-
|-
|style="background:LightGrey; color:black"|Batch
|style="background:LightGrey; color:black"|Batch
|style="background:WhiteSmoke; color:black" align="center" colspan="2"|
|style="background:WhiteSmoke; color:black" colspan="2"|
1  
1  
|-  
|-  
Line 445: Line 455:


<sup>1)</sup> Not available yet. Requires tool change. Would require safety investigation (and a dedicated project) to become available.
<sup>1)</sup> Not available yet. Requires tool change. Would require safety investigation (and a dedicated project) to become available.
===Light intensity and uniformity after lamp ignition===
[[image:MA6-2_light_intensity_and_uniformity_from_ignition.png|400px|right|thumb|Light intensity and uniformity after lamp ignition. CP is constant power mode, CI is constant intensity mode. The accept limits for the CI mode is ±5%, the accept limit for the non-uniformity is 2%. CP mode has no accept limit.]]
The lamp requires about 15 minutes of on-time after lamp ignition, before the light output has stabilized, and users using the constant power mode should always wait for 15 minutes after ignition, before starting processing.
If users use the constant intensity mode, this waiting time is not necessary, since the internal light sensor automatically adjusts the output to match the setpoint value.
The non-uniformity of the light, in CI mode, is unaffected by the stabilization time.


<br clear="all" />
<br clear="all" />


== Aligner: Maskless 01 ==
== Aligner: Maskless 01 ==
'''Feedback to this section''': '''[mailto:lithography@Nanolab.dtu.dk?Subject=Feed%20back%20from%20page%20http://labadviser.nanolab.dtu.dk/index.php/Specific_Process_Knowledge/Lithography/UVExposure#Aligner:_Maskless_01 click here]'''


[[Image:Heidelberg_MLA100.jpg|300x300px|thumb|Aligner: Maskless 01 positioned in E-5 (2017)]]
[[Image:Heidelberg_MLA100.jpg|300x300px|thumb|Aligner: Maskless 01 positioned in E-5 (2017)]]
Line 460: Line 478:
The system offers top side alignment with high accuracy.
The system offers top side alignment with high accuracy.


'''The user manual and contact information can be found in LabManager:'''
 
Equipment info in [http://labmanager.dtu.dk/function.php?module=Machine&view=view&mach=422 LabManager]
The user manual and contact information can be found in LabManager:
 
Equipment info in [http://labmanager.dtu.dk/function.php?module=Machine&view=view&mach=422 LabManager] - '''requires login'''


===Exposure dose and defocus===
===Exposure dose and defocus===
Line 475: Line 495:
=== Equipment performance and process related parameters ===
=== Equipment performance and process related parameters ===


{| border="2" cellspacing="0" cellpadding="2"  
{|border="1" cellspacing="1" cellpadding="10" style="text-align:left;"  
 


!style="background:silver; color:black;" align="center" width="60"|Purpose  
!style="background:silver; color:black;" align="center" width="60"|Purpose  
|style="background:LightGrey; color:black"|
|style="background:LightGrey; color:black"|
|style="background:WhiteSmoke; color:black" align="center" colspan="2"|
|style="background:WhiteSmoke; color:black" colspan="2"|
Alignment and UV exposure  
Alignment and UV exposure  
|-
|-
Line 487: Line 506:


|style="background:LightGrey; color:black"|Exposure mode
|style="background:LightGrey; color:black"|Exposure mode
|style="background:WhiteSmoke; color:black" align="center" colspan="2"|
|style="background:WhiteSmoke; color:black" colspan="2"|
Projection
Projection
|-
|-
| style="background:LightGrey; color:black"|Exposure light
| style="background:LightGrey; color:black"|Exposure light
|style="background:WhiteSmoke; color:black" align="center" colspan="2"|
|style="background:WhiteSmoke; color:black" colspan="2"|
365nm (LED), FWHM=8nm
365nm (LED), FWHM=8nm
|-
|-
|style="background:LightGrey; color:black"|Focusing method
|style="background:LightGrey; color:black"|Focusing method
|style="background:WhiteSmoke; color:black" align="center" colspan="2"|
|style="background:WhiteSmoke; color:black" colspan="2"|
Pneumatic
Pneumatic
|-
|-
|style="background:LightGrey; color:black"|Minimum structure size
|style="background:LightGrey; color:black"|Minimum structure size
|style="background:WhiteSmoke; color:black" align="center" colspan="2"|
|style="background:WhiteSmoke; color:black" colspan="2"|
down to 1µm
down to 1µm
|-
|-
|style="background:LightGrey; color:black"|Design formats
|style="background:LightGrey; color:black"|Design formats
|style="background:WhiteSmoke; color:black" align="center" colspan="2"|
|style="background:WhiteSmoke; color:black" colspan="2"|
*'''GDS-II'''
*'''GDS-II'''
*CIF
*CIF
Line 511: Line 530:
|-
|-
|style="background:LightGrey; color:black"|Alignment modes
|style="background:LightGrey; color:black"|Alignment modes
|style="background:WhiteSmoke; color:black" align="center" colspan="2"|
|style="background:WhiteSmoke; color:black" colspan="2"|
Top side only
Top side only


Line 517: Line 536:
!style="background:silver; color:black" align="center" valign="center" rowspan="3"|Substrates
!style="background:silver; color:black" align="center" valign="center" rowspan="3"|Substrates
|style="background:LightGrey; color:black"|Substrate size
|style="background:LightGrey; color:black"|Substrate size
|style="background:WhiteSmoke; color:black" align="center" colspan="2"|
|style="background:WhiteSmoke; color:black" colspan="2"|
* maximum writing area: 125x125 mm<sup>2</sup>
* maximum writing area: 125x125 mm<sup>2</sup>
* 150 mm wafer
* 150 mm wafer
Line 525: Line 544:
|-
|-
| style="background:LightGrey; color:black"|Allowed materials
| style="background:LightGrey; color:black"|Allowed materials
|style="background:WhiteSmoke; color:black" align="center" colspan="2"|
|style="background:WhiteSmoke; color:black" colspan="2"|
All cleanroom materials
All cleanroom materials
|-
|-
|style="background:LightGrey; color:black"|Batch
|style="background:LightGrey; color:black"|Batch
|style="background:WhiteSmoke; color:black" align="center" colspan="2"|
|style="background:WhiteSmoke; color:black" colspan="2"|
1  
1  
|-  
|-  
Line 537: Line 556:


==Aligner: Maskless 02==
==Aligner: Maskless 02==
MLA150 WMI maskless aligner from Heidelberg Instruments GmbH.
MLA150 WMII maskless aligner from Heidelberg Instruments GmbH (installed 2019 as WMI (0.6µm resolution), rebuilt to WMII 2023).
 
 
'''PC logon password'''<br>
The logon password for the PC is "mla" (without quotation marks).




'''Special features'''
'''Special features'''
*High resolution (Write Mode I, Optical Autofocus)
*Optical Autofocus
*Backside Alignment
*Backside Alignment
*Basic Gray Scale Exposure
*Basic Gray Scale Exposure
Line 551: Line 566:
*High Aspect Ratio Mode for exposure of thick resists
*High Aspect Ratio Mode for exposure of thick resists
*200 x 200 mm exposure field
*200 x 200 mm exposure field
 
*Separate conversion PC
 
'''<span style="color:red">Special restrictions</span>'''
*No pneumatic autofocus
*405 nm laser is not allowed
*Soft resists are not allowed
*Resist thickness must be ≤10 µm
*No crystal bonded chips
*Thickness variations across the substrate must be <50 µm
*Wafer bow must be <50 µm




'''[https://www.youtube.com/playlist?list=PLjWVU97LayHAiCabstMfAUeeWyQoQI_cV Training videos]'''
'''[https://www.youtube.com/playlist?list=PLjWVU97LayHAiCabstMfAUeeWyQoQI_cV Training videos]'''


Equipment info in [http://labmanager.dtu.dk/function.php?module=Machine&view=view&mach=440 LabManager]
Equipment info in [http://labmanager.dtu.dk/function.php?module=Machine&view=view&mach=440 LabManager] - '''requires login'''


===Exposure dose and defocus===
===Exposure dose and defocus===
Line 578: Line 584:
=== Equipment performance and process related parameters ===
=== Equipment performance and process related parameters ===


{| border="2" cellspacing="0" cellpadding="2"  
{|border="1" cellspacing="1" cellpadding="10" style="text-align:left;"  




!style="background:silver; color:black;" align="center" width="60"|Purpose  
!style="background:silver; color:black;" align="center" width="60"|Purpose  
|style="background:LightGrey; color:black"|
|style="background:LightGrey; color:black"|
|style="background:WhiteSmoke; color:black" align="center" colspan="2"|
|style="background:WhiteSmoke; color:black" colspan="2"|
Alignment and UV exposure  
Alignment and UV exposure  
|-
|-
Line 590: Line 596:


|style="background:LightGrey; color:black"|Exposure mode
|style="background:LightGrey; color:black"|Exposure mode
|style="background:WhiteSmoke; color:black" align="center" colspan="2"|
|style="background:WhiteSmoke; color:black" colspan="2"|
Projection
Projection
|-
|-
| style="background:LightGrey; color:black"|Exposure light
| style="background:LightGrey; color:black"|Exposure light
|style="background:WhiteSmoke; color:black" align="center" colspan="2"|
|style="background:WhiteSmoke; color:black" colspan="2"|
*375 nm
*375 nm<br>
*<s>405 nm</s> NOT ALLOWED
(laser diode arrays)
(laser diode arrays)
|-
|-
|style="background:LightGrey; color:black"|Focusing method
|style="background:LightGrey; color:black"|Focusing method
|style="background:WhiteSmoke; color:black" align="center" colspan="2"|
|style="background:WhiteSmoke; color:black" colspan="2"|
*Optical
*Optical
*<s>Pneumatic</s> DISABLED
*Pneumatic
|-
|-
|style="background:LightGrey; color:black"|Minimum structure size
|style="background:LightGrey; color:black"|Minimum structure size
|style="background:WhiteSmoke; color:black" align="center" colspan="2"|
|style="background:WhiteSmoke; color:black" colspan="2"|
down to 600nm <sup>1)</sup>
down to 1 µm
|-
|-
|style="background:LightGrey; color:black"|Design formats
|style="background:LightGrey; color:black"|Design formats
|style="background:WhiteSmoke; color:black" align="center" colspan="2"|
|style="background:WhiteSmoke; color:black" colspan="2"|
*'''GDS-II'''
*'''GDS-II'''
*CIF
*CIF
Line 617: Line 622:
|-
|-
|style="background:LightGrey; color:black"|Alignment modes
|style="background:LightGrey; color:black"|Alignment modes
|style="background:WhiteSmoke; color:black" align="center" colspan="2"|
|style="background:WhiteSmoke; color:black" colspan="2"|
*Top side alignment
*Top side alignment
*Backside alignment
*Backside alignment
Line 625: Line 630:
!style="background:silver; color:black" align="center" valign="center" rowspan="3"|Substrates
!style="background:silver; color:black" align="center" valign="center" rowspan="3"|Substrates
|style="background:LightGrey; color:black"|Substrate size
|style="background:LightGrey; color:black"|Substrate size
|style="background:WhiteSmoke; color:black" align="center" colspan="2"|
|style="background:WhiteSmoke; color:black" colspan="2"|
* maximum writing area: 200x200 mm<sup>2</sup>
* maximum writing area: 200x200 mm<sup>2</sup>
* 200 mm wafer
* 200 mm wafer
Line 634: Line 639:
|-
|-
| style="background:LightGrey; color:black"|Allowed materials
| style="background:LightGrey; color:black"|Allowed materials
|style="background:WhiteSmoke; color:black" align="center" colspan="2"|
|style="background:WhiteSmoke; color:black" colspan="2"|
All cleanroom materials
All cleanroom materials
|-
|-
|style="background:LightGrey; color:black"|Batch
|style="background:LightGrey; color:black"|Batch
|style="background:WhiteSmoke; color:black" align="center" colspan="2"|
|style="background:WhiteSmoke; color:black" colspan="2"|
1  
1  
|-  
|-  
Line 648: Line 653:


==Aligner: Maskless 03==
==Aligner: Maskless 03==
MLA150 WMII maskless aligner from Heidelberg Instruments GmbH.
MLA150 WMII maskless aligner from Heidelberg Instruments GmbH, installed 2020.
 
 
'''PC logon password'''<br>
The logon password for the PC is "mla" (without quotation marks).




Line 663: Line 664:
'''[https://www.youtube.com/playlist?list=PLjWVU97LayHAiCabstMfAUeeWyQoQI_cV Training videos]'''
'''[https://www.youtube.com/playlist?list=PLjWVU97LayHAiCabstMfAUeeWyQoQI_cV Training videos]'''


Equipment info in [http://labmanager.dtu.dk/function.php?module=Machine&view=view&mach=464 LabManager]
Equipment info in [http://labmanager.dtu.dk/function.php?module=Machine&view=view&mach=464 LabManager] - '''requires login'''


===Exposure dose and defocus===
===Exposure dose and defocus===
Line 676: Line 677:
=== Equipment performance and process related parameters ===
=== Equipment performance and process related parameters ===


{| border="2" cellspacing="0" cellpadding="2"  
{|border="1" cellspacing="1" cellpadding="10" style="text-align:left;"  




!style="background:silver; color:black;" align="center" width="60"|Purpose  
!style="background:silver; color:black;" align="center" width="60"|Purpose  
|style="background:LightGrey; color:black"|
|style="background:LightGrey; color:black"|
|style="background:WhiteSmoke; color:black" align="center" colspan="2"|
|style="background:WhiteSmoke; color:black" colspan="2"|
Alignment and UV exposure  
Alignment and UV exposure  
|-
|-
Line 688: Line 689:


|style="background:LightGrey; color:black"|Exposure mode
|style="background:LightGrey; color:black"|Exposure mode
|style="background:WhiteSmoke; color:black" align="center" colspan="2"|
|style="background:WhiteSmoke; color:black" colspan="2"|
Projection
Projection
|-
|-
| style="background:LightGrey; color:black"|Exposure light
| style="background:LightGrey; color:black"|Exposure light
|style="background:WhiteSmoke; color:black" align="center" colspan="2"|
|style="background:WhiteSmoke; color:black" colspan="2"|
405nm (laser diode array)
405nm<br>(laser diode array)
|-
|-
|style="background:LightGrey; color:black"|Focusing method
|style="background:LightGrey; color:black"|Focusing method
|style="background:WhiteSmoke; color:black" align="center" colspan="2"|
|style="background:WhiteSmoke; color:black" colspan="2"|
Pneumatic
Pneumatic
|-
|-
|style="background:LightGrey; color:black"|Minimum structure size
|style="background:LightGrey; color:black"|Minimum structure size
|style="background:WhiteSmoke; color:black" align="center" colspan="2"|
|style="background:WhiteSmoke; color:black" colspan="2"|
down to 1µm
down to 1 µm
|-
|-
|style="background:LightGrey; color:black"|Design formats
|style="background:LightGrey; color:black"|Design formats
|style="background:WhiteSmoke; color:black" align="center" colspan="2"|
|style="background:WhiteSmoke; color:black" colspan="2"|
*'''GDS-II'''
*'''GDS-II'''
*CIF
*CIF
Line 712: Line 713:
|-
|-
|style="background:LightGrey; color:black"|Alignment modes
|style="background:LightGrey; color:black"|Alignment modes
|style="background:WhiteSmoke; color:black" align="center" colspan="2"|
|style="background:WhiteSmoke; color:black" colspan="2"|
*Top side alignment
*Top side alignment
*Backside alignment
*Backside alignment
Line 720: Line 721:
!style="background:silver; color:black" align="center" valign="center" rowspan="3"|Substrates
!style="background:silver; color:black" align="center" valign="center" rowspan="3"|Substrates
|style="background:LightGrey; color:black"|Substrate size
|style="background:LightGrey; color:black"|Substrate size
|style="background:WhiteSmoke; color:black" align="center" colspan="2"|
|style="background:WhiteSmoke; color:black" colspan="2"|
* maximum writing area: 150x150 mm<sup>2</sup>
* maximum writing area: 150x150 mm<sup>2</sup>
* 150 mm wafer
* 150 mm wafer
Line 728: Line 729:
|-
|-
| style="background:LightGrey; color:black"|Allowed materials
| style="background:LightGrey; color:black"|Allowed materials
|style="background:WhiteSmoke; color:black" align="center" colspan="2"|
|style="background:WhiteSmoke; color:black" colspan="2"|
All cleanroom materials
All cleanroom materials
|-
|-
|style="background:LightGrey; color:black"|Batch
|style="background:LightGrey; color:black"|Batch
|style="background:WhiteSmoke; color:black" align="center" colspan="2"|
|style="background:WhiteSmoke; color:black" colspan="2"|
1  
1  
|-  
|-  
Line 739: Line 740:
<br clear="all" />
<br clear="all" />


<!--
==Inclined UV Lamp==
==Inclined UV Lamp==
[[Image:Inclined UV lamp_1.jpg|300x300px|right|thumb|Inclined UV lamp is placed in CX-1]]
[[Image:Inclined UV lamp_1.jpg|300x300px|right|thumb|Inclined UV lamp is placed in CX-1]]
Line 748: Line 750:
The technical specification and the general outline of the equipment can be found in LabManager.   
The technical specification and the general outline of the equipment can be found in LabManager.   


'''The user manual(s), quality control procedure(s) and results and contact information can be found in [http://labmanager.dtu.dk/function.php?module=Machine&view=view&mach=259 LabManager].'''
 
The user manual(s), quality control procedure(s) and results and contact information can be found in [http://labmanager.dtu.dk/function.php?module=Machine&view=view&mach=259 LabManager] - '''requires login'''
<br clear="all" />
<br clear="all" />
=== Equipment performance and process related parameters ===
=== Equipment performance and process related parameters ===


{| border="2" cellspacing="0" cellpadding="2"  
{|border="1" cellspacing="1" cellpadding="10" style="text-align:left;"  




!style="background:silver; color:black;" align="center" width="60"|Purpose  
!style="background:silver; color:black;" align="center" width="60"|Purpose  
|style="background:LightGrey; color:black"|
|style="background:LightGrey; color:black"|
|style="background:WhiteSmoke; color:black" align="center" colspan="2"|
|style="background:WhiteSmoke; color:black" colspan="2"|
UV exposure  
UV exposure  
|-
|-
Line 764: Line 768:


|style="background:LightGrey; color:black"|Exposure mode
|style="background:LightGrey; color:black"|Exposure mode
|style="background:WhiteSmoke; color:black" align="center" colspan="2"|
|style="background:WhiteSmoke; color:black" colspan="2"|
*Flood exposure
*Flood exposure
*Proximity exposure with mask possible for 4inch substrate
*Proximity exposure with mask possible for 4inch substrate
Line 771: Line 775:
|-
|-
| style="background:LightGrey; color:black"|Exposure light/filters
| style="background:LightGrey; color:black"|Exposure light/filters
|style="background:WhiteSmoke; color:black" align="center" colspan="2"|
|style="background:WhiteSmoke; color:black" colspan="2"|
* Near UV(350-450nm)
* Near UV(350-450nm)
* Mid UV (260-320nm)
* Mid UV (260-320nm)
Line 778: Line 782:
|-
|-
|style="background:LightGrey; color:black"|Minimum structure size
|style="background:LightGrey; color:black"|Minimum structure size
|style="background:WhiteSmoke; color:black" align="center" colspan="2"|
|style="background:WhiteSmoke; color:black" colspan="2"|


|-
|-
|style="background:LightGrey; color:black"|Mask size
|style="background:LightGrey; color:black"|Mask size
|style="background:WhiteSmoke; color:black" align="center" colspan="2"|
|style="background:WhiteSmoke; color:black" colspan="2"|
5x5inch optinal
5x5inch optinal
|-
|-
|style="background:LightGrey; color:black"|Alignment modes
|style="background:LightGrey; color:black"|Alignment modes
|style="background:WhiteSmoke; color:black" align="center" colspan="2"|
|style="background:WhiteSmoke; color:black" colspan="2"|
No alignment possible
No alignment possible


Line 792: Line 796:
!style="background:silver; color:black" align="center" valign="center" rowspan="3"|Substrates
!style="background:silver; color:black" align="center" valign="center" rowspan="3"|Substrates
|style="background:LightGrey; color:black"|Substrate size
|style="background:LightGrey; color:black"|Substrate size
|style="background:WhiteSmoke; color:black" align="center" colspan="2"|
|style="background:WhiteSmoke; color:black" colspan="2"|
Up to 8inch substrates, different shapes  
Up to 8inch substrates, different shapes  
|-
|-
| style="background:LightGrey; color:black"|Allowed materials
| style="background:LightGrey; color:black"|Allowed materials
|style="background:WhiteSmoke; color:black" align="center" colspan="2"|
|style="background:WhiteSmoke; color:black" colspan="2"|
All cleanroom materials
All cleanroom materials
|-
|-
|style="background:LightGrey; color:black"|Batch
|style="background:LightGrey; color:black"|Batch
|style="background:WhiteSmoke; color:black" align="center" colspan="2"|
|style="background:WhiteSmoke; color:black" colspan="2"|
1  
1  
|-  
|-  
Line 806: Line 810:


<br clear="all" />
<br clear="all" />
-->

Latest revision as of 15:53, 13 November 2024

The contents on this page, including all images and pictures, was created by DTU Nanolab staff unless otherwise stated.

Feedback to this page: click here

UV Exposure Comparison Table

Equipment Aligner: MA6-1 Aligner: MA6-2 Aligner: Maskless 01 Aligner: Maskless 02 Aligner: Maskless 03
Purpose
  • Top Side Alignment
  • Back Side Alignment
  • UV exposure
  • Top Side Alignment
  • Back Side Alignment
  • UV exposure
  • (DUV exposure)
  • Bond alignment
  • Top Side Alignment
  • Maskless UV exposure
  • Top Side Alignment
  • Back Side Alignment
  • Maskless UV exposure
  • Top Side Alignment
  • Back Side Alignment
  • Maskless UV exposure
Performance Minimum feature size

~1 µm

~1 µm

~1 µm

~1 µm

~1 µm

Exposure light/filters/spectrum
  • 350W Hg lamp
  • i-line filter (365nm notch filter)
    Intensity in Constant Intensity mode:
    8mW/cm2 @ 365nm
  • 303nm filter optional
  • 500W Hg-Xe lamp
  • i-line filter (365nm notch filter)
    Intensity in Constant Intensity mode:
    11mW/cm2 @ 365nm
  • UV350 optics optional
  • UV250 optics optional

365nm LED

375nm laserdiodes

405nm laserdiodes

Exposure mode
  • Flood exposure
  • Proximity
  • Contact:
    • Soft contact
    • Hard contact
    • Vacuum contact
  • Flood exposure
  • Proximity
  • Contact:
    • Soft contact
    • Hard contact
    • Vacuum contact
  • Projection:
    • Pneumatic auto-focus
  • Projection:
    • Optical auto-focus
    • Pneumatic auto-focus
  • Projection:
    • Pneumatic auto-focus
Substrates Batch size
  • 1 small sample, down to 10x10 mm2
  • 1 50 mm wafer
  • 1 100 mm wafer
  • 1 150 mm wafer
  • 1 small sample, down to 10x10 mm2
  • 1 50 mm wafer
  • 1 100 mm wafer
  • 1 150 mm wafer
  • 1 small sample, down to 5x5 mm2
  • 1 50 mm wafer
  • 1 100 mm wafer
  • 1 150 mm wafer
  • 1 small sample, down to 3x3 mm2
  • 1 50 mm wafer
  • 1 100 mm wafer
  • 1 150 mm wafer
  • 1 200 mm wafer
  • 1 small sample, down to 5x5 mm2
  • 1 50 mm wafer
  • 1 100 mm wafer
  • 1 150 mm wafer
Allowed materials
  • All cleanroom materials except copper and steel
  • Dedicated 2inch chuck for III-V materials
  • All cleanroom materials except copper and steel
  • Dedicated chuck for III-V materials
  • All cleanroom materials
  • All cleanroom materials
  • All cleanroom materials

Aligner: MA6-1

The Aligner: MA6-1 is located in PolyFabLab

SUSS Mask Aligner MA6 is designed for high resolution photolithography. The 365nm exposure wavelength version is capable of 1.25 (1.0) um resolution in vacuum contact. All contact exposure programs (vacuum, hard, soft, proximity) are supplied. Two alignment options are available: top side alignment (TSA) with a split field or a video microscope and back side alignment (BSA) with BSA microscope. It is also possible to make IR- light alignment.


Training videos:

Operation

Alignment


The user manual(s), quality control procedure(s) and results and contact information can be found in LabManager - requires login

Exposure dose

Information on UV exposure dose

Process information

The Aligner: MA6-1 has an i-line notch filter installed. This results in an exposure light peak around 365nm with a FWHM of 7nm.

Quality Control (QC)

Quality Control (QC) for KS Aligner
QC Recipe: Manual intensity measurement
Measurement area 4"
Number of measurements 5
QC limits Aligner: MA6-1
Nominal intensity 8 mW/cm2 @ 365 nm
Intensity deviation from nominal ≤5%
Intensity non-uniformity ≤2%

Power supply and/or lamp will be adjusted if intensity is outside the limit.

Equipment performance and process related parameters

Purpose

Alignment and UV exposure

Performance Exposure mode

vacuum contact, hard contact, soft contact, proximity, flood exposure

Exposure light/filters
  • broadband (i-, g-, h-line)
  • 365 nm (i-line)
  • 303 nm
Minimum structure size

down to 1.25µm

Mask size
  • 5x5 inch
  • 7x7 inch
  • special holder for 4 x 2" designs on 5x5 inch
Alignment modes
  • Top side (TSA)
  • Backside (BSA)
Substrates Substrate size
  • small pieces 1x1cm
  • 50 mm wafers
  • 100 mm wafers
  • 150 mm wafers
Allowed materials

All cleanroom materials

Dedicated chuck for III-V materials

Batch

1


Aligner: MA6-2

The Aligner: MA6-2 is placed in E-4

The Süss MicroTek Mask Aligner MA6 is designed for high resolution photolithography. The 365nm exposure wavelength version is capable of 1.25 (1.0) um resolution in vacuum contact. All contact exposure programs (vacuum, low vacuum, hard, soft, proximity) are supplied. Two alignment options are available: top side alignment (TSA) with a split field or a video microscope and back side alignment (BSA) with BSA (video) microscope.

The aligner can also be used for bond alignment (for Wafer Bonder 02). Special training is required.


Training videos:

Operation

Alignment


The user manual, quality control procedures and results, user APVs, and contact information can be found in LabManager - requires login

Exposure dose

Information on UV exposure dose

Process information

The Aligner: MA6-2 has an i-line notch filter installed. This results in an exposure light peak at 365nm. Compared to exposure on the KS Aligner, the optimal dose should be very similar. The 500W Hg-Xe lamp also enables exposure in the DUV range around 240nm. This functionality is not established yet, partly due to safety concerns.

Quality Control (QC)

Quality Control (QC) for Aligner: MA6-2
QC Recipe: Manual intensity measurement
Measurement area 4"
Number of measurements 5
QC limits Aligner: MA6-2
Nominal intensity 11 mW/cm2 @ 365 nm
Intensity deviation from nominal ≤5%
Intensity non-uniformity ≤2%

Power supply and/or lamp will be adjusted if intensity is outside the limit.

Alignment

Top Side Alignment:

  • TSA microscope standard objectives: 5X, and 10X (20X available)
  • TSA microscope special objectives: 11.25X offset (for smaller separation)
  • Minimum distance between TSA microscope objectives: 33 mm (8 mm for special objectives)
  • Maximum distance between TSA microscope objectives: 160 mm
  • TSA microscope travel range: X +/- 25mm; Y +20mm / -75mm (towards flat)

BackSide Alignment:

  • Minimum distance between BSA microscope objectives: 15 mm
  • Maximum distance between BSA microscope objectives: 100 mm
  • BSA microscope travel range: X +50mm / -16mm; Y +50mm / -20mm (towards flat)
  • BSA chuck view ranges:
    • 2": X +/- 8-22mm; Y +/- 0-6mm
    • 4": X +/- 14-46mm; Y +/- 0-10mm
    • 6": X +/- 14-69mm; Y +/- 0-10mm


Microscope field of view (W x H, splitfield):

  • TSA 5X
    • Oculars: 1.3mm x 2.6mm (Ø2.6mm full field)
    • Camera: 350µm x 500µm (700µm x 500µm full field)
  • TSA 10X
    • Oculars: 0.6mm x 1.3mm (Ø1.3mm full field)
    • Camera: 150µm x 250µm (350µm x 250µm full field)
  • TSA special
    • Oculars: 0.55mm x 1.1mm (Ø1.1mm full field)
    • Camera: 150µm x 200µm (300µm x 200µm full field)
  • BSA camera
    • Low: 1.5mm x 2mm (3mm x 2mm full field)
    • High: 450µm x 650µm (950µm x 650µm full field)

Equipment performance and process related parameters

Purpose

Mask alignment and UV exposure, potentially DUV exposure 1)

Bond alignment

Performance Exposure mode

vacuum contact, hard contact, soft contact, proximity, flood exposure

Exposure light/filters
  • broadband (i-, g-, h-line)
  • 365 nm (i-line)
  • "UV300" (280-350 nm)
  • DUV (240 nm) 1)
Minimum structure size

Typically 1.25 µm, possibly down to 0.8 µm 1)

Mask size
  • 5x5 inch
  • 7x7 inch
  • special holder for 4 x 2" designs on 5x5 inch
Alignment modes
  • Top side (TSA)
  • Backside (BSA)
Substrates Substrate size
  • small pieces 1x1cm
  • 50 mm wafers
  • 100 mm wafers
  • 150 mm wafers
Allowed materials

All cleanroom materials except copper and steel

Dedicated chuck for III-V materials

Batch

1

1) Not available yet. Requires tool change. Would require safety investigation (and a dedicated project) to become available.

Light intensity and uniformity after lamp ignition

Light intensity and uniformity after lamp ignition. CP is constant power mode, CI is constant intensity mode. The accept limits for the CI mode is ±5%, the accept limit for the non-uniformity is 2%. CP mode has no accept limit.

The lamp requires about 15 minutes of on-time after lamp ignition, before the light output has stabilized, and users using the constant power mode should always wait for 15 minutes after ignition, before starting processing.

If users use the constant intensity mode, this waiting time is not necessary, since the internal light sensor automatically adjusts the output to match the setpoint value.

The non-uniformity of the light, in CI mode, is unaffected by the stabilization time.


Aligner: Maskless 01

Aligner: Maskless 01 positioned in E-5 (2017)

The logon password for the PC is "mla" (without quotation marks).

The MLA 100 Maskless Aligner located in the E-4 cleanroom is a direct exposure lithography tool installed in 2017. It is a UV LED exposure system, that exposes the patterns directly on photosensitive resists on chips, 2, 4, and 6 inch substrates, without prior fabrication of the mask. The system offers top side alignment with high accuracy.


The user manual and contact information can be found in LabManager:

Equipment info in LabManager - requires login

Exposure dose and defocus

Information on UV exposure dose

Process information

Equipment performance and process related parameters

Purpose

Alignment and UV exposure

Performance Exposure mode

Projection

Exposure light

365nm (LED), FWHM=8nm

Focusing method

Pneumatic

Minimum structure size

down to 1µm

Design formats
  • GDS-II
  • CIF
  • DXF
  • Gerber
  • HIMT format
Alignment modes

Top side only

Substrates Substrate size
  • maximum writing area: 125x125 mm2
  • 150 mm wafer
  • 100 mm wafer
  • 50 mm wafer
  • pieces down to 5x5 mm2
Allowed materials

All cleanroom materials

Batch

1


Aligner: Maskless 02

MLA150 WMII maskless aligner from Heidelberg Instruments GmbH (installed 2019 as WMI (0.6µm resolution), rebuilt to WMII 2023).


Special features

  • Optical Autofocus
  • Backside Alignment
  • Basic Gray Scale Exposure
  • Advanced Field Alignment Mode for alignment to individual chips/devices on the substrate
  • High Aspect Ratio Mode for exposure of thick resists
  • 200 x 200 mm exposure field
  • Separate conversion PC


Training videos

Equipment info in LabManager - requires login

Exposure dose and defocus

Information on UV exposure dose

Process information

Equipment performance and process related parameters

Purpose

Alignment and UV exposure

Performance Exposure mode

Projection

Exposure light
  • 375 nm

(laser diode arrays)

Focusing method
  • Optical
  • Pneumatic
Minimum structure size

down to 1 µm

Design formats
  • GDS-II
  • CIF
  • DXF
  • Gerber
  • HIMT format
Alignment modes
  • Top side alignment
  • Backside alignment
  • Field alignment (chip-by-chip TSA)
Substrates Substrate size
  • maximum writing area: 200x200 mm2
  • 200 mm wafer
  • 150 mm wafer
  • 100 mm wafer
  • 50 mm wafer
  • pieces down to 3x3 mm2 1)
Allowed materials

All cleanroom materials

Batch

1

1) with optical autofocus


Aligner: Maskless 03

MLA150 WMII maskless aligner from Heidelberg Instruments GmbH, installed 2020.


Special features:

  • Backside Alignment
  • Advanced Field Alignment Mode for alignment to individual chips/devices on the substrate
  • Separate conversion PC (Power PC)


Training videos

Equipment info in LabManager - requires login

Exposure dose and defocus

Information on UV exposure dose

Process information

Equipment performance and process related parameters

Purpose

Alignment and UV exposure

Performance Exposure mode

Projection

Exposure light

405nm
(laser diode array)

Focusing method

Pneumatic

Minimum structure size

down to 1 µm

Design formats
  • GDS-II
  • CIF
  • DXF
  • Gerber
  • HIMT format
Alignment modes
  • Top side alignment
  • Backside alignment
  • Field alignment (chip-by-chip TSA)
Substrates Substrate size
  • maximum writing area: 150x150 mm2
  • 150 mm wafer
  • 100 mm wafer
  • 50 mm wafer
  • pieces down to 5x5 mm2
Allowed materials

All cleanroom materials

Batch

1