=The FEI Nova 600 NanoSEM=
''This section is written by DTU Nanolab internal if nothing else is stated.''
[[Category:314]]
[[Category:314-Microscopy]]


The FEI Nova 600 NanoSEM Scanning Electron Microscope is a very versatile characterization instrument that produces enlarged images of a variety of specimens, achieving magnifications of over 500.000 times providing ultra high resolution imaging. At the time being this microscope is mainly designated for microstructural characterization using [http://en.wikipedia.org/wiki/Electron_backscatter_diffraction electron backscatter diffraction (EBSD)] , [http://en.wikipedia.org/wiki/Energy-dispersive_X-ray_spectroscopy Energy-dispersive X-ray spectroscopy (EDS)] and forward scatter electron imaging by detectors recently bought from [http://www.bruker.com/ Bruker]. 
=The FEI Nova nanoSEM 600=
= Who may operate the FEI SEM =
Currently, [http://www.dtu.dk/english/Service/Phonebook/Person?id=44570&tab=1 Dr. Alice Bastos Fanta] is co-ordinating all the activities that take place on FEI Nova 600 NanoSEM. So she must be informed about the planned activities before hand in the bi-weekly meetings which are held at DTU Nanolab. Alice makes the bookings in agreement with all the users.
The rules of who may operate the FEI SEM and to what extent they may operate it are:


* '''Only users who have been trained and approved by DTU Nanolab personnel may operate the instrument. '''Irregardless of their prior experience any new user with no official DTU Nanolab training or approval cannot operate the instrument, not even under close supervision by experienced users.
The Nova is, in it's normal configuration, a very versatile characterisation instrument. Capable of producing enlarged images of a variety of specimens and achieving magnifications of over 500.000 times with ultra high resolution imaging in both high and low vacuum. The Nova has become highly evolved in it’s life at DTU Nanolab and it is possible to carry out advanced processes including in-situ experiments using heat, gas injection and probing. At the time being this microscope is mainly designated for advanced microstructural characterisation and particle analysis techniques using [http://en.wikipedia.org/wiki/Electron_backscatter_diffraction electron backscatter diffraction (EBSD)] ,[http://en.wikipedia.org/wiki/Energy-dispersive_X-ray_spectroscopy Energy-dispersive X-ray spectroscopy (EDS)] and forward scatter electron imaging by detectors recently bought from [http://www.bruker.com/ Bruker].
 
= Who may operate the Nova =
There are restrictions on who is given access to the Nova due to the development activities of the Bruker detectors fitted to the tool.
Currently, [http://www.dtu.dk/english/Service/Phonebook/Person?id=44570&tab=1 Dr. Alice Bastos Fanta] is co-ordinating the activities that take place on Nova. Alice is consulted about the planned activities beforehand in the bi-weekly meetings which are held at DTU Nanolab and coordinates the bookings in agreement with all the users.
The rules of who may operate the Nova and to what extent they may operate it are:
 
* '''Only users who have been trained and approved by DTU Nanolab personnel may operate the instrument. '''Irregardless of their prior experience any new user without official DTU Nanolab training or approval may not operate the instrument, not even under close supervision by experienced users.


*'''Users may only use the instrument to the extent they have been trained.''' This means that one should not try to operate the following options/capabilities without explicit training:
*'''Users may only use the instrument to the extent they have been trained.''' This means that one should not try to operate the following options/capabilities without explicit training:
Line 14: Line 21:
** EDX capability
** EDX capability
** Backscatter detectors
** Backscatter detectors
**STEM


= How to operate the FEI Nova 600 NanoSEM =
= How to operate the Nova =


To find the basic instructions for operating the instrument, the reader is referred to the labmanager manual.
To find the basic instructions for operating the instrument, the reader is referred to the labmanager manual.


= Processing guidelines on the FEI SEM =
= Processing guidelines on the Nova =


When one combines the 6 different detectors, 2 vacuum modes and 3 SEM modes with the traditional SEM parameters such as high voltage, working distance, spot sizes etc. it is clear that the FEI SEM has a huge number of different ways to be operated in. The number of ways that produce ultra high quality images is, however, limited. It is therefore crucial that the operators share their knowledge of how to obtain great images.
When one combines the 6 different detectors, 2 vacuum modes and 3 SEM modes with the traditional SEM parameters such as high voltage, working distance, spot sizes etc. it is clear that the FEI SEM has a huge number of different ways to be operated in. The number of ways that produce ultra high quality images is, however, limited. It is therefore crucial that the operators share their knowledge of how to obtain great images.
Line 40: Line 48:
: The TLD and Helix detectors do not provide large field of view and depth of focus when operated in SEM mode II.
: The TLD and Helix detectors do not provide large field of view and depth of focus when operated in SEM mode II.


== Detectors on the FEI SEM==
== Detectors on the Nova==


The FEI SEM has 6 electron detectors and 1 photon (X-ray) detector.
The Nova has 8 electron detectors and 2 photon (X-ray) detectors.


{| border="2" cellspacing="0" cellpadding="4" align="center"
{| border="2" cellspacing="0" cellpadding="4" align="left"
!Tag  
!Tag  
!Name
!Name
Line 69: Line 77:
|1, 2 and 3
|1, 2 and 3
|-
|-
!EBSD  
!EBSD/TKD
|Electron Backscatter Diffraction
|Electron Backscatter Diffraction
|HiVac
|HiVac+LowVac
|BSE
|BSE
|1 and 3
|1 and 2
|-
|-
!ARGUS  
!ARGUS  
|ARGUS(TM)
|ARGUS(TM)
|HiVac
|HiVac+LowVac
|BSE
|BSE,FSE and STEM (BF and DF)
|1 and 3
|1 and 2
|-
|-
!LVD
!LVD
|Low Vacuum Detector
|Low Vacuum Detector
|LoVac
|LowVac
|SE + BSE
|SE + BSE
|1 and 3
|1 and 3
Line 89: Line 97:
!Helix
!Helix
|High resolution Low Vacuum detector
|High resolution Low Vacuum detector
|LoVac
|LowVac
|SE
|SE
|1, 2 and 3
|1, 2 and 3
Line 95: Line 103:
!GAD
!GAD
|Gaseous Analytical Detector
|Gaseous Analytical Detector
|HiVac and Lovac
|HiVac and Lowvac
|SE + BSE
|BSE
|1, 2 and 3
|1, 2 and 3
|-  
|-  
!EDX
!EDS
|Energy Dispersive X-ray detector
|Energy Dispersive X-ray detector
|HiVac and Lovac
|HiVac and Lowvac
|X-ray photons
|1,2 and 3
|-
!EDS
|Flat Quad X-ray detector
|HiVac ONLY
|X-ray photons
|X-ray photons
|2 and 3  
|1,2 and 3  
|}
|}


<br clear="all" />
==Equipment performance and process related parameters==
==Equipment performance and process related parameters==
{| border="2" cellspacing="0" cellpadding="0"  
{| border="2" cellspacing="0" cellpadding="0"  
!colspan="2" border="none" style="background:silver; color:black;" align="center |Equipment  
!colspan="2" border="none" style="background:silver; color:black;" align="center |Equipment  
|style="background:WhiteSmoke; color:black" align="center"|QFEG Cryo SEM FEI Quanta FEG 200
|style="background:WhiteSmoke; color:black" align="center"|'''"Nova"''' FEI Nova nanoSEM 600
|-
|-
!style="background:silver; color:black;" align="center" width="60"| Purpose  
!style="background:silver; color:black;" align="center" width="60"| Purpose  
|style="background:LightGrey; color:black" "rowspan="2"| Vizualization and Microanalysis
|style="background:LightGrey; color:black" "rowspan="2" align="center"| Visualisation and Microanalysis
|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
* Vizualization of surfaces (topography and Z contrast)  
* Visualisation of surfaces (topography and Z contrast)  
* Vizualization of projected image (BF STEM image)
* Visualisation of projected image (BF STEM image)
* Energy Dispersive X-ray analysis (EDS)
* Energy Dispersive X-ray analysis (EDS)
*Microstructural characterisation (EBSD/TKD/STEM)
|-
|-
!style="background:silver; color:black" align="center" valign="center" rowspan="2"|Performance
!style="background:silver; color:black" align="center" valign="center" rowspan="2"|Performance
|style="background:LightGrey; color:black" rowspan="2"|Resolution
|style="background:LightGrey; color:black" rowspan="2" align="center"|Resolution
|style="background:Whitesmoke; color:black" colspan="1" align="left"|The resolution of QFEG dependends on the sample and the operation mode!
|style="background:Whitesmoke; color:black" colspan="1" align="center"|The resolution of Nova depends on the sample and is in Mode II
|-
|-
|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
* 2 nm at 30 keV for Au on C sample with the ETD detector
*High Vacuum operation in Mode II:
**1.0 nm at 15 kV (TLD detector and optimum working distance)
**1.8 nm at 1 kV (TLD detector and optimum working distance)
*Low Vacuum operation in Mode II:
**1.5 nm at 10 kV (Helix detector and optimum working distance)
**1.8 nm at 3 kV (Helix detector and optimum working distance)1 nm at  15keV for Au on C sample with the TLD detector
|-
|-
!style="background:silver; color:black" align="center" valign="center" rowspan="7"|Instrument specifics
!style="background:silver; color:black" align="center" valign="center" rowspan="7"|Instrument specifics
|style="background:LightGrey; color:black"|Detectors
|style="background:LightGrey; color:black" align="center"|Detectors
|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
* ETD- Everhart-Thornley for secondary electrons
* ETD- Everhart-Thornley for secondary electrons
Line 138: Line 159:
* CCD camera  
* CCD camera  
|-
|-
|style="background:LightGrey; color:black"|Electron source
|style="background:LightGrey; color:black" align="center"|Electron source
|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
* Field Emission - Tungsten filament
* Field Emission - Tungsten filament
|-
|-
|style="background:LightGrey; color:black"|Stage (room temperature)
|style="background:LightGrey; color:black" align="center"|Stage
|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
* X, Y: 25 &times; 25 mm
* X, Y: 75 &times; 75 mm (piezo)
* T: 0 to 60<sup>o</sup> (Full tilt require removment of cryo cold trap.)
* Z 10mm (DC motor)
* R: 360<sup>o</sup>  
* T: 0 to 70<sup>o</sup> (DC motor)
* R: 360<sup>o</sup> (piezo)
|-
|-
|style="background:LightGrey; color:black"|Peltier stage
|style="background:LightGrey; color:black" align="center"|FlatQuad EDS
|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
* FEI Peltier stage -10<sup>o</sup> to 22<sup>o</sup>
* Bruker FlatQuad (High Vacuum ONLY)
|-
|-
|style="background:LightGrey; color:black"|Cryo stage
|style="background:LightGrey; color:black" align="center"|EBSD/TKD
|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
* Quorum PP2000 Cryo System
* Bruker eFlash
|-
|-
|style="background:LightGrey; color:black"|EDS  
|style="background:LightGrey; color:black" align="center"|EDS  
|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
* Oxford Instruments 80 mm<sup>2</sup> X-Max silicon drift detector, MnKα resolution at 124 eV
* Bruker 50 mm<sup>2</sup> silicon drift detector, MnKα resolution at 124 eV
|-
|-
|style="background:LightGrey; color:black"|Operating pressures
|style="background:LightGrey; color:black" align="center"|Operating pressures
|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
* High vacuum (10<sup>-4</sup> Pa), Low vacuum and Enviromental mode (up to 2700Pa)
* High vacuum (10<sup>-4</sup> Pa), Low vacuum mode (up to 130Pa)
|-
|-
!style="background:silver; color:black" align="center" valign="center" rowspan="3"|Substrates
!style="background:silver; color:black" align="center" valign="center" rowspan="3"|Samples
|style="background:LightGrey; color:black"|Sample sizes
|style="background:LightGrey; color:black" align="center"|Sample sizes
|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
* No actual limit (limitted by stage movment and detector position).
* No actual limit (limitted by stage movment and detector position).
|-
|-
| style="background:LightGrey; color:black"|Allowed materials
| style="background:LightGrey; color:black" align="center"|Allowed materials
|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
* Conductors, Semiconductors,Insulators,Wet Samples, Biological (not pathogents!)
* Conductors, Semiconductors,Insulators,Wet Samples, Biological (not pathogents!)
|-  
|-  
|}
|}
'''Feedback to this page''': '''[mailto:labadviser@nanolab.dtu.dk?Subject=Feed%20back%20from%20page%20http://labadviser.nanolab.dtu.dk/index.php?title=LabAdviser/314/SEM/QFEG click here]'''




Line 183: Line 202:
*[[/Transmission Kikuchi diffraction|Transmission Kikuchi diffraction]]
*[[/Transmission Kikuchi diffraction|Transmission Kikuchi diffraction]]
*[[/Micro 4-point probe|Micro 4-point probe]]
*[[/Micro 4-point probe|Micro 4-point probe]]
 
*[[:File:In situ x-ray diffraction and electron microscopy investigations of noncatalytic metal oxide nanowire growth.pdf]]
 
'''Feedback to this page''': '''[mailto:labadviser@nanolab.dtu.dk?Subject=Feed%20back%20from%20page%20http://labadviser.nanolab.dtu.dk/index.php/LabAdviser/314/SEM/Nova click here]'''

Latest revision as of 14:48, 27 June 2023

Feedback to this page: click here

This section is written by DTU Nanolab internal if nothing else is stated.

The FEI Nova nanoSEM 600

The Nova is, in it's normal configuration, a very versatile characterisation instrument. Capable of producing enlarged images of a variety of specimens and achieving magnifications of over 500.000 times with ultra high resolution imaging in both high and low vacuum. The Nova has become highly evolved in it’s life at DTU Nanolab and it is possible to carry out advanced processes including in-situ experiments using heat, gas injection and probing. At the time being this microscope is mainly designated for advanced microstructural characterisation and particle analysis techniques using electron backscatter diffraction (EBSD) ,Energy-dispersive X-ray spectroscopy (EDS) and forward scatter electron imaging by detectors recently bought from Bruker.

Who may operate the Nova

There are restrictions on who is given access to the Nova due to the development activities of the Bruker detectors fitted to the tool. Currently, Dr. Alice Bastos Fanta is co-ordinating the activities that take place on Nova. Alice is consulted about the planned activities beforehand in the bi-weekly meetings which are held at DTU Nanolab and coordinates the bookings in agreement with all the users. The rules of who may operate the Nova and to what extent they may operate it are:

  • Only users who have been trained and approved by DTU Nanolab personnel may operate the instrument. Irregardless of their prior experience any new user without official DTU Nanolab training or approval may not operate the instrument, not even under close supervision by experienced users.
  • Users may only use the instrument to the extent they have been trained. This means that one should not try to operate the following options/capabilities without explicit training:
    • Low vacuum
    • EBSD capability
    • EDX capability
    • Backscatter detectors
    • STEM

How to operate the Nova

To find the basic instructions for operating the instrument, the reader is referred to the labmanager manual.

Processing guidelines on the Nova

When one combines the 6 different detectors, 2 vacuum modes and 3 SEM modes with the traditional SEM parameters such as high voltage, working distance, spot sizes etc. it is clear that the FEI SEM has a huge number of different ways to be operated in. The number of ways that produce ultra high quality images is, however, limited. It is therefore crucial that the operators share their knowledge of how to obtain great images.

Steps to follow

There is no single general approach to taking good images on the FEI SEM. These points below are only guidelines.

What kind of images are needed

It is important to consider what kind of images are needed before you start. The different requirements/conditions could be some of the following:

High resolution at high magnification (above 10.000x) images are wanted
For conducting samples: Use TLD and mount the sample so that short WD's are possible.
For non-conducting samples: Mount the Helix detector and use short (app. 5 mm) WD.
Images of very thin (> 10 nm) metal films on low atom number materials
Use the backscatter detectors
Images with high resolution, large field of view and depth of focus are wanted
The TLD and Helix detectors do not provide large field of view and depth of focus when operated in SEM mode II.

Detectors on the Nova

The Nova has 8 electron detectors and 2 photon (X-ray) detectors.

Tag Name Vacuum mode Signal SEM modes
ETD Everhardt Thornley Detector HiVac SE, BSE (detector bias adjustable) 1 and 3
TLD Through Lens Detector HiVac SE, BSE (detector bias adjustable) 1, 2 and 3
BSED BackScatter Electron Detector HiVac BSE 1, 2 and 3
EBSD/TKD Electron Backscatter Diffraction HiVac+LowVac BSE 1 and 2
ARGUS ARGUS(TM) HiVac+LowVac BSE,FSE and STEM (BF and DF) 1 and 2
LVD Low Vacuum Detector LowVac SE + BSE 1 and 3
Helix High resolution Low Vacuum detector LowVac SE 1, 2 and 3
GAD Gaseous Analytical Detector HiVac and Lowvac BSE 1, 2 and 3
EDS Energy Dispersive X-ray detector HiVac and Lowvac X-ray photons 1,2 and 3
EDS Flat Quad X-ray detector HiVac ONLY X-ray photons 1,2 and 3


Equipment performance and process related parameters

Equipment "Nova" FEI Nova nanoSEM 600
Purpose Visualisation and Microanalysis
  • Visualisation of surfaces (topography and Z contrast)
  • Visualisation of projected image (BF STEM image)
  • Energy Dispersive X-ray analysis (EDS)
  • Microstructural characterisation (EBSD/TKD/STEM)
Performance Resolution The resolution of Nova depends on the sample and is in Mode II
  • High Vacuum operation in Mode II:
    • 1.0 nm at 15 kV (TLD detector and optimum working distance)
    • 1.8 nm at 1 kV (TLD detector and optimum working distance)
  • Low Vacuum operation in Mode II:
    • 1.5 nm at 10 kV (Helix detector and optimum working distance)
    • 1.8 nm at 3 kV (Helix detector and optimum working distance)1 nm at 15keV for Au on C sample with the TLD detector
Instrument specifics Detectors
  • ETD- Everhart-Thornley for secondary electrons
  • BSD- Solid state Back Scattered Detector
  • LFD- Large Field Detector for secondary electrons
  • GSED- Gaseous Secondary Electron Detector
  • GBSD- Gaseous Backscattered Electron Detector
  • GAD- Gaseous Analystical Detector
  • STEM- right field Scanning Tramission Electron detector
  • vCD- low voltage BSED
  • CCD camera
Electron source
  • Field Emission - Tungsten filament
Stage
  • X, Y: 75 × 75 mm (piezo)
  • Z 10mm (DC motor)
  • T: 0 to 70o (DC motor)
  • R: 360o (piezo)
FlatQuad EDS
  • Bruker FlatQuad (High Vacuum ONLY)
EBSD/TKD
  • Bruker eFlash
EDS
  • Bruker 50 mm2 silicon drift detector, MnKα resolution at 124 eV
Operating pressures
  • High vacuum (10-4 Pa), Low vacuum mode (up to 130Pa)
Samples Sample sizes
  • No actual limit (limitted by stage movment and detector position).
Allowed materials
  • Conductors, Semiconductors,Insulators,Wet Samples, Biological (not pathogents!)


Characterization Techniques