'''Feedback to this page''': '''[mailto:labadviser@nanolab.dtu.dk?Subject=Feed%20back%20from%20page%20http://labadviser.nanolab.dtu.dk/index.php/LabAdviser/Microscopy_314-307/314/ATEM/HAADF-detector click here]''' | '''Feedback to this page''': '''[mailto:labadviser@nanolab.dtu.dk?Subject=Feed%20back%20from%20page%20http://labadviser.nanolab.dtu.dk/index.php/LabAdviser/Microscopy_314-307/314/ATEM/HAADF-detector click here]''' | ||
(''content by Jens Kling, January 2021'') | (''content by Jens Kling @DTU Nanolab, January 2021'') | ||
[[Category:314]] | [[Category:314]] | ||
[[Category:314-Microscopy]] | [[Category:314-Microscopy]] |
Latest revision as of 14:26, 27 June 2023
Feedback to this page: click here
(content by Jens Kling @DTU Nanolab, January 2021)
HAADF detector doesn't give a signal
Eventhough the HAADF detector is inserted properly, it sometimes doesn't give a signal. This is typically the case at the beginning of the session, when the detector is used for the first time. If this is the case, try the following:
- Go to the "STEM Detector (User)" control panel
- Open the flapout
- Un-check "HAADF"
- Check "HAADF"
- Try imaging again. The detector should give a signal.