= The Hummer Sputter coater =
<!-- = The Hummer Sputter coater =


[[image:Hummer.jpg|300x300px|right|thumb|The Hummer sputter coater located in service area Cx1]]
[[image:Hummer.jpg|300x300px|right|thumb|The Hummer sputter coater located in service area Cx1]]
Line 118: Line 120:
|}
|}


= The Sputter coater 03 (Cressington sputter coater) =
-->
 
= The Sputter coater 03 (Cressington) =


[[image:Cressington_LA_1.jpg|300x300px|right|thumb|Sputter coater 03 (Cressington) located in the 346 basement, room 907]]
[[image:Cressington_LA_1.jpg|300x300px|right|thumb|Sputter coater 03 (Cressington) located in the 346 basement, room 907]]
Line 126: Line 130:
One of the most effective ways to counter problems with charging on non-conducting samples during SEM investigations is to apply a conducting coating to the surface. The sputter coater allows you to coat small samples with a thin gold layer. When securely grounded, i.e. connected electrically with a sample holder using aluminium tape or clamps, the gold layer will effectively eliminate all charging problems in the SEM.
One of the most effective ways to counter problems with charging on non-conducting samples during SEM investigations is to apply a conducting coating to the surface. The sputter coater allows you to coat small samples with a thin gold layer. When securely grounded, i.e. connected electrically with a sample holder using aluminium tape or clamps, the gold layer will effectively eliminate all charging problems in the SEM.
   
   
The sputter coater consists of a sputter chamber with a gold target and a stage. The stage consist of four smaller stages. These small stages can rotate individually in a planetary motion, when the entire stage rotates at the same time. The rotation speed is controlled by use of a black knob in front of the sputter chamber.Small samples can be mounted on a SEM pinstub holder (only use the smallest pinstubs) with carbon pad and then placed on one of the small stages. The maximum sample size is a 4 inch wafer. If your sample is larger than one of the four small stages, or if the sample is not fastened on a pinstub, the stage rotation should not be activated.   
The sputter coater consists of a sputter chamber with a gold target and a stage. The stage consist of four smaller stages. These small stages can rotate individually in a planetary motion, when the entire stage rotates at the same time. The rotation speed can be controlled. Small samples can be mounted on a SEM pinstub holder (only use the smallest pinstubs) with carbon pads and then placed on one of the small stages. The maximum sample size is a 4 inch wafer. If your sample is larger than one of the four small stages, or if the sample is not fastened on a pinstub, the stage rotation should not be activated.   
   
   
The machine is controlled by a control panel. The sputter process will be done automatically, but the sputter time can be adjusted.
The machine is controlled by a control panel. The sputter process will be done automatically, but the sputter time can be adjusted.
Line 134: Line 138:
Argon is used as a sputter gas to generate a plasma in the sputter chamber and to vent the chamber.  
Argon is used as a sputter gas to generate a plasma in the sputter chamber and to vent the chamber.  


The Balzer sputter coater is located in the 346 basement, room 907 (next to the SEM Supra 1) .  
The Sputter coater 03 (Cressington) is located in the 346 basement, room 907 (next to the SEM Supra 1) .  
 


'''The user manual and contact information can be found in LabManager:'''  
'''The user manual and contact information can be found in LabManager:'''  


[http://labmanager.danchip.dtu.dk/function.php?module=Machine&view=view&mach=303 Sputter coater 03]
[http://labmanager.dtu.dk/function.php?module=Machine&view=view&mach=400 Sputter coater 03]
 


==Process knowledge==
==Process knowledge==


*[[Specific_Process_Knowledge/Thin_film_deposition/Deposition_of_Gold|Deposition of gold]]
*[[Specific_Process_Knowledge/Thin_film_deposition/Deposition_of_Gold|Deposition of Au]]
 
*[[Specific_Process_Knowledge/Thin film deposition/Deposition of Tungsten/Sputtering of W in Sputter Coater 3|Deposition of W]]


==Overview of the performance of the Sputter coater 03 (Cressington) and some process related parameters==
==Overview of the performance of the Sputter coater 03 (Cressington) and some process related parameters==
Line 181: Line 189:
|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
*All samples that are allowed in the SEM Supra 1
*All samples that are allowed in the SEM Supra 1
|-
|-
|}
= The Sputter coater 04 (Agar Scientific) =
[[image:IMG_6513.JPG|300x300px|right|thumb|Sputter coater 04 (Agar Scientific) located in the cleanroom Cx1]]
The Sputter coater is used to sputter a thin gold layer on different samples up to 100 mm wafers. It is not allowed to use the sputter coater for other purposes.
One of the most effective ways to counter problems with charging on non-conducting samples during SEM investigations is to apply a conducting coating to the surface. The sputter coater allows you to coat small samples with a thin gold layer. When securely grounded, i.e. connected electrically with a sample holder using aluminium tape or clamps, the gold layer will effectively eliminate all charging problems in the SEM.
The sputter coater consists of a sputter chamber with a gold target and a stage. Small samples can be mounted on a SEM pinstub holder (only use the smallest pinstubs) with carbon pads and then placed on one of the small holes. The maximum sample size is a 100 mm wafer. 
The machine is controlled by a control panel. The sputter process will be done automatically, but the sputter time can be adjusted.
Please note that the sputter rate and uniformity have not been measured yet.
Argon is used as a sputter gas to generate a plasma in the sputter chamber and to vent the chamber.
The Sputter coater 04 (Agar Scientific) is located in the cleanroom Cx1
'''The user manual and contact information can be found in LabManager:'''
[http://labmanager.dtu.dk/function.php?module=Machine&view=view&mach=421 Sputter coater 04]
==Process knowledge==
*[[Specific_Process_Knowledge/Thin_film_deposition/Deposition_of_Gold|Deposition of gold ]]
==Overview of the performance of the Sputter coater 04 (Agar Scientific) and some process related parameters==
{| border="2" cellspacing="0" cellpadding="10"
|-
!style="background:silver; color:black;" align="left"|Purpose
|style="background:LightGrey; color:black"| ||style="background:WhiteSmoke; color:black"|
*Gold sputter coating of different samples before SEM characterization
|-
!style="background:silver; color:black" align="left" valign="top" rowspan="2"|Performance
|style="background:LightGrey; color:black"|Film thickness||style="background:WhiteSmoke; color:black"|
*Not measured. 
|-
|style="background:LightGrey; color:black"|Deposition rate
|style="background:WhiteSmoke; color:black"|
*Not measured
|-
!style="background:silver; color:black" align="left" valign="top" rowspan="3"|Process parameter range
|style="background:LightGrey; color:black"|Pressure||style="background:WhiteSmoke; color:black"|
*0.01 mbar (default value)
|-
|style="background:LightGrey; color:black"|Current
|style="background:WhiteSmoke; color:black"|
*40 mA (default value)
|-
|style="background:LightGrey; color:black"|Sputter time
|style="background:WhiteSmoke; color:black"|
*Maximum 30 s
|-
!style="background:silver; color:black" align="left" valign="top" rowspan="2"|Substrates
|style="background:LightGrey; color:black"|Batch size
|style="background:WhiteSmoke; color:black"|
*One sample smaller than a 100 mm wafer
*Several smaller samples
|-
| style="background:LightGrey; color:black"|Allowed materials
|style="background:WhiteSmoke; color:black"|
*All samples that are allowed in the SEM Supra 2 and 3
|-  
|-  
|-  
|-  
|}
|}

Latest revision as of 10:44, 6 June 2023

Feedback to this page: click here

Unless otherwise stated, this page is written by DTU Nanolab internal


The Sputter coater 03 (Cressington)

Sputter coater 03 (Cressington) located in the 346 basement, room 907

The Sputter coater is used to sputter a thin gold layer on different small samples (smaller than 100 mm wafers). It is not allowed to use the sputter coater for other purposes.

One of the most effective ways to counter problems with charging on non-conducting samples during SEM investigations is to apply a conducting coating to the surface. The sputter coater allows you to coat small samples with a thin gold layer. When securely grounded, i.e. connected electrically with a sample holder using aluminium tape or clamps, the gold layer will effectively eliminate all charging problems in the SEM.

The sputter coater consists of a sputter chamber with a gold target and a stage. The stage consist of four smaller stages. These small stages can rotate individually in a planetary motion, when the entire stage rotates at the same time. The rotation speed can be controlled. Small samples can be mounted on a SEM pinstub holder (only use the smallest pinstubs) with carbon pads and then placed on one of the small stages. The maximum sample size is a 4 inch wafer. If your sample is larger than one of the four small stages, or if the sample is not fastened on a pinstub, the stage rotation should not be activated.

The machine is controlled by a control panel. The sputter process will be done automatically, but the sputter time can be adjusted.

Please note that the sputter rate and uniformity have not been measured yet.

Argon is used as a sputter gas to generate a plasma in the sputter chamber and to vent the chamber.

The Sputter coater 03 (Cressington) is located in the 346 basement, room 907 (next to the SEM Supra 1) .


The user manual and contact information can be found in LabManager:

Sputter coater 03


Process knowledge

Overview of the performance of the Sputter coater 03 (Cressington) and some process related parameters

Purpose
  • Gold sputter coating of different samples before SEM characterization
Performance Film thickness
  • Not measured.
Deposition rate
  • Not measured
Process parameter range Pressure
  • 0.01 mbar (default value)
Current
  • 40 mA (default value)
Sputter time
  • Maximum 30 s
Substrates Batch size
  • One sample smaller than a 100 mm wafer
  • Several smaller samples
Allowed materials
  • All samples that are allowed in the SEM Supra 1

The Sputter coater 04 (Agar Scientific)

Sputter coater 04 (Agar Scientific) located in the cleanroom Cx1

The Sputter coater is used to sputter a thin gold layer on different samples up to 100 mm wafers. It is not allowed to use the sputter coater for other purposes.

One of the most effective ways to counter problems with charging on non-conducting samples during SEM investigations is to apply a conducting coating to the surface. The sputter coater allows you to coat small samples with a thin gold layer. When securely grounded, i.e. connected electrically with a sample holder using aluminium tape or clamps, the gold layer will effectively eliminate all charging problems in the SEM.

The sputter coater consists of a sputter chamber with a gold target and a stage. Small samples can be mounted on a SEM pinstub holder (only use the smallest pinstubs) with carbon pads and then placed on one of the small holes. The maximum sample size is a 100 mm wafer.

The machine is controlled by a control panel. The sputter process will be done automatically, but the sputter time can be adjusted.

Please note that the sputter rate and uniformity have not been measured yet.

Argon is used as a sputter gas to generate a plasma in the sputter chamber and to vent the chamber.

The Sputter coater 04 (Agar Scientific) is located in the cleanroom Cx1


The user manual and contact information can be found in LabManager:

Sputter coater 04


Process knowledge

Overview of the performance of the Sputter coater 04 (Agar Scientific) and some process related parameters

Purpose
  • Gold sputter coating of different samples before SEM characterization
Performance Film thickness
  • Not measured.
Deposition rate
  • Not measured
Process parameter range Pressure
  • 0.01 mbar (default value)
Current
  • 40 mA (default value)
Sputter time
  • Maximum 30 s
Substrates Batch size
  • One sample smaller than a 100 mm wafer
  • Several smaller samples
Allowed materials
  • All samples that are allowed in the SEM Supra 2 and 3