Specific Process Knowledge/Lithography/Descum/plasmaAsher04
Plasma Asher 4
Product name: PVA Tepla Gigabatch 380M
Year of purchase: 2024
Descum of AZ 5214E on 100 mm wafers. The descum process development was done for a single substrate, as well as 3 substrates (for decreased ashing rate and improved ashing uniformity). The substrates were placed vertically in the glass boat.
-
Single vertical substrate
-
3 vertical substrates:
Dummy - Test - Dummy
For the 3 substrates only the center substrate was used for testing, the front and back wafers were used as dummy wafers. The wafers were placed in consecutive slots - when running multiple wafers in this way, the first and last wafers should always be dummy wafers.
The user manual, risk assessment, and contact information can be found in LabManager - requires login
Typical descum parameters
Tested with 1.5 µm AZ 5214E on 100 mm silicon substrate.
- O2: 100 sccm
- N2: 100 sccm
- Pressure (DSC): 1.3 mbar
- Power: 200 W
- Chamber temperature at start (with door closed): 30°C
- Time (single wafer): 5-10 minutes = 35-72 nm ashed
- Time (multiple wafers): 10-15 minutes = 40-80 nm ashed
| Single substrate | Center of 3 substrates | |
|---|---|---|
| Test results: ashing rate | 5.7 ±2.1 nm/min | 3.8 ±1.6 nm/min |
| Test results: non-uniformity | 0.6 ±0.4% | 0.4 ±0.2% |
| Wafers | 1 | 3 |
| Wafer size | 100 mm | 100 mm |
| Boat position | Center of chamber | Center of chamber |
| Test wafer position | Center of boat | Center of boat |
| Total gas flow rate | 200 sccm | 200 sccm |
| Gas mix ratio | 50% N2 | 50% N2 |
| Chamber pressure | 1.3 mbar | 1.3 mbar |
| Power | 200 W | 200 W |
| Test processing time | Tested parameter | Tested parameter |
| Test average temperature | 33°C | 33°C |
Single wafer descum ashing rate and uniformity for plasma asher 4 & 5
| Ashing time [min]: | 1 | 2 | 5 | 10 | 15 |
|---|---|---|---|---|---|
| Ashing amount [nm]: | 5.2 | 6.2 | 35.1 | 72.3 | 87.1 |
| Ashing rate [nm/min]: | 5.2 | 3.1 | 7.0 | 7.2 | 5.8 |
| Ashing time [min]: | 1 | 2 | 5 | 10 | 15 |
|---|---|---|---|---|---|
| Pre-descum film thickness range [nm]: | 11 | 12 | 10 | 11 | 9 |
| Pre-descum non-uniformity [%]: | 0.37 | 0.40 | 0.33 | 0.37 | 0.30 |
| Post-descum film thickness range [nm]: | 10 | 10 | 12 | 19 | 33 |
| Post-descum non-uniformity [%]: | 0.33 | 0.33 | 0.41 | 0.66 | 1.18 |
Multi wafer descum ashing rate and uniformity for plasma asher 4 & 5
| Ashing time [min]: | 1 | 2 | 5 | 10 | 15 |
|---|---|---|---|---|---|
| Ashing amount [nm]: | 4.3 | 6.7 | 10.1 | 39.5 | 78.8 |
| Ashing rate [nm/min]: | 4.3 | 3.4 | 2.0 | 4.0 | 5.3 |
| Ashing time [min]: | 1 | 2 | 5 | 10 | 15 |
|---|---|---|---|---|---|
| Pre-descum film thickness range [nm]: | 11 | 13 | 11 | 12 | 14 |
| Pre-descum non-uniformity [%]: | 0.37 | 0.43 | 0.37 | 0.40 | 0.46 |
| Post-descum film thickness range [nm]: | 11 | 9 | 10 | 12 | 21 |
| Post-descum non-uniformity [%]: | 0.37 | 0.30 | 0.33 | 0.41 | 0.74 |
Comparison between single substrate processing and multi substrate processing for plasma asher 4 & 5
Processing a single substrate using the standard descum settings will provide users with a relatively controllable and uniform process. Adding dummy substrates in close proximity with the process substrate will reduce the ashing rate and improve the uniformity:
-
Ashing amount
-
Ashing rate
-
Non-uniformity
Comparison of ashing rate between substrate sizes
It is assumed that the low power descum rates, for different substrate sizes, follows the same pattern as the high power ashing rates, which can be seen here.