Specific Process Knowledge/Lithography/Descum/plasmaAsher03
Appearance
Plasma Asher 3: Descum
Product name: Diener Pico Plasma Asher
Year of purchase: 2014
The user manual(s), quality control procedure(s) and results and contact information can be found in LabManager - requires login
Plasma Asher 3 is specifically used for controlled descum process after lithography. Please note that you only can process a single 100 mm wafer, or one small sample, at a time. The plasma asher is equipped with 2 gaslines: oxygen and nitrogen, but all standard processes use only oxygen (as recommended by Diener).
Process parameters
You can manipulate two different descum process development parameters: you can either change power or chamber pressure.
Power testing - AZ MiR 701

Recipe settings:
- Resist: AZ MiR 701
- O2 flow: 5 sccm
- N2 flow: 0
- Pressure: 0.2 mbar
- Power: Varied
| Forward/reverse | C2/C1 | Power | |
|---|---|---|---|
| recipe 1 | 50/0 | 52/31 | 50% |
| recipe 2 | 100/0 | 53/31 | 100% |
| recipe 3 | 20/0 | 51/34 | 20% |
Pressure testing - AZ MiR 701

Recipe settings:
- Resist: AZ MiR 701
- O2 flow: varied
- N2 flow: 0
- Pressure: varied
- Power: V100% (100 W)
| Forward/reverse | C2/C1 | Oxygen | Pressure | |
|---|---|---|---|---|
| recipe 1 | 100/0 | 52/31 | 5 | 0.2 |
| recipe 2 | 100/0 | 37/38 | 45 | 0.8 |
Pressure testing - AZ 5214E

Recipe settings:
- Resist: AZ 5214E
- O2 flow: varied
- N2 flow: 0
- Pressure: varied
- Power: V100% (100 W)
| Forward/reverse | C2/C1 | Oxygen | Pressure | |
|---|---|---|---|---|
| recipe 1 | 100/0 | 52/31 | 17 | 0.4 |
| recipe 2 | 100/0 | 37/39 | 45 | 0.8 |