Jump to content

Specific Process Knowledge/Characterization/SEM Supra 3: Difference between revisions

Jmli (talk | contribs)
No edit summary
Reet (talk | contribs)
Equipment performance: added QC section
Line 88: Line 88:
*Any standard cleanroom material
*Any standard cleanroom material
|-  
|-  
|}
==Quality control of length measurement ==
{| border="1" cellspacing="2" cellpadding="2" colspan="3"
|bgcolor="#98FB98" |'''Quality Control (QC) for SEM Supra 3'''
|-
|
*[http://labmanager.dtu.dk/d4Show.php?id=2493&mach=361 QC procedure for SEM Supra 3]<br>
*[https://labmanager.dtu.dk/view_binary.php?type=data&mach=361 Newest QC data for SEM Supra 3]<br>
'''QC limits:'''
{| {{table}}
| align="center" valign="top"|
{| border="2" cellspacing="1" cellpadding="2" align="center" style="width:500px"
! Detector
! Settings
! Magnification
! Measured dimensions
! Calibration limit
! Action limit
|-
|SE2
| EHT 10 kV,
WD 10 mm
| 2.440 k
(mag range 1)
| 70x70 μm
| ± 2 %
| ± 3 %
|-
|SE2
| EHT 10 kV,
WD 10 mm
| 17.000 k
(mag range 2)
| 10x10 μm
| ± 2 %
| ± 3 %
|-
|SE2
| EHT 5 kV,
WD 5 mm
| 17.000 k
| 10x10 μm
| ± 2 %
| ± 3 %
|-
|InLens
| EHT 2 kV,
WD minimum
| 17.000 k
| 10x10 μm
| ± 2 %
| ± 3 %
|-
|HDAsB (Compo)
| EHT 10 kV,
WD 5 mm
| 17.000 k
| 10x10 μm
| ± 2 %
| ± 3 %
|-
| VPSE
| EHT 10 kV,
WD 10 mm,
17 Pa
| 17.000 k
| 10x10 μm
| ± 2 %
| ± 3 %
|-
| InLens
| EHT 5 kV,
WD 5 mm
| 200.000 k
(mag range 3)
| 1x1 μm
| ± 2 %
| ± 3 %
|-
| InLens
| EHT 10 kV,
WD 3 mm
| 1.000.000 k
| (imaging only)
| NA
| NA
|-
|}
|-
|}
|}
|}