==Quality control of length measurement == | |||
{| border="1" cellspacing="2" cellpadding="2" colspan="3" | |||
|bgcolor="#98FB98" |'''Quality Control (QC) for SEM Supra 3''' | |||
|- | |||
| | |||
*[http://labmanager.dtu.dk/d4Show.php?id=2493&mach=361 QC procedure for SEM Supra 3]<br> | |||
*[https://labmanager.dtu.dk/view_binary.php?type=data&mach=361 Newest QC data for SEM Supra 3]<br> | |||
'''QC limits:''' | |||
{| {{table}} | |||
| align="center" valign="top"| | |||
{| border="2" cellspacing="1" cellpadding="2" align="center" style="width:500px" | |||
! Detector | |||
! Settings | |||
! Magnification | |||
! Measured dimensions | |||
! Calibration limit | |||
! Action limit | |||
|- | |||
|SE2 | |||
| EHT 10 kV, | |||
WD 10 mm | |||
| 2.440 k | |||
(mag range 1) | |||
| 70x70 μm | |||
| ± 2 % | |||
| ± 3 % | |||
|- | |||
|SE2 | |||
| EHT 10 kV, | |||
WD 10 mm | |||
| 17.000 k | |||
(mag range 2) | |||
| 10x10 μm | |||
| ± 2 % | |||
| ± 3 % | |||
|- | |||
|SE2 | |||
| EHT 5 kV, | |||
WD 5 mm | |||
| 17.000 k | |||
| 10x10 μm | |||
| ± 2 % | |||
| ± 3 % | |||
|- | |||
|InLens | |||
| EHT 2 kV, | |||
WD minimum | |||
| 17.000 k | |||
| 10x10 μm | |||
| ± 2 % | |||
| ± 3 % | |||
|- | |||
|HDAsB (Compo) | |||
| EHT 10 kV, | |||
WD 5 mm | |||
| 17.000 k | |||
| 10x10 μm | |||
| ± 2 % | |||
| ± 3 % | |||
|- | |||
| VPSE | |||
| EHT 10 kV, | |||
WD 10 mm, | |||
17 Pa | |||
| 17.000 k | |||
| 10x10 μm | |||
| ± 2 % | |||
| ± 3 % | |||
|- | |||
| InLens | |||
| EHT 5 kV, | |||
WD 5 mm | |||
| 200.000 k | |||
(mag range 3) | |||
| 1x1 μm | |||
| ± 2 % | |||
| ± 3 % | |||
|- | |||
| InLens | |||
| EHT 10 kV, | |||
WD 3 mm | |||
| 1.000.000 k | |||
| (imaging only) | |||
| NA | |||