LabAdviser/314/Microscopy 314-307/TEM/ETEM/dose-calibration

From LabAdviser

Feedback to this page: click here

(content by Jens Kling @DTU Nanolab, March 2020)


Set dose calibration for OneView camera in Digital Micrograph


Whenever the high tension of the microscope is changed, the dose calibration in Digital Micrograph for the OneView camera needs to be adjusted. This is done by running an installed script:

1. Go to "Customs" and "SetDoseCalibrationTag_OneView_mod-DTU"

Run script


2. Type in the high tension the microscope is running in the pop-up window

Select high tension


3. Don't change the conversion factor; just click on "OK"

Don't change conversion factor