LabAdviser/314/Microscopy 314-307/TEM/ETEM/dose-calibration
(Redirected from LabAdviser/314/TEM/ETEM/dose-calibration)
Feedback to this page: click here
(content by Jens Kling @DTU Nanolab, March 2020)
Set dose calibration for OneView camera in Digital Micrograph
Whenever the high tension of the microscope is changed, the dose calibration in Digital Micrograph for the OneView camera needs to be adjusted. This is done by running an installed script:
1. Go to "Customs" and "SetDoseCalibrationTag_OneView_mod-DTU"
2. Type in the high tension the microscope is running in the pop-up window
3. Don't change the conversion factor; just click on "OK"