We have Espacer in stock and approved for use in the cleanroom. You can find a guideline for a process flow here: Process_Flow_ZEP_ESPACER.docx. Technical information of Espacer can be found here: Espacer_300_Technical_Info.pdf, Espacer_catalog.pdf.
Please contact Lithography if you wish to test this material.
Process | No ESPACER | ESPACER |
---|---|---|
Resist | AR-P 6200/2 AllResist E-beam resist | |
Spin Coat | 1 min @ 4000 rpm, softbake 2 min @ 150 degC, thickness ~140nm | |
ESPACER | 1 min @ 2000 rpm (no softbake), Manual Spinner 1 (Laurell) | |
E-beam exposure | 2 nA, aperture 5, dose 150-310 muC/cm2, SHOT A,10 | |
Rinse | Rinsed with hand shower in fumehood, dried with N2 gun | |
Developing | SX-AR 600-54/6 60 sec, 60 sec IPA rinse | |
Characterization | ||
Nikon ECLIPSE optical microscope, E-5 | ||
Nikon ECLIPSE optical microscope, E-5 | ||
Nikon ECLIPSE optical microscope, E-5 | ||
Sputter Coated | <10nm Pt at DTU CEN | |
Zeiss SEM supra 60VP, E-3 |