Specific Process Knowledge/Characterization/AFM: Atomic Force Microscopy/Workspaces: Difference between revisions

From LabAdviser
Bghe (talk | contribs)
Bghe (talk | contribs)
Line 72: Line 72:
! Explanation
! Explanation
! SEM images
! SEM images
! Conclusion
|-  
|-  
|Box1 no. 1||bghe||A lot of triangles in the image||[[File:Box1 no1 tilt0_26.jpg|200px]][[File:Box1 no1 tilt0ny2_56.jpg|200px]]
|Box1 no. 1||bghe||A lot of triangles in the image||[[File:Box1 no1 tilt0_26.jpg|200px]][[File:Box1 no1 tilt0ny2_56.jpg|200px]]||Tip end has broken off
|-
|-
|Box1 no. 2||bghe/smurthy||Round pillars became misshaped||[[File:Box1 no2 tilt0ny2_55.jpg|200px]][[File:Box1 no2 tilt21_04.jpg|200px]]
|Box1 no. 2||bghe/smurthy||Round pillars became misshaped||[[File:Box1 no2 tilt0ny2_55.jpg|200px]][[File:Box1 no2 tilt21_04.jpg|200px]]||Tip end has broken off or maybe if was not good to begin with
|-  
|-  
|Box1 no. 3||bghe/zhongli||Force curve bad, could not get the tip off the surface||[[File:Box1 no3 tilt0_28.jpg|200px]][[File:Box1 no3 tilt0_29.jpg|200px]][[File:box1 no3 90dg_03.jpg|200px]]
|Box1 no. 3||bghe/zhongli||Force curve bad, could not get the tip off the surface||[[File:Box1 no3 tilt0_28.jpg|200px]][[File:Box1 no3 tilt0_29.jpg|200px]][[File:box1 no3 90dg_03.jpg|200px]]||Tip end has broken off
|-
|-
|Box1 no. 4||mattod||Used dirty sample, force curve very bad, tip probably broken||[[File:Box1 no4 tilt21_10.jpg|200px]]
|Box1 no. 4||mattod||Used dirty sample, force curve very bad, tip probably broken||[[File:Box1 no4 tilt21_10.jpg|200px]]||The hole cantilever has broken off
|-
|-
|Box1 no. 5||mattod||Tip broken during tip change||[[File:Box1 no5 tilt21_20.jpg|200px]]
|Box1 no. 5||mattod||Tip broken during tip change||[[File:Box1 no5 tilt21_20.jpg|200px]]||Tip/cantilever has been damaged
|-
|-
|Box1 no. 6||abshir||Bad force curve (fluctuates)||[[File:Box1 no6 tilt0_34.jpg|200px]][[File:Box1 no6 tilt0_35.jpg|200px]][[File:box1 no6 90dg_00.jpg|200px]]
|Box1 no. 6||abshir||Bad force curve (fluctuates)||[[File:Box1 no6 tilt0_34.jpg|200px]][[File:Box1 no6 tilt0_35.jpg|200px]][[File:box1 no6 90dg_00.jpg|200px]]||Tip end has broken off
|-
|-
|Box1 no. 7||bghe||Was OK on large structures but they got larger||[[File:Box1 no7 tilt0_32.jpg|200px]][[File:Box1 no7 tilt0_33.jpg|200px]]
|Box1 no. 7||bghe||Was OK on large structures but they got larger||[[File:Box1 no7 tilt0_32.jpg|200px]][[File:Box1 no7 tilt0_33.jpg|200px]]||
|-
|-
|Box1 no. 8||mattod||Bad force curve||[[File:Box1 no8 tilt0_30.jpg|200px]][[File:Box1 no8 tilt0_31.jpg|200px]]
|Box1 no. 8||mattod||Bad force curve||[[File:Box1 no8 tilt0_30.jpg|200px]][[File:Box1 no8 tilt0_31.jpg|200px]]||Tip end has broken off
|-
|-
|Box1 no. 9||kabi||Blunt tip . Definitely (triangles)||[[File:Box1 no9 tilt0ny2_49.jpg|200px]][[File:Box1 no9 tilt21_17.jpg|200px]]
|Box1 no. 9||kabi||Blunt tip . Definitely (triangles)||[[File:Box1 no9 tilt0ny2_49.jpg|200px]][[File:Box1 no9 tilt21_17.jpg|200px]]||Tip end has broken off
|-
|-
|Box1 no. 10||kabi||Blunt tip. Definitely (triangles)||[[File:Box1 no10 tilt0ny2_48.jpg|200px]][[File:box1 no10 90dg_05.jpg|200px]]
|Box1 no. 10||kabi||Blunt tip. Definitely (triangles)||[[File:Box1 no10 tilt0ny2_48.jpg|200px]][[File:box1 no10 90dg_05.jpg|200px]]||Tip end has broken off
|-
|-
|Box2 no. 1||mattod||Bad force curve||[[File:Box2 no1 tilt0ny2_46.jpg|200px]][[File:Box2 no1 tilt0ny2_47.jpg|200px]][[File:box2 no1a 90dg_13.jpg|200px]]
|Box2 no. 1||mattod||Bad force curve||[[File:Box2 no1 tilt0ny2_46.jpg|200px]][[File:Box2 no1 tilt0ny2_47.jpg|200px]][[File:box2 no1a 90dg_13.jpg|200px]]||Tip seems ok, or at least very close to okay
|-
|-
|Box2 no. 2||mattod||Broken tip||[[File:Box2 no2 tilt0ny_44.jpg|200px]][[File:Box2 no2 tilt0ny_45.jpg|200px]][[File:box2 no2 90dg_11.jpg|200px]]
|Box2 no. 2||mattod||Broken tip||[[File:Box2 no2 tilt0ny_44.jpg|200px]][[File:Box2 no2 tilt0ny_45.jpg|200px]][[File:box2 no2 90dg_11.jpg|200px]]||It looks like the tip end has been contaminated
|-
|-
|Box2 no. 3||bghe||Got large triangle scanning the tip checker||[[File:box2 no3 90dg_07.jpg|200px]]
|Box2 no. 3||bghe||Got large triangle scanning the tip checker||[[File:box2 no3 90dg_07.jpg|200px]]||Tip end has broken off
|-
|-
|}
|}

Revision as of 06:34, 30 July 2015

Feedback to this page: click here

What scanning mode, experiment/workspace and probe should I select

Roughness measurements Topographic measurements with no steep/abrupt steps Steep/abrupt steps but no high aspect ratio High aspect ratio measurements
Recommended mode Roughness measurement can be done in both Contact, Tapping and ScanAsyst mode. We recommend ScanAsyst mode because of less chance of (non-uniform) deformation of the sample and less tip wear - and because of ease of use. For Topographic measurements with no steep/abrupt steps we also recommend ScanAsyst mode due to less chance of (non-uniform) deformation of the sample and less tip wear - and because of ease of use. For Steep/abrupt steps but no high aspect ratio structures we still recommend ScanAsyst mode due to ease of use and lower top wear. For High aspect ratio measurements we recommend Tapping mode. There is not yet a ScanAsyst probe developed for high aspect ratio. Therefore we recommend Tapping mode.
Recommended probes ScanAsyst in air ScanAsyst in air ScanAsyst in air for ScanAsyst mode

TAP300Al-G for Tapping mode

AR5T-NCHR

FIB6-400A

Recommended experiment/Workspace

QNM in air*

QNM in air*

QNM in air* TappingMode 300nm trench (for steps <~1µm)

Tappping mode in air - 6µm Deep Trench (for steps >1~µm)

*By choosing the work space/experiment QNM in air (Mechanical properties - QNM in air) scanAsyst is included and so are the nanomechanical properties like modulus, adhesion, dissipation and deformation.


More information about de different modes

Contact, Tapping and Peak Force Tapping

Please flowing this link to Brukers homepage: Introduction_to_Brukers_ScanAsyst_and_PeakForce_Tapping_Atomic_Force_Microscopy_Technology_AFM_AN133.pdf

PeakForce tapping and quantitative nanomechanical mapping

Please flowing this link to Brukers homepage: PeakForce tapping and Nanomechanical option

Peak Force KPFM (Kelvin Probe Force Microscopy)

Please flowing this link to Brukers homepage: pf-kpfm.html

Evaluation of used probes

Box number and probe number User Initials Explanation SEM images Conclusion
Box1 no. 1 bghe A lot of triangles in the image Tip end has broken off
Box1 no. 2 bghe/smurthy Round pillars became misshaped Tip end has broken off or maybe if was not good to begin with
Box1 no. 3 bghe/zhongli Force curve bad, could not get the tip off the surface Tip end has broken off
Box1 no. 4 mattod Used dirty sample, force curve very bad, tip probably broken The hole cantilever has broken off
Box1 no. 5 mattod Tip broken during tip change Tip/cantilever has been damaged
Box1 no. 6 abshir Bad force curve (fluctuates) Tip end has broken off
Box1 no. 7 bghe Was OK on large structures but they got larger
Box1 no. 8 mattod Bad force curve Tip end has broken off
Box1 no. 9 kabi Blunt tip . Definitely (triangles) Tip end has broken off
Box1 no. 10 kabi Blunt tip. Definitely (triangles) Tip end has broken off
Box2 no. 1 mattod Bad force curve Tip seems ok, or at least very close to okay
Box2 no. 2 mattod Broken tip It looks like the tip end has been contaminated
Box2 no. 3 bghe Got large triangle scanning the tip checker Tip end has broken off