=Helios 5 Hydra UX DualBeam=
=Helios 5 Hydra UX DualBeam=


The Helios 5 Hydra UX DualBeam consists of two independent charged particle sources: 1- electrons, 2- plasma ions (choose from Xe, Ar, O or N).  FEG-SEM with versatile characterization capabilities. At the time being, this microscope SEM provides highest resolution among all operational SEMs at DTU Nanolab, when operated for high resolution imaging (i.e. using low kV , in-lens detector, and  in immersion mode)('''Correct? IF so remove from helios'''). This microscope is equipped with the following detectors BSE (backscatter electrons), [http://en.wikipedia.org/wiki/Electron_backscatter_diffraction electron backscatter diffraction (EBSD)] (backscatter diffraction pattern), [http://en.wikipedia.org/wiki/Energy-dispersive_X-ray_spectroscopy Energy-dispersive X-ray spectroscopy (EDS)] (X-Rays), ICE (secondary ions and electrons), etc.
The Helios 5 Hydra UX DualBeam consists of two independent charged particle sources: 1- electrons, 2- plasma ions (choose from Xe, Ar, O or N).  FEG-SEM with versatile characterization capabilities. At the time being, this microscope SEM provides highest resolution among all operational SEMs at DTU Nanolab, when operated for high resolution imaging (i.e. using low kV , in-lens detector, and  in immersion mode). This microscope is equipped with the following detectors: ETD (secondary electrons), ABS/CBS (backscatter electrons), ICE (secondary ions and electrons) etc.
Plasma focused ion beam (PFIB) with inductively coupled plasma source supporting four ion species (Xe, Ar, O, N) provides the possibility of characterization of microstructures by imaging and serial sectioning via sputtering. The microscope is equipped with a micro ('''Nano?''') manipulator for high quality TEM sample preparation and the microscope can with the AutoTEM software do automated lift-out procedures with high precision.
The plasma focused ion beam (PFIB) with inductively coupled plasma source supporting four ion species (Xe, Ar, O, N) provides the possibility of characterization of microstructures by imaging and serial sectioning via sputtering. The microscope is equipped with a nano manipulator for high quality TEM sample preparation and can with the AutoTEM software do automated lift-out procedures with high precision.
The combination of PFIB sputtering capabilities, and 2-dimentional imaging and mapping (in case of EDS and EBSD) facilitates 3-dimentional characterization.
The combination of PFIB sputtering capabilities, and 2-dimentional imaging facilitates 3-dimentional imaging.
With the Maps software, it is possible to make automated mapping of large areas for a higher resolution of the image.


=Reading Materials=
=Reading Materials=
Focused Ion Beam (FIB) [[:File:FIB-Lecture_2011 By Alice.pdf ]]


Gas Injection System (GIS) [[:File:GIS_summary2013 by Alice.pdf]]


=Technical Notes =
=Technical Notes =
Line 29: Line 33:
- PFIB column with unique inductively coupled plasma (ICP) source supporting four ion species with fast switching capability. Ion species available: Xe, Ar, O, N
- PFIB column with unique inductively coupled plasma (ICP) source supporting four ion species with fast switching capability. Ion species available: Xe, Ar, O, N


- MultiChem Gas Injection System (GIS) (''for W and Pt both E and I beams. C deposition E beam only. XeF2 and H2O etching Ion beam only.'' '''Check this''')
- MultiChem Gas Injection System (GIS) (for W, Pt and C both E and I beams.)


- Thermo Scientific EasyLift™ NanoManipulator for precise in situ sample manipulation
- Thermo Scientific EasyLift™ NanoManipulator for precise in situ sample manipulation


- Everhart-Thornley SE detector (ETD)
- Everhart-Thornley detector (ETD)


- In-chamber electron and ion detector (ICE) for secondary ions (SI) and electrons (SE)
- Charged particle detector (ICE) near the end of the ion column for detection of secondary ions (SI) and electrons (SE)
 
- EDS detector


- In-chamber Nav-Cam Sample Navigation Camera
- In-chamber Nav-Cam Sample Navigation Camera
Line 43: Line 45:
- Integrated plasma cleaner
- Integrated plasma cleaner


May have: '''(need to check)'''
- In-lens SE/BSE detector (TLD-SE, TLD-BSE)
 
- Thermo Scientific CryoCleaner Decontamination Device
 
- WDS


- EBSD
- In-column SE/BSE detector (ICD)


- Retractable low-voltage, high-contrast directional solid-state backscatter electron detector (DBS)
- Retractable low-voltage Directional Backscattered Detector (DBS), including Angular Backscattered Detector (ABS) and Concentric Backscattered Detector (CBS)  
 
- Elstar Column in-lens SE/BSE detector (TLD-SE, TLD-BSE)
 
- Elstar Column in-column SE/BSE detector (ICD)*


- STEM 3+ detector


To find the basic instructions for operating the instrument, the reader is referred to the labmanager manual.
To find the basic instructions for operating the instrument, the reader is referred to the labmanager manual.


=Important Notes=
=Important Notes=
* ALWAYS make sure the EDS (and EBSD '''relevant?''') detectors are RETRACTED when opening and closing the chamber.
* Do NOT adjust the Nitrogen regulator!!!
* Do NOT adjust the Nitrogen regulator!!! ''' Relevant?'''
* Always open and close the chamber GENTLY whilst viewing the live CCD image.
* Always open and close the chamber GENTLY whilst viewing the live CCD image.
* Only home the stage with open door or if in vacuum only with holders that you are 100% sure it is safe to do so.
* Only home the stage with open door or if in vacuum only with holders that you are 100% sure it is safe to do so.
* Make sure you delete all saved positions before finishing your microscope section.  
* Make sure you delete all saved positions before finishing your microscope section.  
* When moving between saved positions, BE CAREFUL! If you are unsure move the Z height to the lowest point and lock it so as not to collide with the pole piece.
* When moving between saved positions, BE CAREFUL! If you are unsure move the Z height to the lowest point and lock it so as not to collide with the pole piece. Enable safe stage movement.
* Use the logbook
* Use the logbook
* NEVER work without a valid link between the Z and the FWD
* NEVER work without a valid link between the Z and the FWD
* ONLY INSERT EBSD CAMERA if the sample is at the correct position (in case of a special EBSD holder the stage must be tilted to 16 °). '''Relevant?'''
* Always use GLOVES. Only vacuum-ready samples and holders are allowed inside the chamber.  
* Always use GLOVES. Only vacuum-ready samples and holders are allowed inside the chamber.  
* DO NOT install ANY software on either the microscope or support PC,.
* DO NOT install ANY software on either the microscope or support PC,.
Line 80: Line 73:


*'''Users may only use the instrument to the extent they have been trained.''' This means that one should not try to operate the following options/capabilities without explicit training:
*'''Users may only use the instrument to the extent they have been trained.''' This means that one should not try to operate the following options/capabilities without explicit training:
** EBSD capability '''Relevant?''''
** NanoManipulator (EasyLift needle)
** EDX capability
** NanoManipulator
** Plasma cleaner
** Plasma cleaner
 
** STEM3+ detector
'''Correct? Anything else?'''
** ABS/CBS


=Equipment performance and process related parameters=
=Equipment performance and process related parameters=
Line 91: Line 82:


=Process information=
=Process information=
=Tips=
Support PC can be accessed from the microscope PC using the remote desktop app (See Windows taskbar).
Alternatively switch input on the left monitor. HDMI is support PC. DP is microscope PC.

Latest revision as of 10:11, 27 June 2023

THIS PAGE IS UNDER CONSTRUCTIONUnder construction.png

Feedback to this page: click here

This section is written by DTU Nanolab internal if nothing else is stated.

Helios 5 Hydra UX DualBeam

The Helios 5 Hydra UX DualBeam consists of two independent charged particle sources: 1- electrons, 2- plasma ions (choose from Xe, Ar, O or N). FEG-SEM with versatile characterization capabilities. At the time being, this microscope SEM provides highest resolution among all operational SEMs at DTU Nanolab, when operated for high resolution imaging (i.e. using low kV , in-lens detector, and in immersion mode). This microscope is equipped with the following detectors: ETD (secondary electrons), ABS/CBS (backscatter electrons), ICE (secondary ions and electrons) etc. The plasma focused ion beam (PFIB) with inductively coupled plasma source supporting four ion species (Xe, Ar, O, N) provides the possibility of characterization of microstructures by imaging and serial sectioning via sputtering. The microscope is equipped with a nano manipulator for high quality TEM sample preparation and can with the AutoTEM software do automated lift-out procedures with high precision. The combination of PFIB sputtering capabilities, and 2-dimentional imaging facilitates 3-dimentional imaging. With the Maps software, it is possible to make automated mapping of large areas for a higher resolution of the image.

Reading Materials

Focused Ion Beam (FIB) File:FIB-Lecture_2011 By Alice.pdf

Gas Injection System (GIS) File:GIS_summary2013 by Alice.pdf

Technical Notes

Resolution

  • Electron beam:
    • At optimum WD: 0.7 nm at 1 kV • 1.0 nm at 500 V (ICD)
    • At coincident point: 0.6 nm at 15 kV • 1.2 nm at 1 kV
  • Xe Ion beam resolution at coincident point:
    • <20 nm at 30 kV using preferred statistical method
    • <10 nm at 30 kV using selective edge method

Attachments

- PFIB column with unique inductively coupled plasma (ICP) source supporting four ion species with fast switching capability. Ion species available: Xe, Ar, O, N

- MultiChem Gas Injection System (GIS) (for W, Pt and C both E and I beams.)

- Thermo Scientific EasyLift™ NanoManipulator for precise in situ sample manipulation

- Everhart-Thornley detector (ETD)

- Charged particle detector (ICE) near the end of the ion column for detection of secondary ions (SI) and electrons (SE)

- In-chamber Nav-Cam Sample Navigation Camera

- Integrated plasma cleaner

- In-lens SE/BSE detector (TLD-SE, TLD-BSE)

- In-column SE/BSE detector (ICD)

- Retractable low-voltage Directional Backscattered Detector (DBS), including Angular Backscattered Detector (ABS) and Concentric Backscattered Detector (CBS)

- STEM 3+ detector

To find the basic instructions for operating the instrument, the reader is referred to the labmanager manual.

Important Notes

  • Do NOT adjust the Nitrogen regulator!!!
  • Always open and close the chamber GENTLY whilst viewing the live CCD image.
  • Only home the stage with open door or if in vacuum only with holders that you are 100% sure it is safe to do so.
  • Make sure you delete all saved positions before finishing your microscope section.
  • When moving between saved positions, BE CAREFUL! If you are unsure move the Z height to the lowest point and lock it so as not to collide with the pole piece. Enable safe stage movement.
  • Use the logbook
  • NEVER work without a valid link between the Z and the FWD
  • Always use GLOVES. Only vacuum-ready samples and holders are allowed inside the chamber.
  • DO NOT install ANY software on either the microscope or support PC,.
  • Transfer your files at the end of each session. We will be deleting files from the support computer without notice.

Who may operate Helios 5 Hydra UX DualBeam

The rules of who may operate the Helios 5 Hydra UX DualBeam and to what extent they may operate it are:

  • Only users who have been trained and approved by DTU Nanolab personnel may operate the instrument. Irregardless of their prior experience any new user with no official DTU Nanolab training or approval cannot operate the instrument, not even under close supervision by experienced users.
  • Users may only use the instrument to the extent they have been trained. This means that one should not try to operate the following options/capabilities without explicit training:
    • NanoManipulator (EasyLift needle)
    • Plasma cleaner
    • STEM3+ detector
    • ABS/CBS

Equipment performance and process related parameters

Process information

Tips

Support PC can be accessed from the microscope PC using the remote desktop app (See Windows taskbar). Alternatively switch input on the left monitor. HDMI is support PC. DP is microscope PC.