Specific Process Knowledge/Characterization/Probe station: Difference between revisions

From LabAdviser
Jump to navigation Jump to search
Line 2: Line 2:


'''Feedback to this page''': '''[mailto:labadviser@danchip.dtu.dk?Subject=Feed%20back%20from%20page%20http://labadviser.danchip.dtu.dk/index.php/Specific_Process_Knowledge/Characterization/Probe_Station click here]'''  
'''Feedback to this page''': '''[mailto:labadviser@danchip.dtu.dk?Subject=Feed%20back%20from%20page%20http://labadviser.danchip.dtu.dk/index.php/Specific_Process_Knowledge/Characterization/Probe_Station click here]'''  
==Thickness measurer==
[[Image:Thicknessmeasurer2007.jpg|thumb|300x300px|Thickness Measurer. Positioned in cleanroom D-3]]
The purpose is to measure the thickness of wafers, depths of larger grooves or height of larger mesas.
During a KOH etch it can be helpful to ensure no over-etching by making a thickness measurement during the etching.
'''The user manual, technical information and contact information can be found in LabManager:'''
'''[http://www.labmanager.danchip.dtu.dk/function.php?module=Machine&view=view&mach=198 Thickness measurer]'''
==Quality Control - Recipe Parameters and Limits==
{| border="1" cellspacing="2" cellpadding="2" colspan="3"
|bgcolor="#98FB98" |'''Quality Control (QC) for the Thickness measurer'''
|-
|
The measured standard thickness is 0.1 mm. The measured result has to be within ± 0.005 mm. The QC is preformed once a year.
*[http://www.labmanager.danchip.dtu.dk/d4Show.php?id=1831&mach=198 The QC procedure]<br>
*[http://www.labmanager.danchip.dtu.dk/view_binary.php?fileId=2062 The newest QC data]<br>
|}
==Equipment performance and process related parameters==
{| border="2" cellspacing="2" cellpadding="3"
|-
!style="background:silver; color:black;" align="center"|
Purpose
|style="background:LightGrey; color:black"|
Thickness measurer
|style="background:WhiteSmoke; color:black"|
*Wafer thickness
*Depths of larger grooves
*Heigth of larger mesas
|-
!style="background:silver; color:black" align="center"|
Performance
|style="background:LightGrey; color:black"|
Thickness resolution
|style="background:WhiteSmoke; color:black"|
*< 5 µm
|-
!style="background:silver; color:black" align="center" valign="center" rowspan="2"|
Substrates
|style="background:LightGrey; color:black"|
Batch size
|style="background:WhiteSmoke; color:black"|
*One sample
|-
| style="background:LightGrey; color:black"|Substrate materials allowed
|style="background:WhiteSmoke; color:black"|
*No restrictions
|-
|}





Revision as of 12:52, 21 September 2017

Probe Station

Feedback to this page: click here

Thickness measurer

Thickness Measurer. Positioned in cleanroom D-3

The purpose is to measure the thickness of wafers, depths of larger grooves or height of larger mesas.

During a KOH etch it can be helpful to ensure no over-etching by making a thickness measurement during the etching.


The user manual, technical information and contact information can be found in LabManager:

Thickness measurer

Quality Control - Recipe Parameters and Limits

Quality Control (QC) for the Thickness measurer

The measured standard thickness is 0.1 mm. The measured result has to be within ± 0.005 mm. The QC is preformed once a year.


Equipment performance and process related parameters

Purpose

Thickness measurer

  • Wafer thickness
  • Depths of larger grooves
  • Heigth of larger mesas

Performance

Thickness resolution

  • < 5 µm

Substrates

Batch size

  • One sample
Substrate materials allowed
  • No restrictions


The Probe station is a - EPS150Triax - Cascade for I/V measurement, ohmic measurements etc, it has 4 individually adjustable probes, but can be fitted with more. It has several source meters, multi meters and a computer attached.

It can be used from pieces up to 6" wafers

File:Probestation.jpg
Probe station: positioned in Service room CX1

The user manual, technical information and contact information can be found in LabManager:

Probe station