Specific Process Knowledge/Characterization/KLA-Tencor Surfscan 6420/Acceptance test results

From LabAdviser
Jump to navigation Jump to search

This page is written by DTU Nanolab internal

Calibration wafers

DTU Nanolab has two (earlier tree, but one has disappeared) calibration wafers for the KLA-Tencor Surfscan. The calibration wafers are 6" silicon wafers with a number of polystyrene latex spheres (PLS) uniformily spread over the surface.


Certified mean diameter and number PLS's on each wafer:

Calibration wafer 1:

  • PLS mean diameter: 0.204 µm +/- 0.008 µm
  • Number of PLS's: 7234 +/- 10%

Calibration wafer 2:

  • PLS mean diameter: 1.112 µm +/- 0.018 µm
  • Number of PLS's: 7700 +/- 10%

Calibration wafer 3:

  • PLS mean diameter: 4.000 µm +/- 0.043 µm
  • Number of PLS's: 7587 +/- 10%

Repeatability check

Acceptance criterion:

10 scans are done on each of the three PLS calibration wafers. The wafers are loaded/unloaded to the cassette every time. The repeatbility of +/- 1σ has to be less than 1%.

Results

Calibration wafer 1:

  • Measured average PLS diameter: 0.2063 µm. Repeatability: 0,24%
  • Average number of PLS's: 7065. Repeatability: 0.11%

Calibration wafer 2:

  • Measured average PLS diameter: 1.100 µm. Repeatability: 0%
  • Average number of PLS's: 7427. Repeatability: 0.20%

Calibration wafer 3:

  • Measured average PLS diameter: 3.970 µm. Repeatability: 0,24%
  • Average number of PLS's: 7101. Repeatability: 0.10%


Uniformity check

Acceptance criterion:

Measure the average PLS diameter by boxing an area in five different locations on a PLS wafer. The measurement has to be within 3%.

Results

Calibration wafer 2:

  • Non-uniformity: 1,36%

Calibration wafer 3:

  • Non-uniformity: 0.38%


Handling test

Acceptance criterion:

Run a cassette of 25 wafers. The machine should be able to error-free handle the cassette 40 times, i.e. it should be able to handle 1000 wafers without failing.

Results

6" cassette":

A cassette of 25 wafers was handled 40 times over night, i.e. the acceptance criterion was meet.

4" and 8" cassette:

A cassette of 25 wafers was handled two times. The scanning was then stopped to save time as handling of more cassettes is not expected to cause problems.