Specific Process Knowledge/Characterization/XPS

From LabAdviser

XPS-ThermoScientific

The XPS system placed at Danchip (room 904, building 346).


A X-ray Photoecectron Spectroscopy (XPS) system can be used at Danchip. The system is a Thermo K-Alpha system, and is found in LabManager under the name XPS-ThermoScientific.

Elemental analysis

The XPS technique can be used to do elemental analysis. A comparision about techniques and intstruments used for elemental analysis at Danchip can be found on the page Element analysis.

XPS technique

XPS is a surface sensitive and non destructive technique used for analysis of the elemental composition. Only the outermost atomic layers (some nanometers) are probed, but with an ion gun etch it is possibile to probe deeper laying layers.


In the XPS spectrometer system the probed samples are irradiated by photons with a specific energy, and the photoelectrons that leaves the sample are detected. The energy levels of the electrons are elemental specific, and by measuring the energy of the outgoing electrons, it is possible to detect which elements that are present in a sample.

You can read futher about the technique here: [1]


The technique can be used for different purposes:


Elemental composition

XPS spectrum of a sample consisting of the elements silicon, oxygen and carbon.


Each element give a specific "finger-print" in the XPS spectrum. The binding energy of the electrons in atoms are different for all elements, and when measuring a photoelectron spectrum over a wide range of energies, the main line from each element will be placed at a specific energy in the spectrum.

Here is shown a spectrum measured over the energy range 0-1350 eV, and characteristic lines from three elements (C,O and Si) are seen and indicated in the spectrum.

The instrument program can use this information to give an estimate of the sample composition, giving the atomic percentage of the different elements.






Chemical state

XPS Si2p spectrum of a Si reference sample (red curve), and a Si sample that was treated in HF shortly before the measurment was done (green curve).

Due to the so called Chemical shift, it is possible to get information about the chemical state of the probed atoms. The core electrons of the atoms are affected, meaning that the binding energy of the electrons are slightly shifted, when an atom is bonded to atoms of other elements.

This gives an excellent tool for examining the chemistry of a surfaces, and how it is affected by different surface treatments.

The figure to the left gives an illustration of the effect. An XPS Si2p spectrum of a Si reference sample and a Si sample that was treated in HF shortly before the measurement, is clearly showing two different curves. The untreated spectrum has a clear feature at about 103 eV due to Si atoms bonded to oxygen. In the spectrum from the HF treated sample, only the feature steaming from Si-Si interaction is present. Note that both curves only shows the Si signal, but with an clear indication of the chemical state of the Si atoms in the samples.

If you study polymers, you can detect the precense of different chemical groups, for example (C-C),(C-OH),(C=O),(CF3) or (CF2-CH2) in the polymeric layer. And after surface treatments, you may examine differences in the polymeric layer.

Note that binding energies for different chemical states often can be found in the litterature.



Depth profiling

The composition of a NiCr as a function of film depth (etch time). The relationship between Cr and Ni is quite constant through the 70nm thick film. Measurements done with XPS-ThermoScientific.

The analysis is made on a chosen spot on the sample surface (choosen with the system camera). The technique is, as written above, very surface sensitive and probes only the top nanometers of the sample.

With the ion beam gun on the system a etch of the sample can be done. The system measures the desired spectra, do a etch step and measures again. A series of etch cycles can be set up, measuring the composition of the sample at different depths (for example at different depth of a film).


Example: NiCr film

As an illustration, a figure to the left, shows an elemental analysis through a metallic film consisting of Ni and Cr. The metallic layer was about 70 nm thick, and the atomic percentage of Ni and Cr was measured through the layer.

In the graph, you see the atomic % as a function of etch depth, and it is possible to detect that the relationship between Ni and Cr is fairly constant through the metallic film.


A rough overview of XPS-ThermoScientific characteristics

Purpose Chemical analysis
  • Probing elemental composition
  • Chemical state identification
  • Non destructive technique
  • Surface sensitive
  • Depth profiling possible by ion beam etch of sample
Performance Spot size Can be set between 30µm - 400µm
Probing depth Depending on probed element. Max probe depth lies within 10-200 Å.
Resolution Dependent on probed elements. Concentrations down to about 0,5 atomic % can in some cases be detected.
Charge compensation

Flood gun can be used for charge compensation of non conductive samples

Finding structures Choose measuring spot from camera image (magnified)
Depth profiling Purpose With ion beam etch the top layer of the material can be removed, to do a depth profiling
Ion beam size About 0,3x1 mm
Substrates Substrate size

Max 60x60 mm

Substrate thickness

Max height about 20 mm