Specific Process Knowledge/Lithography/EBeamLithography/RaithElphyManual

From LabAdviser

Feedback to this page: click here

Purpose, location and technical specifications

The Raith Elphy system is a pattern generator built onto the LEO Scanning Electron Microscope (SEM) in cleanroom F-2. All users must therefore acquire license to use the SEM LEO before acquiring license to the Raith Elphy system.

Techical Specification

The system can be characterized as follows:

  • Electron-beam scanning speeds, f, up to X MHz are available (which is maximum scan speed).
  • The acceleration voltage is maximum 25 kV.
  • The maximum field-size without stitching is X µm x Y µm.
  • The machine has a chip holder that fits to 2 - 3 chips with sizes of approximately 1 x 1 cm2 each. A larger chip or wafer can also be mounted on a holder without a Faraday's cup.

Mounting of chips or wafers into chamber

It can be difficult to align the beam properly to a chip coated with e-beam resist. It is thus recommended to remove the resist on a corner of the chip, preferably a place on the chip where you easily can adjust the beam, i.e. on metallised structures on the chip.


  1. Ventilate the SEM chamber
  2. Mount your chips(s) on the Raith Specimen holder
  3. Mount the holder on the SEM stage, evacuate the SEM chamber
Chip holder to Raith Elphy E-beam writer system; the holder has three clamps to hold up to 3 chips and a Faraday's cup to measure beam current. The holder is equipped with a dovetail that fits to the stage of the SEM.


Software login

  1. Start the Raith ELPHY Quantum software from the desktop of the Raith computer (screen to the right).
  2. Login with the following password:
User ID          administrator
Password         Renrum12
  1. Start RemCon32 from the desktop on the SEM computer (screen to the left), choose 'COM 6', right-click and press 'start'

Pepare beam for writing

Before you start preparing your beam, a few recommendations:

  • It is recommended to move the stage (by joystick) instead of deflecting the beam (ctrl + tab); this to ensure that you work with an undeflected beam while preparing the beam for patterning and while patterning your GDS.
  • It is recommended not to rotate the scan by the 'scan' knob; this...

Prepare beam

  1. Measure beam current:
    1. Click 'Stage Control/Positions/Faraday's cup', make sure the stage moves to the center of the Faraday's cup, increase the magnification to 100.000x or more
    2. Toggle beam blanker to switch on beam
    3. Make sure you operate at a working distance of 5 mm
    4. Click 'XX' to measure beam current
  2. Move the stage to a corner of your chip. With the joystick (or from the 'stage' tab in the SEM software), rotate the stage to align the chip
  3. Adjust beam quality, i.e. focus, astigmatism, wobbling at a magnification of 100.000 x or more
  4. Move to a new spot on the chip and switch the SEM to 'spot mode'; burn a spot in the resist (approximately 20s). Correct astigmatism and aperture alignment on that spot.
  5. Burn a spot with a larger magnification and adjust beam quality again.
  6. Burn a spot at a lower magnification to see that the spot is circular at this magnification

Stage adjustment

XY is stage coordinate system, UV is sample coordinate system:

  1. Origin correction:
    1. Click 'Adjustment/Adjust UV/Origin correction'
    2. Move the stage to the lower left corner of the chip
    3. Enter (U,V)=(0,0) and click 'adjust'
  2. Angle correction:
    1. Click 'Adjustment/Adjust UV/Angle Correction', make sure the system is in 'Global' mode
    2. Turn on the crosshair (SmartSEM/View)
    3. Set magnification to approximately 100x
    4. turn off beam and click 'Label1/flash'-icon to move stage to the origin.
    5. Click 'label1/pippette'-icon
    6. Move stage to the lower right corner of the chip
    7. click 'label2/pipette'-icon.
    8. Click adjust: the button hereafter turns green


Writing

  1. Click 'Design/open' to open a GDS file. Click 'View/Hierarchy/Max' from the menu bar to see full structure
  2. Set the writing field:
    1. Click 'Raith/Writefield Manager', right-click on the writing field you wish to use and click 'SET'