Specific Process Knowledge/Wafer and sample drying

From LabAdviser

Feedback to this page: click here

Drying Comparison Table

Equipment Spin dryer 1 Spin dryer 2 Spin dryer 3 Spin dryer 4 Critical point dryer
Location
  • C-1
  • B-1
  • D-4
  • E-5
  • D-4
Purpose
  • Drying
  • Drying
  • Drying
  • Drying
  • Rinsing + drying
  • Drying sensitive samples. E.g. with cantilevers
Substrates Batch size
  • # 100 mm wafers
  • # 100 mm wafers
  • # 100 mm wafers
  • # 150 mm wafers
  • # 50 mm wafers
  • # 100 mm wafers
  • # 150 mm wafers
  • 1 to 5 wafers per run. Sizes: 2”, 4" or 6"
  • Pieces (up to 10x10mm)
Allowed materials
  • No restrictions
  • Only for RCA cleaned wafers
  • No restrictions
  • No restrictions
  • Si,SiO2, Si3N4
  • Quartz and Pyrex
  • InAlP, GaAs
  • SU8


Choose a drying equipment

Process information

Standard process is 120 seconds for 100mm wafers or 180 seconds for 150mm wafers, at 2000 RPM - drying only.

If rinsing is used as well, the standard process is 60 seconds of rinse at 500 RPM, followed by 120/180 seconds of drying at 2000 RPM.