Specific Process Knowledge/Wafer and sample drying: Difference between revisions

From LabAdviser
Jehem (talk | contribs)
Rkch (talk | contribs)
 
(11 intermediate revisions by 3 users not shown)
Line 11: Line 11:
|[[Specific_Process_Knowledge/Wafer_and_sample_drying #Spin dryer 2|Spin dryer 2]]
|[[Specific_Process_Knowledge/Wafer_and_sample_drying #Spin dryer 2|Spin dryer 2]]
|[[Specific_Process_Knowledge/Wafer_and_sample_drying #Spin dryer 3|Spin dryer 3]]
|[[Specific_Process_Knowledge/Wafer_and_sample_drying #Spin dryer 3|Spin dryer 3]]
|[[Specific_Process_Knowledge/Wafer_and_sample_drying #Spin dryer 4|Spin dryer 4]]
|[[Specific_Process_Knowledge/Wafer_and_sample_drying #Spin dryer 5|Spin dryer 5]]
|[[Specific_Process_Knowledge/Wafer_and_sample_drying #Spin dryer 5|Spin dryer 5]]
|[[Specific_Process_Knowledge/Wafer_and_sample_drying #Spin dryer 6|Spin dryer 6]]
|[[Specific_Process_Knowledge/Wafer_and_sample_drying/Critical Point Dryer|Critical point dryer]]
|[[Specific_Process_Knowledge/Wafer_and_sample_drying/Critical Point Dryer|Critical point dryer]]
|[[Specific_Process_Knowledge/Wafer_and_sample_drying #Ethanol fume dryer|Ethanol fume dryer]]
|[[Specific_Process_Knowledge/Wafer_and_sample_drying #Nitrogen guns|N<sub>2</sub> guns]]
|[[Specific_Process_Knowledge/Wafer_and_sample_drying #Nitrogen guns|N<sub>2</sub> guns]]
|-
|-
Line 30: Line 29:
|
|
*D-3
*D-3
|
*C-1
|
|
*In fumehoods and at chemical benches
*In fumehoods and at chemical benches
Line 47: Line 44:
*Drying
*Drying
*Rinsing + drying
*Rinsing + drying
|
*Drying sensitive samples. E.g. with cantilevers
|
|
*Drying sensitive samples. E.g. with cantilevers
*Drying sensitive samples. E.g. with cantilevers
Line 58: Line 53:
!Batch size
!Batch size
|
|
*1-25 100 mm wafers
*100 mm wafers: 1-25
*1-25 150 mm wafers
*150 mm wafers: 1-25
|
|
*1-25 100 mm wafers
*100 mm wafers: 1-25
*1-25 150 mm wafers
*150 mm wafers: 1-25
|
|
*1-25 100 mm wafers
*100 mm wafers: 1-25
*1-25 150 mm wafers
*150 mm wafers: 1-25
|
|
*1-25 100 mm wafers
*100 mm wafers: 1-25
*1-25 150 mm wafers
*150 mm wafers: 1-25
*1-25 200 mm wafers  
*200 mm wafers: 1-25
|
|
*1 to 5 wafers per run. Sizes: 2”, 4" or 6"
*1 to 5 wafers per run. Sizes: 2”, 4" or 6"
* Pieces (up to 10x10mm)
* Pieces (up to 10x10mm)
|
*1-25 50 mm wafers
*1-25 100 mm wafers
*Pieces if a suitable carrier is available
|
|
*One sample at a time
*One sample at a time
Line 96: Line 87:
*InAlP, GaAs
*InAlP, GaAs
*SU8
*SU8
|
*No restriction except for polymers
|
|
*No restriction
*No restriction
Line 112: Line 101:




'''The user manual, technical information, and contact information can be found in LabManager: [http://labmanager.danchip.dtu.dk/function.php?module=Machine&view=view&mach=157 Spin dryer 2] [http://labmanager.danchip.dtu.dk/function.php?module=Machine&view=view&mach=158 Spin dryer 3]'''
'''The user manual, technical information, and contact information can be found in LabManager: [http://labmanager.danchip.dtu.dk/function.php?module=Machine&view=view&mach=157 Spin dryer 2] [http://labmanager.danchip.dtu.dk/function.php?module=Machine&view=view&mach=158 Spin dryer 3]  [https://labmanager.dtu.dk/function.php?module=Machine&view=view&mach=408 Spin dryer 5] [https://labmanager.dtu.dk/function.php?module=Machine&view=view&mach=323 Spin dryer 6 (8")]'''




Line 131: Line 120:
== Spin dryer 2 ==
== Spin dryer 2 ==


[[File:Spin_dryer_2ny.jpg|640px|thumb|right]]
[[File:spinDryer_2.jpg|640px|thumb|right]]


Spin dryer 2 is for RCA processing only, and is located in B-1 next to the RCA bench.
Spin dryer 2 is for RCA processing only, and is located in B-1 next to the RCA bench.
Line 141: Line 130:
== Spin dryer 3 ==
== Spin dryer 3 ==


[[File:Spin_dryer_3.jpg|640px|thumb|right]]
[[File:spinDryer_3.jpg|640px|thumb|right]]


Spin dryer 3 is located in D-3 between the working table and wet bench 03.
Spin dryer 3 is located in D-3 between the working table and wet bench 03.
Line 149: Line 138:
<br clear="all" />
<br clear="all" />


== Spin dryer 4 ==
== Spin dryer 5 ==


[[File:Spin_dryer_3.jpg|640px|thumb|right]]
[[File:spinDryer_4.jpg|640px|thumb|right]]


Spin dryer 4 is located in D-3 between the HF vapor phase etcher and wet bench 07.
Spin dryer 5 is located in D-3 between the HF vapor phase etcher and wet bench 07.
*Top chamber: 100 mm wafers
*Top chamber: 100 mm wafers
*Bottom chamber: 150 mm wafers
*Bottom chamber: 150 mm wafers
Line 159: Line 148:
<br clear="all" />
<br clear="all" />


== Spin dryer 5 ==
== Spin dryer 6 ==


[[File:Spin_dryer_5.JPG|640px|thumb|right]]
[[File:spinDryer_5.jpg|640px|thumb|right]]


Spin dryer 5 is located in F-3 Next to the Developer: TMAH stepper. It has a single process chamber that is capable of processing 100 mm, 150 mm and 200 mm wafers. The wafer size is easily changed, by changing the rotor insert frame.
Spin dryer 6 is located in F-3 Next to the Developer: TMAH stepper. It has a single process chamber that is capable of processing 100 mm, 150 mm and 200 mm wafers. The wafer size is easily changed, by changing the rotor insert frame.


<br clear="all" />
<br clear="all" /><br clear="all" />
 
= Ethanol fume dryer =
 
If your wafers are fragile or having thin membranes, cantilevers or suspended bridges that risk to stick or adhere to the surface it can be a good idea to dry your wafers in the ethanol fume dryer. Ethanol is heated to 70°C and your wafers are placed in the fumes of Ethanol which will make the water evaporate from the surface.
The Ethanol dryer is placed on a shelf or trolly in D-3 and looks like a US bath. Describtion on how to use is written on the bath.
 
 
<br clear="all" />


= Nitrogen guns =
= Nitrogen guns =

Latest revision as of 08:45, 20 November 2024

Feedback to this page: click here

Unless anything else is stated, everything on this page, text and pictures are made by DTU Nanolab.

Drying Comparison Table

Equipment Spin dryer 2 Spin dryer 3 Spin dryer 5 Spin dryer 6 Critical point dryer N2 guns
Location
  • B-1
  • D-3
  • D-3
  • F-3
  • D-3
  • In fumehoods and at chemical benches
Purpose
  • Drying
  • Drying
  • Drying
  • Drying
  • Rinsing + drying
  • Drying sensitive samples. E.g. with cantilevers
  • Drying all kind of samples.
Batch size
  • 100 mm wafers: 1-25
  • 150 mm wafers: 1-25
  • 100 mm wafers: 1-25
  • 150 mm wafers: 1-25
  • 100 mm wafers: 1-25
  • 150 mm wafers: 1-25
  • 100 mm wafers: 1-25
  • 150 mm wafers: 1-25
  • 200 mm wafers: 1-25
  • 1 to 5 wafers per run. Sizes: 2”, 4" or 6"
  • Pieces (up to 10x10mm)
  • One sample at a time
Allowed materials
  • Only for RCA cleaned wafers
  • No restrictions
  • No restrictions
  • No restrictions
  • Si,SiO2, Si3N4
  • Quartz and Pyrex
  • InAlP, GaAs
  • SU8
  • No restriction

Spin dryers

A tool used for drying wafers. A single cassette of wafers is placed in a stainless steel rotor. The wafers are centered just off the axis of rotation, and rotated at high RPM so that centrifugal force “locks” the wafers in place.

Some of the tools have DI water connected, and can be used for both rinsing and drying of wafers, while other tools do not and can only be used for drying. All wafers have to be washed in a wetbench cleaning station before going into the spin dryer, regardless of using the rinse function or not.


The user manual, technical information, and contact information can be found in LabManager: Spin dryer 2 Spin dryer 3 Spin dryer 5 Spin dryer 6 (8")


Process information

Standard process is:

  1. Rinse (with DI water) for 60 seconds at 500 RPM
  2. Dry with nitrogen for 150 seconds at 1800 RPM
  3. Dry with nitrogen and anti-static purge for 30 seconds at 1800 RPM

All spin dryers use the following carriers:

  • 100 mm wafers: A72-40MB
  • 150 mm wafers: A182-60MB
  • 200 mm wafers: A192-80M

It is not allowed to use other carriers in the spin dryers, as this may damage the motor due to incorrect mass distribution. It may also cause fragile wafers to break due to excessive vibrations caused by the mass imbalance.

Spin dryer 2

Spin dryer 2 is for RCA processing only, and is located in B-1 next to the RCA bench.

  • Top chamber: 100 mm wafers
  • Bottom chamber: 150 mm wafers


Spin dryer 3

Spin dryer 3 is located in D-3 between the working table and wet bench 03.

  • Top chamber: 100 mm wafers
  • Bottom chamber: 150 mm wafers


Spin dryer 5

Spin dryer 5 is located in D-3 between the HF vapor phase etcher and wet bench 07.

  • Top chamber: 100 mm wafers
  • Bottom chamber: 150 mm wafers


Spin dryer 6

Spin dryer 6 is located in F-3 Next to the Developer: TMAH stepper. It has a single process chamber that is capable of processing 100 mm, 150 mm and 200 mm wafers. The wafer size is easily changed, by changing the rotor insert frame.



Nitrogen guns

Nitrogen guns are valuable for drying small substrate pieces or samples that otherwise cannot go into the outer drying tools.