Specific Process Knowledge/Characterization/X-Ray Diffractometer: Difference between revisions

From LabAdviser
Bghe (talk | contribs)
Reet (talk | contribs)
No edit summary
 
(3 intermediate revisions by 2 users not shown)
Line 1: Line 1:
'''Feedback to this page''': '''[mailto:danchipsupport@danchip.dtu.dk?Subject=Feed%20back%20from%20page%20http://labadviser.danchip.dtu.dk/index.php?title=Specific_Process_Knowledge/Characterization/X-Ray_Diffractometer click here]'''
'''Feedback to this page''': '''[mailto:danchipsupport@danchip.dtu.dk?Subject=Feed%20back%20from%20page%20http://labadviser.danchip.dtu.dk/index.php?title=Specific_Process_Knowledge/Characterization/X-Ray_Diffractometer click here]'''


==X-Ray Diffractometer==
''Text and photo by staff at DTU Nanolab or DTU Electro (previously DTU Photonics Engineering)''
 
==X-Ray Diffractometer <span style="color:Red">This machine has been decommissioned - see other XRD setups at DTU Nanolab [[Specific Process Knowledge/Characterization/XRD| HERE]] </span>==


[[file:X-ray_diffractometer.jpg|300px|right|thumb|X-ray diffractometer]]
[[file:X-ray_diffractometer.jpg|300px|right|thumb|X-ray diffractometer]]
Line 8: Line 10:


''The X-Ray Diffractometer is maintained by DTU fotonik (not Nanolab) and is therefore not in LabManager!''
''The X-Ray Diffractometer is maintained by DTU fotonik (not Nanolab) and is therefore not in LabManager!''
Nanolab also has a X-Ray Diffractometer. You can see it by clicking [[Specific Process Knowledge/Characterization/XRD| HERE]]


X-ray diffraction is a non-destructive technique to measure the lattice mismatch of epitaxial grown layers. The resulting measurements are also know as rocking-curves. In this way it is possible to get the relative content of e.g. In in Ga<sub>x</sub>In<sub>1-x</sub>As grown on InP. Ga<sub>0.47</sub>In<sub>0.53</sub>As is lattice-matched to InP. Compunds containing three different materials are also called ternaries.
X-ray diffraction is a non-destructive technique to measure the lattice mismatch of epitaxial grown layers. The resulting measurements are also know as rocking-curves. In this way it is possible to get the relative content of e.g. In in Ga<sub>x</sub>In<sub>1-x</sub>As grown on InP. Ga<sub>0.47</sub>In<sub>0.53</sub>As is lattice-matched to InP. Compunds containing three different materials are also called ternaries.

Latest revision as of 09:53, 9 June 2023

Feedback to this page: click here

Text and photo by staff at DTU Nanolab or DTU Electro (previously DTU Photonics Engineering)

X-Ray Diffractometer This machine has been decommissioned - see other XRD setups at DTU Nanolab HERE

X-ray diffractometer

The Philips DCD IIH x-ray diffractometer used to characterize the layers on epitaxial semiconductor structures - usually III-V compound semiconductors.

The X-Ray Diffractometer is maintained by DTU fotonik (not Nanolab) and is therefore not in LabManager!

X-ray diffraction is a non-destructive technique to measure the lattice mismatch of epitaxial grown layers. The resulting measurements are also know as rocking-curves. In this way it is possible to get the relative content of e.g. In in GaxIn1-xAs grown on InP. Ga0.47In0.53As is lattice-matched to InP. Compunds containing three different materials are also called ternaries. For more complex compounds containing 4 elements, known as quaternaries, e.g. GaxIn1-xAsyP1-y, it is possible to obtain both x and y by combining x-ray diffraction measurements and PL-measurements (see PL Mapper).

Equipment performance and process related parameters

Performance Excitation
  • x-ray tube Cu K-alpha (K) 1.5418 Å
Detection
  • counter
Sample size
  • from 5x5mm up to 4" approximately
Resolution
  • ?? arc-sec
Materials Allowed substrate materials
  • III-V
  • Silicon