== Fluence ==
'''Unless anything else is stated, everything on this page, text and pictures are made by DTU Nanolab.'''


The fluence produces by a laser is, by definition, the energy delivered by a laser pulse (<math>E_{pulse}</math>) per surface unit (<math>S</math>) and it is given in J/cm2.  
'''All links to Kemibrug (SDS) and Labmanager Including APV and QC requires login.'''




The function <math>E_{pulse}</math> is defined as :
== Fluence definition ==
 
The fluence produces by a laser is, by definition, the energy delivered by a laser pulse (<math>E_{pulse}</math>) per surface unit (<math>A</math>) and it is given in J/cm2.
 
 
The function <math>E_{pulse}</math> is defined as:




: <math>E_{pulse} = \frac{\overline{P}}{f}\,</math>
: <math>E_{pulse} = \frac{\overline{P}}{f}\,</math>


Where <math>\overline{P}</math> (average power) and is measured with the laser sensor power measurement and the repetion rate (<math>f</math>) is set in the settings.
Where <math>\overline{P}</math> (average power) is measured with the laser sensor power measurement (see [http://labmanager.danchip.dtu.dk/d4Show.php?id=2885&mach=309|laser manual],section 5.3) and the repetion rate (<math>f</math>) is set in the parameter file under " laser frequency".
 
The fluence is therefore determinated as such:
 
 
: <math>Fluence = \frac{E_{pulse}}{A}\,</math>


The fluence is therefore determinate as such :
The area (<math>A</math>) can be determinate with the theoretical min spot diameter (1/e2), available [[Specific Process Knowledge/Back-end processing/Laser Micromachining Tool/Spot size and optics|here]]


== Fluence results measurement==


: <math>Fluence = \frac{E_{pulse}}{S}\,</math>
The results below have been calculated with the theoretical [[Specific Process Knowledge/Back-end processing/Laser Micromachining Tool/Spot size and optics|spot size]] provided by 3D Micromac.


The theoretical spot sizes (<math>S</math>) can be found [[Specific Process Knowledge/Back-end processing/Laser Micromachining Tool/Spot size and optics|here]]
{| border="1" cellspacing="2" cellpadding="2"
![[image:161214_FluenceVSrepRate_1064nm_255mm.jpg|500x500px|thumb|center|Fluence result for IR wavelength (1064nm, F255mm)]]
![[image:161214_FluenceVSrepRate_532nm_255mm.jpg|500x500px|thumb|center|Fluence result for green wavelength (532nm, F255mm)]]
![[image:161214_FluenceVSrepRate_355nm_103mm.jpg|500x500px|thumb|center|Fluence result for UV wavelength (355nm, F103mm)]]
|}
{| border="1" cellspacing="2" cellpadding="2"
![[image:060815_FluenceVSrepRate_355nm_255mm.jpg|500x500px|thumb|center|Fluence result for UV wavelength (355nm, F255mm)]]
![[image:140618_FluenceVSrepRate_1064nm_IPG.jpg|500x500px|thumb|center|Fluence result for IR wavelength (IPG 1064nm, F255mm). The fluence at 100% and 10kHz must be expected around 25J/cm2]]
|}

Latest revision as of 12:46, 6 February 2023

Unless anything else is stated, everything on this page, text and pictures are made by DTU Nanolab.

All links to Kemibrug (SDS) and Labmanager Including APV and QC requires login.


Fluence definition

The fluence produces by a laser is, by definition, the energy delivered by a laser pulse () per surface unit () and it is given in J/cm2.


The function is defined as:


Where (average power) is measured with the laser sensor power measurement (see manual,section 5.3) and the repetion rate () is set in the parameter file under " laser frequency".

The fluence is therefore determinated as such:


The area () can be determinate with the theoretical min spot diameter (1/e2), available here

Fluence results measurement

The results below have been calculated with the theoretical spot size provided by 3D Micromac.

Fluence result for IR wavelength (1064nm, F255mm)
Fluence result for green wavelength (532nm, F255mm)
Fluence result for UV wavelength (355nm, F103mm)
Fluence result for UV wavelength (355nm, F255mm)
Fluence result for IR wavelength (IPG 1064nm, F255mm). The fluence at 100% and 10kHz must be expected around 25J/cm2