This page presents the results of Cr deposition using DC sputtering in  Sputter-system Metal-Nitride (PC3), now commonly known as "Cluster Lesker". The deposition target is Cr 3-inch. Source #2 (DC/p-DC) was used.
This page presents the results of Cr deposition using DC sputtering in  Sputter-system Metal-Nitride (PC3), now commonly known as "Cluster Lesker". The deposition target is Cr 3-inch. Source #2 (DC/p-DC) was used.



Revision as of 16:50, 3 January 2023

Feedback to this page: click here

This page presents the results of Cr deposition using DC sputtering in Sputter-system Metal-Nitride (PC3), now commonly known as "Cluster Lesker". The deposition target is Cr 3-inch. Source #2 (DC/p-DC) was used.

The fabrication and characterization described below were conducted in 2022 by Evgeniy Shkondin, DTU Nanolab. The focus of the study was the deposition conditions and uniformity check across 150mm wafer.

The prepared film was investigated by the X-ray Reflectivity method. The process recipe is following:

Recipe:
  • Power unit: Power Supply 5 (DC)
  • Gun: Src2
  • Deposition type: DC
  • Gas: Ar
  • Power used: 350 W
  • Pressure: 3 mTorr
  • Temperature: 20°C (no heating)
  • Deposition time: 30 s
  • Observed DC bias: 412V


  • Deposition Rate: 0.468 nm/s


Deposition rate for Cr is a bit higher in PC3 compared to PC1 at the same power, This is due to the shorter distance between the source and the wafer (12.5 cm in PC3 vs. 15 cm in PC1).

X-ray reflectivity analysis

X-ray analysis (XRR) of Cr samples have been performed to investigate the thicknesses, roughness, and density profiles.

The scans has been obtained using Rigaku XRD SmartLab equipment. The voltage and current settings for the Cu X-ray tube were standard 40kV and 30mA. The incident optics contained a IPS (incident parallel slit) adaptor with 5 ° Soller slit. Other slits: IS=0.03mm RS1=0.03mm and RS2=0.075mm. Step size: 0.01 and measurement time - 5s for each point. Fitting procesure was performed using commercial GlobalFit software assuming the model based on Si substrates with native oxide followed by the deposited Cr film with thin oxides and moisture surfaces. The results are summarized in a tables below.




XRR results for Cr layer deposited at 3 mTorr, 350W, 30s. 150mm wafer
Top layer (moisture) CrOx Cr main layer Si native oxide Si substrate
Deposition name Moisture thickness (nm) Moisture density (g/cm3) Moisture roughness (nm) CrOx thickness (nm) CrOx density (g/cm3) CrOx roughness (nm) Cr thickness (nm) Cr density (g/cm3) Cr roughness (nm) SiO2 thickness (nm) SiO2 density (g/cm3) SiO2 roughness (nm) Si thickness (nm) Si density (g/cm3) Si roughness (nm)
Center 1.05 1.27 0.67

1.21

5.29 0.74

12.83

7.00 0.82 1.51 2.38 0.40 2.33 0.00
Edge 0.73 0.74 0.89

1.07

3.94 0.75

13.22

7.11 0.83 1.33 2.38 0.43 2.33 0.00




Fitting parameters
Sample Moisture CrOx Cr Si native oxide Si substrate Fitting parameters
Delta Beta Delta Beta Delta Beta Delta Beta Delta Beta Intensity Background Fitting area () R
Center 3.8129E-6 8.8542E-8 1.6021E-5 1.1148E-6 2.0572E-5 2.1158E-6 7.7460E-6 1.7987E-7 7.5860E-6 1.7616E-7 1.19063E+000 1.00877E-006 0.4775 - 5.0080 0.00695 0.00081
Edge 2.3226E-6 5.3933E-8 1.1725E-5 8.1591E-6 2.0899E-5 2.1494E-6 7.7588E-6 1.8017E-7 7.5860E-6 1.7616E-7 1.29242E+000 1.48454E-006 0.4775 - 5.0080 0.01019 0.00170