Specific Process Knowledge/Thin film deposition/ALD Picosun R200/AZO deposition using ALD: Difference between revisions
Line 15: | Line 15: | ||
|- | |- | ||
!# cycles | !# cycles | ||
|colspan="4" align="center"| | |colspan="4" align="center"|19 | ||
|colspan="4" align="center"|1 | |colspan="4" align="center"|1 | ||
|- | |- | ||
Line 86: | Line 86: | ||
|- | |- | ||
!# cycles | !# cycles | ||
|colspan="4" align="center"| | |colspan="4" align="center"|24 | ||
|colspan="4" align="center"|1 | |colspan="4" align="center"|1 | ||
|- | |- | ||
Line 216: | Line 216: | ||
|- | |- | ||
!# cycles | !# cycles | ||
|colspan="4" align="center"| | |colspan="4" align="center"|34 | ||
|colspan="4" align="center"|1 | |colspan="4" align="center"|1 | ||
|- | |- |
Revision as of 16:55, 16 March 2017
THIS PAGE IS UNDER CONSTRUCTION![Under construction.png](/images/thumb/f/f8/Under_construction.png/200px-Under_construction.png.jpeg)
Et2Zn or Zn(CH3CH2)2 = DEZ (diethylzinc)
Recipe: AZO 20T
Maximum deposition thickness: xx nm
Temperature: x oC - xx oC
# macrocycles | n | |||||||
---|---|---|---|---|---|---|---|---|
# cycles | 19 | 1 | ||||||
Precursor | DEZ | DEZ | H2O | H2O | TMA | TMA | H2O | H2O |
Nitrogen flow | 200 sccm | 200 sccm | 150 sccm | 150 sccm | 200 sccm | 200 sccm | ||
Pulse time | 0.1 s | 0.1 s | 0.1 s | 0.1 s | 0.1 s | 0.1 s | 0.1 s | 0.1 s |
Purge time | 0.5 s | 20.0 s | 0.5 s | 20.0 s | 0.5 s | 20.0 s | 0.5 s | 20.0 s |
How to fill out the process log in LabManager:
The pulse time for each precursor equals the total pulse time times the number of cycles in each macroscycle.
DEZ pulse time: (0.1 + 0.1) s * 20 = 4 s
TMA pulse time: (0.1 + 0.1 )s * 1 = 0.2 s
H2O pulse time: [(0.1 + 0.1) s * 20] + [(0.1 + 0.1) s * 1] = 4.2 s
The number of cycles for each precursor now equals the number of macrocycles (n).
Recipe: AZO 25T
Maximum deposition thickness: xx nm
Temperature: x oC - xx oC
# macrocycles | n | |||||||
---|---|---|---|---|---|---|---|---|
# cycles | 24 | 1 | ||||||
Precursor | DEZ | DEZ | H2O | H2O | TMA | TMA | H2O | H2O |
Nitrogen flow | 200 sccm | 200 sccm | 150 sccm | 150 sccm | 200 sccm | 200 sccm | ||
Pulse time | 0.1 s | 0.1 s | 0.1 s | 0.1 s | 0.1 s | 0.1 s | 0.1 s | 0.1 s |
Purge time | 0.5 s | 20.0 s | 0.5 s | 20.0 s | 0.5 s | 20.0 s | 0.5 s | 20.0 s |
How to fill out the process log in LabManager:
The pulse time for each precursor equals the total pulse time times the number of cycles in each macroscycle.
DEZ pulse time: (0.1 + 0.1) s * 25 = 5 s
TMA pulse time: (0.1 + 0.1 )s * 1 = 0.2 s
H2O pulse time: [(0.1 + 0.1) s * 20] + [(0.1 + 0.1) s * 1] = 5.2 s
The number of cycles for each precursor now equals the number of macrocycles (n).
Recipe: AZO 30T
Maximum deposition thickness: xx nm
Temperature: x oC - xx oC
# macrocycles | n | |||||||
---|---|---|---|---|---|---|---|---|
# cycles | 30 | 1 | ||||||
Precursor | DEZ | DEZ | H2O | H2O | TMA | TMA | H2O | H2O |
Nitrogen flow | 200 sccm | 200 sccm | 150 sccm | 150 sccm | 200 sccm | 200 sccm | ||
Pulse time | 0.1 s | 0.1 s | 0.1 s | 0.1 s | 0.1 s | 0.1 s | 0.1 s | 0.1 s |
Purge time | 0.5 s | 20.0 s | 0.5 s | 20.0 s | 0.5 s | 20.0 s | 0.5 s | 20.0 s |
Recipe: AZO 35T
Maximum deposition thickness: xx nm
Temperature: x oC - xx oC
# macrocycles | n | |||||||
---|---|---|---|---|---|---|---|---|
# cycles | 34 | 1 | ||||||
Precursor | DEZ | DEZ | H2O | H2O | TMA | TMA | H2O | H2O |
Nitrogen flow | 200 sccm | 200 sccm | 150 sccm | 150 sccm | 200 sccm | 200 sccm | ||
Pulse time | 0.1 s | 0.1 s | 0.1 s | 0.1 s | 0.1 s | 0.1 s | 0.1 s | 0.1 s |
Purge time | 0.5 s | 20.0 s | 0.5 s | 20.0 s | 0.5 s | 20.0 s | 0.5 s | 20.0 s |