LabAdviser/314/Microscopy 314-307/TEM/ETEM/OneView restart

From LabAdviser
< LabAdviser‎ | 314‎ | Microscopy 314-307‎ | TEM‎ | ETEM
Jump to navigation Jump to search

Feedback to this page: click here

(content by Jens Kling @DTU Nanolab, March 2022)


Communication/Connection problems

An error message about retraction or insertion of the camera in GMS (Digital Micrograph) can point towards a communication/connection problem to the hardware controller. It typically comes together with an IP-address in the error message (siehe example screen shots).

Possible error message 1
Possible error message 2


If this is the case, try the following procedure:

  • Accept the error message and close GMS
  • Go to the bottom rack behind the monitors (behind the table)
  • Open the front door of the rack (push the two lock mechanisms towards the center and remove the door)
  • At the very bottom is the Gatan controller for the OneView camera
  • Hold a finger on the blue lighting Gatan logo (it is a sensor button, not a mechanical button) --> the controller will switch off
  • Wait a few seconds and re-start the controller (finger on the sensor button)
  • Close the front door and re-start GMS