Specific Process Knowledge/Lithography/Descum
Feedback to this page: click here
Descum results
Plasma asher 1
Descum of AZ5214E resist on 50mm silicon wafer. Wafer was placed horisontally in chamber on a 100 mm carier wafer.
Note: Plasma asher was cold before use
|
Conny Hjort & Jesper Hanberg September 2019
Descum of different resists, AZ5214E, AZ701 MiR and AZ2020 nLOF, on a single 100mm wafer was tested. Wafer was placed vertically in the middle of glass carrier.
Recipe setting was the same as in previous test in September 2019: 70ml/min O2, 70 ml/min N2, power 150W, different ashing time 1, 2, 3, 5 and 7 min run. Plasma Asher was cold before use, we observed minor temperature rise during processing, but not more than 5 degrees. Starting chambers pressure was around 0, 5 mbar.
1,5 um AZ5214E resist
|
1,5 um AZ5214E resist placed horizontally in the carrier
|
1,5 um AZ701MiR resist
|
1,5 um AZ 2020nLOF resist
|
Plasma asher 2
Descum of AZ Mir 701 resist on 100mm silicon wafer. Five wafers were placed vertically in chamber.
Experiment parameters:
|
recipe 1
|
recipe 2
|
We can observe linear dependance of etched material on time after etching 7 minutes and more (recipe 2).
Jitka Urbánková & Jesper Hanberg
December 2019
Plasma asher 3
Plasma Asher 3 is specially used for control descum process after lithography. Please notice that you only can process one 4 inch wafer or one small sampel at a time. Machine is equipped with 2 gaslines: oxygen and nitrogen, but all tests run with oxygen as recommended by Diener.
Ashing of AZ MiR701 resist
You can use different strategy planing your descum: you can change power settings or you can vary chamber pressure during descum.
Testing different power settings
Recipe settings: Kept oxygen and pressure settings constant at Oxydgen: 5 sccm under process; Pressure: 0,2mbar and vary power.
Experiment parameters:
|
Testing different pressure settings
Recipe settings: Kept power setting constant at Power: 100% and vary oxydgen flow during process.
Experiment parameters:
|
Ashing of AZ5214E resist
Recipe settings: Kept power setting constant at Power: 100% and vary oxydgen flow during process.
Experiment parameters:
|