=RTP Annealsys - Rapid Thermal Processor=
=RTP Annealsys - Rapid Thermal Processor=
Line 5: Line 5:
''This page is written by Inês Diogo@DTU Nanolab if nothing else is stated.''
''This page is written by Inês Diogo@DTU Nanolab if nothing else is stated.''


''January 2023: The RTP Annealsys is being tested, but it is not released for general use yet. Please contact thinfilm@nanolab.dtu.dk for more information.''
''July 2023: The RTP Annealsys is a research tool. Thus, it is not released for general use, only for selected users. Please contact thinfilm@nanolab.dtu.dk for more information.''


'''RTP Annealsys (RTP AS-Premium, serial number AS0415C4 - 7484, from ANNEALSYS)''' is a research tool available at DTU Nanolab that can reach very high temperatures in a matter of a few minutes or even seconds. Therefore, it is able to perform several types of rapid thermal processing and it can be used as a multi-functional and versatile microfabrication tool, within specific process windows for each process type, using Si/nanofabricated Si structures. Amongst them, there is rapid thermal annealing '''(RTA)''', hydrogenation '''(RTH)''', oxidation '''(RTO)''', nitridation '''(RTN)''', etc. Currently, it is used for '''rapid thermal annealing and smoothing''' of silicon-based samples.
'''RTP Annealsys (RTP AS-Premium, serial number AS0415C4 - 8177, from ANNEALSYS)''' is a research tool available at DTU Nanolab that can reach very high temperatures in just a few minutes or even seconds. Therefore, it is able to perform several types of rapid thermal processing and it can be used as a multi-functional and versatile microfabrication tool, within specific process windows for each process type, using Si/nanofabricated Si structures. Amongst them, there are rapid thermal annealing '''(RTA)''', hydrogenation '''(RTH)''', oxidation '''(RTO)''', nitridation '''(RTN)''', etc. Currently, it is used for '''rapid thermal annealing and smoothing''' of silicon-based samples.


[[File:LL&station Annealsys.png|450px|thumb|right|The RTP Annealsys work station and loadlock are located in the DTU Nanolab cleanroom B-1. Photo: Maria Farinha@DTU Nanolab, February 2023]]
[[File:LL&station Annealsys.png|450px|thumb|right|The RTP Annealsys work station and loadlock are located in the DTU Nanolab cleanroom B-1. Photo: Maria Farinha@DTU Nanolab, February 2023]]
Line 15: Line 15:
The RTP Annealsys system is divided into two main parts: the process chamber and the loadlock.
The RTP Annealsys system is divided into two main parts: the process chamber and the loadlock.


The chamber has stainless steel walls (water-cooled) and a top and bottom halogen lamp-configuration (32 infra-red lamps, in total). As such, the samples are rapidly heated from both sides, simultaneously. In addition, the chamber is enclosed by two quartz-windows (immediately below/above the set of top/bottom lamps) and these are air-cooled. It is also connected to two external vacuum pumps, to a gas inlet and the loadlock. Similarly, the loadlock is connected to two external pumps.
The chamber has stainless steel walls (water-cooled) and a top and bottom halogen lamp-configuration (28 infra-red lamps, in total). As such, the samples are rapidly heated from both sides, simultaneously. In addition, the chamber is enclosed by two quartz-windows (immediately below/above the set of top/bottom lamps) and these are air-cooled. It is also connected to two external vacuum pumps, to a gas inlet and the loadlock. Similarly, the loadlock is connected to two external pumps.


Moreover, the chamber and the loadlock are separated by the gate valve. While processing, the valve remains closed; it is only opened to exchange the substrates when both chamber and loadlock are under turbo vacuum. The substrate transfer is possible due to the mechanical, retractable arm.
Moreover, the chamber and the loadlock are separated by the gate valve. While processing, the valve remains closed; it is only opened to exchange the substrates when both chamber and loadlock are under turbo vacuum. The substrate transfer is possible due to the mechanical, retractable arm.
Line 37: Line 37:
|-
|-
!style="background:Silver; color:black" align="center" align="center" rowspan="2"|Temperature
!style="background:Silver; color:black" align="center" align="center" rowspan="2"|Temperature
|style="background:Silver; color:black"|Temperature Control
|style="background:Silver; color:black"|Pyro Control
|style="background:LightGrey; color:black"|
|style="background:LightGrey; color:black"|
* up to 1250 <sup>o</sup>C
* 700 <sup>o</sup>C to 1200 <sup>o</sup>C
|style="background:WhiteSmoke; color:black;" align="left"|The maximum temperature allowed is '''dependent''' on both '''processing type''' and '''time'''.
|style="background:WhiteSmoke; color:black;" align="left"|The maximum temperature allowed is '''dependent''' on both '''processing type''' and '''time'''.
|-
|-
Line 45: Line 45:
|style="background:LightGrey; color:black"|
|style="background:LightGrey; color:black"|
* 0% - 100%
* 0% - 100%
|style="background:WhiteSmoke; color:black;" align="left"|
|style="background:WhiteSmoke; color:black;" align="left"| '''Power control is NOT ALLOWED for more than 10s.''' Chamber maximum power is '''56 kW'''.
|-
|-
!style="background:silver; color:black" align="center" align="center" rowspan="4"|Process gas
!style="background:silver; color:black" align="center" align="center" rowspan="4"|Process gas
|style="background:Silver; color:black"|Ar
|style="background:Silver; color:black"|Ar
|style="background:LightGrey; color:black"|
|style="background:LightGrey; color:black"|
*up to 2000 SCCM
*Max. 2000 SCCM
|style="background:WhiteSmoke; color:black;" align="center"|
|style="background:WhiteSmoke; color:black;" align="center"|
|-
|-
|style="background:Silver; color:black"|O<sub>2
|style="background:Silver; color:black"|O<sub>2
|style="background:LightGrey; color:black"|
|style="background:LightGrey; color:black"|
*up to 2000 SCCM
*Max. 2000 SCCM
|style="background:WhiteSmoke; color:black;" align="left"|'''Not allowed''' for users.
|style="background:WhiteSmoke; color:black;" align="left"|Limited use.
|-
|-
|style="background:Silver; color:black"|NH<sub>3
|style="background:Silver; color:black"|NH<sub>3
|style="background:LightGrey; color:black"|
|style="background:LightGrey; color:black"|
*up to 2000 SCCM
*Max. 2000 SCCM
|style="background:WhiteSmoke; color:black;" align="left"|'''Not allowed''' for users.
|style="background:WhiteSmoke; color:black;" align="left"|'''Not connected and not tested'''. No recipes available.
|-
|-
|style="background:Silver; color:black"|5% H<sub>2</sub>/Ar
|style="background:Silver; color:black"|5% H<sub>2</sub>/Ar
|style="background:LightGrey; color:black"|
|style="background:LightGrey; color:black"|
*up to 2000 SCCM
*Max. 2000 SCCM
|style="background:WhiteSmoke; color:black;" align="center"|
|style="background:WhiteSmoke; color:black;" align="center"|
|-
|-
Line 76: Line 76:
|style="background:Silver; color:black"|Controller
|style="background:Silver; color:black"|Controller
|style="background:LightGrey; color:black"|
|style="background:LightGrey; color:black"|
*up to 12 mbar
*Max. 13.3(3) mbar
|style="background:WhiteSmoke; color:black;" align="center"|
|-
!style="background:Silver; color:black" align="center" align="center" rowspan="3"|Process time
(Check image below)
|style="background:Silver; color:black"|At 1200 <sup>o</sup>C
|style="background:LightGrey; color:black"|
*Max. 10 min
|style="background:WhiteSmoke; color:black;" align="center"|
|style="background:WhiteSmoke; color:black;" align="center"|
|-
|-
!style="background:Silver; color:black" align="center" align="center" rowspan="1"|Process time
|style="background:Silver; color:black"|At 1100 <sup>o</sup>C
|style="background:Silver; color:black"|
|style="background:LightGrey; color:black"|
|style="background:LightGrey; color:black"|
* 0.1s to 2h46min
*Max. 30 min
|style="background:WhiteSmoke; color:black;" align="left"|
|style="background:WhiteSmoke; color:black;" align="left"|
|-
|-
|style="background:Silver; color:black"|At 1000 <sup>o</sup>C
|style="background:LightGrey; color:black"|
*Max. 60 min
|style="background:WhiteSmoke; color:black;" align="left"|
|-
!style="background:Silver; color:black" align="center" align="center" rowspan="1"|Heating rate
!style="background:Silver; color:black" align="center" align="center" rowspan="1"|Heating rate
|style="background:Silver; color:black"|
|style="background:Silver; color:black"|
|style="background:LightGrey; color:black"|
|style="background:LightGrey; color:black"|
* Max. 150 <sup>o</sup>C/s
* Max. 150 <sup>o</sup>C/s
|style="background:WhiteSmoke; color:black;" align="left"|
|style="background:WhiteSmoke; color:black;" align="left"| Different heating rates can be set under temperature control, as well as the step duration.
|-
|-
!style="background:silver; color:black" align="center" valign="center" rowspan="3"|Substrates
!style="background:silver; color:black" align="center" valign="center" rowspan="3"|Substrates
Line 101: Line 112:
*Chips on carrier
*Chips on carrier
*100 mm or 150 mm wafers
*100 mm or 150 mm wafers
|style="background:WhiteSmoke; color:black;" align="left"| '''Small samples''' must be '''placed''' on '''dedicated carrier wafers'''. These '''do not need''' to be bonded to the carriers.
|style="background:WhiteSmoke; color:black;" align="left"| '''Small samples''' must be '''placed''' on '''dedicated carrier wafers'''. These '''do not need to and should not be''' bonded to the carriers.
|-
|-
| style="background:Silver; color:black"|Allowed materials
| style="background:Silver; color:black"|Allowed materials
Line 108: Line 119:
*Silicon Nitride
*Silicon Nitride
*Aluminum Oxide
*Aluminum Oxide
|style="background:WhiteSmoke; color:black;" align="left"| '''Silicon nitride''' and '''aluminum oxide''' are only used as '''coating/masking materials''' and have been proven capable of sustaining this type of high temperature processing.
|style="background:WhiteSmoke; color:black;" align="left"| During previous testing, '''silicon nitride''' and '''aluminum oxide''' were only used as '''coating/masking materials''' and have been proven capable of sustaining this type of high temperature processing.
|-  
|-  
|}
|}
[[File:PlotAnnealsys.png|650px|thumb|center| Graphic representation of the recommended annealing duration in respect to the annealing temperature. Image: Inês Diogo@DTU Nanolab, July 2023]]


==Previous work on RTP Annealsys==
==Previous work on RTP Annealsys==
Line 116: Line 131:
More information and details about each type of processing - achievable using the RTP Annealsys tool - can be found in the following documents.  
More information and details about each type of processing - achievable using the RTP Annealsys tool - can be found in the following documents.  


*[[media:Report_Annealsys_2022.pdf|Report_Annealsys_December22 by Inês Diogo@DTU Nanolab]]
*[[media:Report_Updated.pdf|Report_Annealsys_Updated_February23 by Inês Diogo@DTU Nanolab]]


*[https://labmanager.dtu.dk/view_binary.php?class=MiscDocument&id=5&name=Dissertation_ID_Rapid+Thermal+Processing+and+its+Effects+on+High+Aspect+Ratio+Silicon+Features.pdf| Rapid Thermal Processing and its Effects on High Aspect Ratio Silicon Features_October22 by Inês Diogo@DTU Nanolab] - requires login
*[http://hdl.handle.net/10362/152031 Rapid Thermal Processing and its Effects on High Aspect Ratio Silicon Features_October22 by Inês Diogo@DTU Nanolab@FCT NOVA]


'''''Important!''''' The '''''RTO sequences''''' that were developed during the previous experimental work on the RTP Annealsys are '''''not available to users'''''. More tests and further investigation are required to prevent damaging the tool.
'''''Important!''''' The '''''RTO sequences''''' that were developed during the previous experimental work on the RTP Annealsys are '''''not available to users'''''. More tests and further investigation are required to prevent damaging the tool.
'''Feedback to this page''': '''[mailto:labadviser@nanolab.dtu.dk?Subject=Feed%20back%20from%20page%20https://labadviser.nanolab.dtu.dk/index.php?title=Specific_Process_Knowledge/Thermal_Process/RTP_Annealsys click here]'''

Latest revision as of 10:52, 6 November 2023

Feedback to this page: click here

RTP Annealsys - Rapid Thermal Processor

This page is written by Inês Diogo@DTU Nanolab if nothing else is stated.

July 2023: The RTP Annealsys is a research tool. Thus, it is not released for general use, only for selected users. Please contact thinfilm@nanolab.dtu.dk for more information.

RTP Annealsys (RTP AS-Premium, serial number AS0415C4 - 8177, from ANNEALSYS) is a research tool available at DTU Nanolab that can reach very high temperatures in just a few minutes or even seconds. Therefore, it is able to perform several types of rapid thermal processing and it can be used as a multi-functional and versatile microfabrication tool, within specific process windows for each process type, using Si/nanofabricated Si structures. Amongst them, there are rapid thermal annealing (RTA), hydrogenation (RTH), oxidation (RTO), nitridation (RTN), etc. Currently, it is used for rapid thermal annealing and smoothing of silicon-based samples.

The RTP Annealsys work station and loadlock are located in the DTU Nanolab cleanroom B-1. Photo: Maria Farinha@DTU Nanolab, February 2023

The Set-Up

The RTP Annealsys system is divided into two main parts: the process chamber and the loadlock.

The chamber has stainless steel walls (water-cooled) and a top and bottom halogen lamp-configuration (28 infra-red lamps, in total). As such, the samples are rapidly heated from both sides, simultaneously. In addition, the chamber is enclosed by two quartz-windows (immediately below/above the set of top/bottom lamps) and these are air-cooled. It is also connected to two external vacuum pumps, to a gas inlet and the loadlock. Similarly, the loadlock is connected to two external pumps.

Moreover, the chamber and the loadlock are separated by the gate valve. While processing, the valve remains closed; it is only opened to exchange the substrates when both chamber and loadlock are under turbo vacuum. The substrate transfer is possible due to the mechanical, retractable arm.

Schematic representation of the rapid thermal processor set-up profile view. The substrate (in green) rests on top of three quartz pins, supported by a quartz holder (in black). The drawing is not to scale. Image: Inês Diogo@DTU Nanolab, December 2022

Temperature Measurement and Control

RTP Annealsys offers an optical pyrometry system for temperature measurement, not requiring any contact with the wafer.

Regarding setting the process temperature, it is also important to mention the tool can function under two distinct modes – power and temperature/pyro control modes. In other words, using power control mode, the tool is programmed to fix the lamp’s power at a specific value (%) and consequently, the temperature inside the chamber varies with time; using the temperature/pyro control mode, the tool is programmed to vary the lamps’ power to maintain the temperature stable and constant inside the chamber at a previously established value in the recipe.

Samples and Process Specifications

The RTP Annealsys, with a base pressure of 10-6 mbar, is a research tool. As such, there are some parameters and details to consider.

Specifics Range Comments
Temperature Pyro Control
  • 700 oC to 1200 oC
The maximum temperature allowed is dependent on both processing type and time.
Power Control
  • 0% - 100%
Power control is NOT ALLOWED for more than 10s. Chamber maximum power is 56 kW.
Process gas Ar
  • Max. 2000 SCCM
O2
  • Max. 2000 SCCM
Limited use.
NH3
  • Max. 2000 SCCM
Not connected and not tested. No recipes available.
5% H2/Ar
  • Max. 2000 SCCM
Pressure Valve (APC)
  • 0o - 90o
Controller
  • Max. 13.3(3) mbar
Process time

(Check image below)

At 1200 oC
  • Max. 10 min
At 1100 oC
  • Max. 30 min
At 1000 oC
  • Max. 60 min
Heating rate
  • Max. 150 oC/s
Different heating rates can be set under temperature control, as well as the step duration.
Substrates Batch size
  • Single-wafer process
Substrate size
  • Chips on carrier
  • 100 mm or 150 mm wafers
Small samples must be placed on dedicated carrier wafers. These do not need to and should not be bonded to the carriers.
Allowed materials
  • Silicon
  • Silicon Nitride
  • Aluminum Oxide
During previous testing, silicon nitride and aluminum oxide were only used as coating/masking materials and have been proven capable of sustaining this type of high temperature processing.


Graphic representation of the recommended annealing duration in respect to the annealing temperature. Image: Inês Diogo@DTU Nanolab, July 2023

Previous work on RTP Annealsys

More information and details about each type of processing - achievable using the RTP Annealsys tool - can be found in the following documents.

Important! The RTO sequences that were developed during the previous experimental work on the RTP Annealsys are not available to users. More tests and further investigation are required to prevent damaging the tool.