Specific Process Knowledge/Etch/AOE (Advanced Oxide Etch)/Remove resist in the AOE

From LabAdviser


Unless otherwise stated, this page is written by DTU Nanolab internal

Remove AZ resist in the AOE using the recipe Descum

Parameter Descum (also setting for TDESC10M 20241029)
Coil Power [W] 2000
Platen Power [W] 30
Platen temperature [oC] 0
O flow [sccm] 99
Pressure [mTorr] 10.0 (strike pressure 20.0 mTorr)


Typical results tested by bge@nanolab
AZ resist etch rate ~41 nm/s

BARC etching

I have tried once etching BARC with recipe SiO2_res:

Parameter Recipe: SiO2_res Recipe: Barcetch
Coil Power [W] 1300 400
Platen Power [W] 200 100
Platen temperature [oC] 0 0
He flow [sccm] 174 0
CF flow [sccm] 5 0
CF4 flow [sccm] 0 4
H flow [sccm] 4 20
Pressure [mTorr] 4 1
Etch rate in BARC [nm/sec] 2 1
Selectivity to DUV resist [Barc:resist] ? [1-2:1]