Specific Process Knowledge/Thermal Process/C4 Aluminium Anneal furnace: Difference between revisions

From LabAdviser
Jump to navigation Jump to search
 
(28 intermediate revisions by 4 users not shown)
Line 1: Line 1:
'''Feedback to this page''': '''[mailto:thinfilm@danchip.dtu.dk?Subject=Feed%20back%20from%20page%20http://labadviser.danchip.dtu.dk/index.php/Specific_Process_Knowledge/Thermal_Process/C4_Aluminium_Anneal_furnace click here]'''
'''Feedback to this page''': '''[mailto:thinfilm@danchip.dtu.dk?Subject=Feed%20back%20from%20page%20http://labadviser.danchip.dtu.dk/index.php/Specific_Process_Knowledge/Thermal_Process/C4_Aluminium_Anneal_furnace click here]'''
''This page is written by DTU Nanolab  internal''
[[Category: Equipment |Thermal C4]]
[[Category: Thermal process|C4]]
[[Category: Furnaces|C4]]


==Aluminium Anneal furnace (C4)==
==Aluminium Anneal furnace (C4)==
[[Image:C4helstak.jpg |thumb|300x300px|Aluminium Anneal furnace (C4). Positioned in cleanroom 2]]
[[Image:C4helstak.jpg |thumb|300x300px|Aluminium Anneal furnace (C4). Positioned in cleanroom B-1. /Photo: DTU Nanolab internal]]


The Aluminium Anneal furnace (C4) is a Tempress horizontal furnace for oxidation and annealing of silicon wafers with aluminium.
The Aluminium Anneal or Al-Anneal furnace (C4) is a Tempress horizontal furnace for annealing and oxidation of different samples.  


This furnace is the lowest of the furnace tubes in the furnace C-stack positioned in cleanroom 2. In this furnace allowed to process wafers that contain aluminium. Please check the cross contamination information in LabManager before you use the furnace.  
The samples can for instance be silicon and quartz wafers or small samples with aluminium or ALD deposited Al<sub>2</sub>O<sub>3</sub> and TiO<sub>2</sub>. Other materials might be allowed in the furnace, but this requires a permision from the Thin Film group. Also please check the cross contamination information in LabManager, before you use the furnace.
 
The furnace is the lowest of the C-stack furnaces positioned in cleanroom B-1. .  


'''The user manual, technical information and contact information can be found in LabManager:'''
'''The user manual, technical information and contact information can be found in LabManager:'''
Line 13: Line 20:


==Process knowledge==
==Process knowledge==
*Annealing: look at the [[Specific Process Knowledge/Thermal Process/Annealing|Annealing]] page
*Annealing: Look at the [[Specific Process Knowledge/Thermal Process/Annealing|Annealing]] page
*Oxidation: Look at the [[Specific Process Knowledge/Thermal Process/Oxidation|Oxidation]] page


==Overview of the performance of Aluminium Anneal furnace and some process related parameters==
==Overview of the performance of the Aluminium Anneal furnace and some process related parameters==


{| border="2" cellspacing="0" cellpadding="2"  
{| border="2" cellspacing="0" cellpadding="2"  
Line 23: Line 31:


|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
*Annealing of wafers with aluminium
*Annealing in N<sub>2</sub>
*Oxidation of wafers with aluminium
*Dry oxidation with O<sub>2</sub>
|-
|-
!style="background:silver; color:black" align="center" valign="center" rowspan="3"|Process parameter range
!style="background:silver; color:black" align="center" valign="center" rowspan="3"|Process parameter range
|style="background:LightGrey; color:black"|Process Temperature
|style="background:LightGrey; color:black"|Process Temperature
|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
*400-500 <sup>o</sup>C
*Normally 350-500 <sup>o</sup>C for wafers and samples with aluminium
*Up to 1150 <sup>o</sup>C for wafers and samples with other materials
|-
|-
|style="background:LightGrey; color:black"|Process pressure
|style="background:LightGrey; color:black"|Process pressure
|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
*1 atm
*1 atm (no vacuum)
|-
|-
|style="background:LightGrey; color:black"|Gas flows
|style="background:LightGrey; color:black"|Gas flows
|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
*N<sub>2</sub>: 0-10 slm
*N<sub>2</sub> (nitrogen): 0-10 SLM
*O<sub>2</sub>: 0-10 slm
*O<sub>2</sub> (oxygen): 0-10 SLM
|-
|-
!style="background:silver; color:black" align="center" valign="center" rowspan="2"|Substrates
!style="background:silver; color:black" align="center" valign="center" rowspan="2"|Substrates
|style="background:LightGrey; color:black"|Batch size
|style="background:LightGrey; color:black"|Batch size
|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
*1-30 100 mm wafers (or 50 mm wafers)  
*1-30 4" wafers (or 2" wafers)  
*1 6" wafer (2 6" wafers with less good uniformity) - Requires training 
*Small samples placed on a 6" dummy wafer - Requires traning
|-
|-
| style="background:LightGrey; color:black"|Substrate materials allowed
| style="background:LightGrey; color:black"|Substrate materials allowed
|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
*Silicon wafers with aluminium.
*Silicon and quartz samples with
*Wafers are allowed enter the furnace after aluminium lift-off or after aluminium etch and resist strip in acetone
**SiO<sub>2</sub> and Si<sub>3</sub>N<sub>4</sub>
|-  
**Aluminium. Wafers are allowed to enter the furnace after aluminium lift-off or aluminium etch and resist strip in acetone
**Al<sub>2</sub>O<sub>3</sub> and TiO<sub>2</sub> deposited by ALD
*Other materials might be allowed, but it requires a permission from the Thin Film group
|-
|}
|}
==Oxidation uniformity for 6" wafers==
In November 2023 an oxygen (O<sub>2</sub>) mass flow controller was connected to the furnace, so that it became possible to oxidize different wafers and samples in the furnace.
The C4 furnace is made for processing of 4" wafers placed vertically in a quartz boat. However, it is possible to process 6" wafers (or small samples placed on a 6" wafer), if they are placed horizontally on the boat.
Normally only one 6" wafer is being processed, because the temperature flat zone, i.e. the area inside the furnace where the temperature is uniform, is only a little longer than one quartz boat. But because there are two quartz boats available for the furnace, it has also been tried to place a 6" wafer on each boat and oxidize them at the same time as explained here.
<b>Sample loading:</b>
One 6" wafer was placed on each of the two quartz boats. The wafer flats were pointing towards each other, and the boats were centered above the arrow on the cantilever. The loading procedure is explained in the user manual for the furnace, but users should not do this without training, as the wafers might collide with the furnace tube, when the door closes (the tube diameter is 160 mm, and the wafer diameter is 150 mm). In order improve the temperature uniformity over the two 6" wafers, two 4" dummy wafers were placed in the end of each boat.
The 6" wafer closest to the furnace door/cleanroom is numbered "w1", and the wafer closest to the furnace/service room is numbered "w2". See the image below.
<b>Process parameters:</b>
For the oxidation, the following process parameters were used:
* Recipe: "DRY1050"
* Oxidation temperature: 1050 <sup>o</sup>C
* Oxidation time: 1 h 40 min
<b>Results:</b>
After the oxidation, the thickness of the oxide layer grown on the 6" wafers was measured in five points on each wafer using the Ellipsometer VASE.
The measurement results are shown below.
<i>Pernille Voss Larsen, DTU Nanolab, November 2023</i>
[[image:Oxidation test 1.jpg|600x600px|left|thumb|Two 6" wafers loaded in the C4 furnace. Two 4" dummy wafers are placed in the end of each boat.]]
[[image:Oxidation test 2.jpg |700x700px|left|thumb|Oxide thickness measured on the two 6" wafers]]

Latest revision as of 15:54, 15 December 2023

Feedback to this page: click here

This page is written by DTU Nanolab internal

Aluminium Anneal furnace (C4)

Aluminium Anneal furnace (C4). Positioned in cleanroom B-1. /Photo: DTU Nanolab internal

The Aluminium Anneal or Al-Anneal furnace (C4) is a Tempress horizontal furnace for annealing and oxidation of different samples.

The samples can for instance be silicon and quartz wafers or small samples with aluminium or ALD deposited Al2O3 and TiO2. Other materials might be allowed in the furnace, but this requires a permision from the Thin Film group. Also please check the cross contamination information in LabManager, before you use the furnace.

The furnace is the lowest of the C-stack furnaces positioned in cleanroom B-1. .

The user manual, technical information and contact information can be found in LabManager:

Aluminium Anneal furnace (C4)

Process knowledge

Overview of the performance of the Aluminium Anneal furnace and some process related parameters

Purpose
  • Annealing in N2
  • Dry oxidation with O2
Process parameter range Process Temperature
  • Normally 350-500 oC for wafers and samples with aluminium
  • Up to 1150 oC for wafers and samples with other materials
Process pressure
  • 1 atm (no vacuum)
Gas flows
  • N2 (nitrogen): 0-10 SLM
  • O2 (oxygen): 0-10 SLM
Substrates Batch size
  • 1-30 4" wafers (or 2" wafers)
  • 1 6" wafer (2 6" wafers with less good uniformity) - Requires training
  • Small samples placed on a 6" dummy wafer - Requires traning
Substrate materials allowed
  • Silicon and quartz samples with
    • SiO2 and Si3N4
    • Aluminium. Wafers are allowed to enter the furnace after aluminium lift-off or aluminium etch and resist strip in acetone
    • Al2O3 and TiO2 deposited by ALD
  • Other materials might be allowed, but it requires a permission from the Thin Film group


Oxidation uniformity for 6" wafers

In November 2023 an oxygen (O2) mass flow controller was connected to the furnace, so that it became possible to oxidize different wafers and samples in the furnace.

The C4 furnace is made for processing of 4" wafers placed vertically in a quartz boat. However, it is possible to process 6" wafers (or small samples placed on a 6" wafer), if they are placed horizontally on the boat.

Normally only one 6" wafer is being processed, because the temperature flat zone, i.e. the area inside the furnace where the temperature is uniform, is only a little longer than one quartz boat. But because there are two quartz boats available for the furnace, it has also been tried to place a 6" wafer on each boat and oxidize them at the same time as explained here.


Sample loading:

One 6" wafer was placed on each of the two quartz boats. The wafer flats were pointing towards each other, and the boats were centered above the arrow on the cantilever. The loading procedure is explained in the user manual for the furnace, but users should not do this without training, as the wafers might collide with the furnace tube, when the door closes (the tube diameter is 160 mm, and the wafer diameter is 150 mm). In order improve the temperature uniformity over the two 6" wafers, two 4" dummy wafers were placed in the end of each boat.

The 6" wafer closest to the furnace door/cleanroom is numbered "w1", and the wafer closest to the furnace/service room is numbered "w2". See the image below.


Process parameters:

For the oxidation, the following process parameters were used:

  • Recipe: "DRY1050"
  • Oxidation temperature: 1050 oC
  • Oxidation time: 1 h 40 min


Results:

After the oxidation, the thickness of the oxide layer grown on the 6" wafers was measured in five points on each wafer using the Ellipsometer VASE.

The measurement results are shown below.

Pernille Voss Larsen, DTU Nanolab, November 2023

Two 6" wafers loaded in the C4 furnace. Two 4" dummy wafers are placed in the end of each boat.
Oxide thickness measured on the two 6" wafers