Specific Process Knowledge/Lithography/Descum: Difference between revisions

From LabAdviser
Jump to navigation Jump to search
No edit summary
 
(10 intermediate revisions by the same user not shown)
Line 1: Line 1:
{{cc-nanolab}}
'''Feedback to this page''': '''[mailto:labadviser@nanolab.dtu.dk?Subject=Feed%20back%20from%20page%20http://labadviser.nanolab.dtu.dk/index.php?title=Specific_Process_Knowledge/Lithography/Descum click here]'''
'''Feedback to this page''': '''[mailto:labadviser@nanolab.dtu.dk?Subject=Feed%20back%20from%20page%20http://labadviser.nanolab.dtu.dk/index.php?title=Specific_Process_Knowledge/Lithography/Descum click here]'''
<!-- Replace "http://labadviser.danchip.dtu.dk/..." with the link to the Labadviser page-->


[[Category: Equipment |Lithography descum]]
[[Category: Equipment|Lithography descum]]
[[Category: Lithography|Descum]]
[[Category: Lithography|Descum]]


=Plasma asher 1=
__TOC__
 
=Plasma Asher 1=
[[File:Descum Results aug 2019.png|640px|thumb|right|Descum results plasma asher 1. September 2019]]
[[File:Descum Results aug 2019.png|640px|thumb|right|Descum results plasma asher 1. September 2019]]


'''The user manual(s), quality control procedure(s) and results and contact information can be found in [http://labmanager.dtu.dk/function.php?module=Machine&view=view&mach=55 LabManager].'''
The user manual(s), quality control procedure(s) and results and contact information can be found in [http://labmanager.dtu.dk/function.php?module=Machine&view=view&mach=55 LabManager] - '''requires login'''


Descum of AZ5214E resist on 50mm silicon wafer. Wafer was placed horisontally in chamber on a 100 mm carier wafer.
Descum of AZ5214E resist on 50mm silicon wafer. Wafer was placed horisontally in chamber on a 100 mm carier wafer.
Line 20: Line 23:
{| {{table}}
{| {{table}}
| align="center" |  
| align="center" |  
{| border="1" cellspacing="1" cellpadding="2" align="left"
{|border="1" cellspacing="1" cellpadding="10" style="text-align:left;"  
|- style="background:LightGrey"
|- style="background:LightGrey"
|'''Ashing time (min)'''|| 1|| 2 || 5 || 7 || 10 || 10   
|'''Ashing time (min)'''|| 1|| 2 || 5 || 7 || 10 || 10   
Line 37: Line 40:
{| {{table}}
{| {{table}}
| align="center" |  
| align="center" |  
{| border="1" cellspacing="1" cellpadding="2" align="left"
{|border="1" cellspacing="1" cellpadding="10" style="text-align:left;"  
|- style="background:LightGrey"
|- style="background:LightGrey"
|'''Ashing time (min)'''|| 1|| 2 || 5 || 7 || 10 || 10   
|'''Ashing time (min)'''|| 1|| 2 || 5 || 7 || 10 || 10   
Line 64: Line 67:
{| {{table}}
{| {{table}}
| align="center" |  
| align="center" |  
{| border="1" cellspacing="1" cellpadding="2" align="center"
{|border="1" cellspacing="1" cellpadding="10" style="text-align:left;"  
   
   
|- style="background:LightGrey"
|- style="background:LightGrey"
Line 78: Line 81:
{| {{table}}
{| {{table}}
| align="center" |  
| align="center" |  
{| border="1" cellspacing="1" cellpadding="2" align="center"
{|border="1" cellspacing="1" cellpadding="10" style="text-align:left;"  
   
   
|- style="background:LightGrey"
|- style="background:LightGrey"
Line 92: Line 95:
{| {{table}}
{| {{table}}
| align="center" |  
| align="center" |  
{| border="1" cellspacing="1" cellpadding="2" align="center"
{|border="1" cellspacing="1" cellpadding="10" style="text-align:left;"  
   
   
|- style="background:LightGrey"
|- style="background:LightGrey"
Line 106: Line 109:
{| {{table}}
{| {{table}}
| align="center" |  
| align="center" |  
{| border="1" cellspacing="1" cellpadding="2" align="center"
{|border="1" cellspacing="1" cellpadding="10" style="text-align:left;"  
   
   
|- style="background:LightGrey"
|- style="background:LightGrey"
Line 121: Line 124:
[[image:descum_graf.jpg|640px|thumb|Descum results plasma asher 2 - recipe 1]]
[[image:descum_graf.jpg|640px|thumb|Descum results plasma asher 2 - recipe 1]]


'''The user manual(s), quality control procedure(s) and results and contact information can be found in [http://labmanager.dtu.dk/function.php?module=Machine&view=view&mach=200 LabManager].'''
The user manual(s), quality control procedure(s) and results and contact information can be found in [http://labmanager.dtu.dk/function.php?module=Machine&view=view&mach=200 LabManager] - '''requires login'''


Descum of AZ  Mir 701 resist on 100mm silicon wafer. Five wafers were placed vertically in chamber.
Descum of AZ  Mir 701 resist on 100mm silicon wafer. Five wafers were placed vertically in chamber.
Line 132: Line 135:
{| {{table}}
{| {{table}}
| align="center" |  
| align="center" |  
{| border="1" cellspacing="1" cellpadding="2" align="center"
{|border="1" cellspacing="1" cellpadding="10" style="text-align:left;"  
   
   
|- style="background:LightGrey"
|- style="background:LightGrey"
Line 153: Line 156:
{| {{table}}
{| {{table}}
| align="center" |  
| align="center" |  
{| border="1" cellspacing="1" cellpadding="2" align="center"
{|border="1" cellspacing="1" cellpadding="10" style="text-align:left;"  
   
   
|- style="background:LightGrey"
|- style="background:LightGrey"
Line 168: Line 171:
<br clear="all" />
<br clear="all" />


=Plasma Asher 3=
=Plasma Asher 3: Descum=
'''The user manual(s), quality control procedure(s) and results and contact information can be found in [http://labmanager.dtu.dk/function.php?module=Machine&view=view&mach=423 LabManager].'''
The user manual(s), quality control procedure(s) and results and contact information can be found in [http://labmanager.dtu.dk/function.php?module=Machine&view=view&mach=423 LabManager] - '''requires login'''


Plasma Asher 3 is specifically used for controlled descum process after lithography. Please note that you only can process a single 100 mm wafer, or one small sample, at a time.
Plasma Asher 3 is specifically used for controlled descum process after lithography. Please note that you only can process a single 100 mm wafer, or one small sample, at a time.
Line 192: Line 195:
{| {{table}}
{| {{table}}
| align="center" |  
| align="center" |  
{| border="1" cellspacing="1" cellpadding="2" align="center"
{|border="1" cellspacing="1" cellpadding="10" style="text-align:left;"  
|- style="background:LightGrey"
|- style="background:LightGrey"
| ||FW/REV|| C2/C1 || Power
| ||FW/REV|| C2/C1 || Power
Line 220: Line 223:
{| {{table}}
{| {{table}}
| align="center" |  
| align="center" |  
{| border="1" cellspacing="1" cellpadding="2" align="center"
{|border="1" cellspacing="1" cellpadding="10" style="text-align:left;"  
|- style="background:LightGrey"
|- style="background:LightGrey"
| ||FW/REV|| C2/C1 || Oxygen || Pressure
| ||FW/REV|| C2/C1 || Oxygen || Pressure
Line 246: Line 249:
{| {{table}}
{| {{table}}
| align="center" |  
| align="center" |  
{| border="1" cellspacing="1" cellpadding="2" align="center"
{|border="1" cellspacing="1" cellpadding="10" style="text-align:left;"  
|- style="background:LightGrey"
|- style="background:LightGrey"
| ||FW/REV|| C2/C1 || Oxygen || Pressure
| ||FW/REV|| C2/C1 || Oxygen || Pressure

Latest revision as of 14:55, 10 May 2023

The contents on this page, including all images and pictures, was created by DTU Nanolab staff unless otherwise stated.

Feedback to this page: click here

Plasma Asher 1

Descum results plasma asher 1. September 2019

The user manual(s), quality control procedure(s) and results and contact information can be found in LabManager - requires login

Descum of AZ5214E resist on 50mm silicon wafer. Wafer was placed horisontally in chamber on a 100 mm carier wafer.


Recipe 1:
Note: Plasma asher was cold before use

  • O2 flow: 70 ml/min
  • N2 flow: 70 ml/min
  • Power: 150 W
Ashing time (min) 1 2 5 7 10 10
Etched Thickness (nm) 14.2 16.3 47.6 123.2 854.3 862.1


Recipe 2:
Note: Plasma asher was cold before use

  • O2 flow: 500 ml/min
  • N2 flow: 0 ml/min
  • Power: 500 W
Ashing time (min) 1 2 5 7 10 10
Etched Thickness (nm) - 8.1 32.9 271.1 495.6 446.2


Descum tests on UV resists

Conny Hjort & Jesper Hanberg, September 2021

Descum results plasma asher 1. August 2021

Descum of different resists, AZ5214E, AZ701 MiR and AZ2020 nLOF, on a single 100mm wafer was tested. Wafer was placed vertically in the middle of glass carrier.

Recipe settings:
Note: plasma Asher was cold before use.

  • O2 flow: 70 ml/min
  • N2 flow: 70 ml/min

Power: 150 W

Minor temperature rise during processing was observed, but not more than 5 degrees. Starting chamber pressure was around 0.5 mbar.


1,5 um AZ5214E resist:

Ashing time (min) 1 2 3 5 7
Etched Thickness (nm) 6,28 102,99 76,92 N/A N/A


1,5 um AZ5214E resist placed horizontally in the carrier:

Ashing time (min) 1 2 3 5 7
Etched Thickness (nm) 63,03 143,32 304,29 372,59 N/A


1,5 um AZ701MiR resist:

Ashing time (min) 1 2 3 5 7
Etched Thickness (nm) 268,88 199,54 219,03 200,86 292,15


1,5 um AZ 2020nLOF resist:

Ashing time (min) 1 2 3 5 7
Etched Thickness (nm) 1,68 76,51 169,72 481,96 272,59


Plasma Asher 2

Jitka Urbánková & Jesper Hanberg, December 2019

Descum results plasma asher 2 - recipe 1

The user manual(s), quality control procedure(s) and results and contact information can be found in LabManager - requires login

Descum of AZ Mir 701 resist on 100mm silicon wafer. Five wafers were placed vertically in chamber.


recipe 1:

  • O2 flow: 100 ml/min
  • N2 flow: 100 ml/min
  • Power: 150 W
Ashing time (min) 1 2 3 4 6 7 8 9 10 12 14 15 20
Etched Thickness (nm) 8,7 5,1 12,5 6,2 31,8 86,0 25,7 46,8 38,3 49,7 59,4 140,1 360,7
Initial temperature (°C) 28 21 31 21 22 28 25 24 21 24 24 22 22
Descum results plasma asher 2 - recipe 2


recipe 2:

  • O2 flow: 500 ml/min
  • N2 flow: 0 ml/min
  • Power: 200 W
Ashing time (min) 1 2 3 4 5 6 7 8 10 12 15 20
Etched Thickness (nm) 8,1 9,4 16,8 55,2 44,0 47,5 42,5 55,1 85,3 122,4 184,8 305,9
Initial temperature (°C) 22 21 21 22 22 22 21 21 20 21 21 22

A linear time dependence was observed after etching 7 minutes or more (recipe 2).

Plasma Asher 3: Descum

The user manual(s), quality control procedure(s) and results and contact information can be found in LabManager - requires login

Plasma Asher 3 is specifically used for controlled descum process after lithography. Please note that you only can process a single 100 mm wafer, or one small sample, at a time. The plasma asher is equipped with 2 gaslines: oxygen and nitrogen, but all standard processes use only oxygen (as recommended by Diener).


Ashing of AZ MiR701 resist:
You can use two different descum process developments: you can either change power settings or processing chamber pressure.


Testing different power settings:

Descum results for different power settings

Recipe settings:

  • O2 flow: 5 sccm
  • N2 flow: 0
  • Pressure: 0.2 mbar
  • Power: Varied


Experiment parameters:

FW/REV C2/C1 Power
recipe 1 50/0 52/31 50%
recipe 2 100/0 53/31 100%
recipe 3 20/0 51/34 20%



Testing different pressure settings:

Descum results for different pressure settings

Recipe settings:

  • O2 flow: varied
  • N2 flow: 0
  • Pressure: varied
  • Power: V100% (100 W)


Experiment parameters:

FW/REV C2/C1 Oxygen Pressure
recipe 1 100/0 53/31 5 0,2
recipe 2 100/0 37/38 45 0,8


Ashing of AZ5214E resist:

Descum results for different pressure settings


Recipe settings:

  • O2 flow: varied
  • N2 flow: 0
  • Pressure: varied
  • Power: V100% (100 W)


Experiment parameters:

FW/REV C2/C1 Oxygen Pressure
recipe 1 100/0 53/31 17 0,4
recipe 2 100/0 37/39 45 0,8